使用卷压力记录,一条尾巴袖套法小鼠的血压测量

, , ,

Cardiovascular Research Center, University of Kentucky

 

Video Article Chapters

Cite this Article: 使用卷压力记录,一条尾巴袖套法小鼠的血压测量

Daugherty, A., Rateri, D., Hong, L., Balakrishnan, A. Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method. J. Vis. Exp. (27), e1291, doi:10.3791/1291 (2009).

Abstract: 使用卷压力记录,一条尾巴袖套法小鼠的血压测量

尾声,8通道高通量非侵入性血压系统测量血压,同时8只小鼠或大鼠。尾声尾袖系统使用卷压力记录(VPR)来衡量确定的尾巴血容量血压。一个特别设计的差动压力传感器和一个闭塞的尾巴袖口措施在尾部的总血量,而不需要获得个人的脉冲信号。特别注意的是给予闭塞袖口的长度,为了获得最准确的血压读数。 VPR可以很容易地获得对皮肤黝黑的啮齿类动物,如C57BL6小鼠,的读数和核磁共振兼容。尾声系统提供的六(6)不同的血压参数的测量,收缩压和舒张压,心率,平均血压,尾巴的血流量,和尾血容量。测量,可无论是清醒或麻醉的小鼠或大鼠。尾声系统包括一个控制器,笔记本电脑,软件,袖口,动物持有人,红外变暖垫,红外测温仪。有七个小大900克8克大鼠,小鼠的不同规模的持有人。

Protocol: 使用卷压力记录,一条尾巴袖套法小鼠的血压测量

  1. 设备设置
    1. 选择面积在实验室中,室温在20 ° C或以上
    2. 避免在空调通风口和/或大声喧哗附近的位置。
    3. 避免暴露动物的气味,可能会引起刺激和/或压力,如香水,古龙水,化学品,和清洁工,。
    4. 打开尾声控制器。
    5. 打开变暖的平台,并设置为3级。如果室温高于25 ° C时,启动1级或2级。
    6. 控制器诊断测试
      1. 打开尾声软件。
      2. 尾声设备上按一下选择。
      3. 点击“选择测试设备”。
      4. 选择袖口和渠道进行测试,然后单击“测试”。
      5. 测试完成后,关闭“设备测试”窗口。
  2. 软件设置
    1. 管理人员和动物
      1. 管理人员
        1. 选择“工具”>“管理人员”。
        2. “研究员”,点击更改名称。
        3. 在随后的行,点击添加额外的研究人员。
        4. “技术员标签”,点击修改或增加的技术人员。
        5. 点击“标本”选项卡,以易于识别的名称的动物和选择动物类型;鼠或鼠标。
        6. “动物”选项卡上点击改变的动物种类和灵敏度。
    2. 新实验
      1. 选择File> New>实验
      2. 开始一个新的实验
        1. 输入实验的名称。
        2. 选择的主要研究人员。
        3. 选择开始日期,然后单击“下一步”。
      3. 基本会话信息
        1. 输入会议名称
        2. 设置的“驯化周期”至5。如果你不喜欢,包括驯化周期设置为零。为了协助acclimating持有人及袖口的动物,你的实验中添加额外的周期和运行连续5天的实验。
        3. “套数”设置为1。
        4. “集之间的时间”设置为30秒多套
        5. 设置“循环”,每套20个周期。
        6. “周期之间的时间”设置为5秒,点击“下一步”。
    3. 样本的选择
        1. 选择从标本池的标本,将其分配给第一个可用的通道,并点击“下一步”。
      1. 会话参数
        1. 设置250mmHg“最大闭塞压”。
        2. 设置“通缩时代”小鼠,大鼠15秒至20秒。
        3. 设置“最低交易量”到15ul。
        4. 设置“显示样式”,每图的一个通道。
        5. 添加额外的研究人员和技术,并点击“下一步”。
      2. 会话脚本
        1. 审查会话脚本,点击“下一步”。
        2. 不要选择完成,直到您完成动物准备部分。
  3. 动物的制备
    1. 动物设置的
      1. 动物持有人
        1. 培养和提高动物的驯化,添加额外的周期,您的实验和运行连续五天的实验。您还可以协助配售持有人的动物,15分钟,连续3天前的实际研究与驯化。
        2. 持有人拿起动物的尾巴,轻轻放入动物的持有人面临的鼻锥开口端的后方放置在每个动物。注:动物在选定的持有人的大小尺寸。动物应该能自由进入的持有人。
        3. 小心地保护持有人的后舱门打开后舱门红色的螺丝帽。应注意避免捏尾巴或身体其他部位,同时确保后方舱口。
        4. 滑向后方舱口的鼻锥,限制动物的运动。鼻锥,应在一个位置,限制从动物转向左右,而里面的持有人。
        5. 将变暖平台上的持有人在指定的位置。如果测试通过一段时间的同一组的动物,可以肯定的旋转变暖的平台上持有人的地位,提高你的血压测量的一致性和准确性。允许的动物,至少5分钟,以适应环境的持有人。
        6. 请勿触摸或处理的动物,而里面的持有人。增加接触可以刺激动物。
        7. 切勿持有人留在无人看管的动物。
      2. 袖口安置
          穿过尾巴“闭塞的袖口。”广场袖口尽可能靠近基地的尾巴,没有力量。
        1. 安全的“闭塞袖口油管”持有人的后上方的缺口。
        2. 穿过“VPR袖口”的尾巴,它放置在2mm的“闭塞袖口”。
        3. 保护持有人的后上方的缺口“VPR袖口油管”。
        4. 将袖口尾声控制器。
      3. 动物温度
        1. 允许的动物,至少5分钟,以thermoregulate。
        2. 经常用红外测温仪记录动物的温度。动物的温度应在32至35 ° C。
  4. 实验
    1. 开始实验
      1. 点击“完成”,并实验将立即开始。
    2. 结束实验
      1. 动物
        1. 立即从袖口和持有人的动物。
        2. 返回后的笼子里的动物的“简单的会话总结报告”显示。
      2. 数据收集
        1. 接受周期尾声应用程序的试算表内格式将被自动显示最后一个周期完成后,点击“导出到Excel”按钮。
        2. 选择放置文件的文件夹,命名该文件,并点击“保存”。
        3. 打开Excel应用程序,然后单击“打开”。
        4. 查看所有文件类型,以显示“。csv文件保存”,在文件上单击。
      3. 数据处理
        1. 有几种方法来处理Excel文件内的数据。通常的做法是获得的平均值和标准偏差。
        2. 删除测量,如果标准偏差大于30。
        3. 此外,拒绝周期可能被视为整个数据库可以很容易地导出到Excel。咨询用户的其他信息手册。

Disclosures: 使用卷压力记录,一条尾巴袖套法小鼠的血压测量

Ask the Author: 使用卷压力记录,一条尾巴袖套法小鼠的血压测量

8 Comments

Do you have any idea how does this system compares with similar product from other vendors like Harvard Apparatus?

1

Reply

Posted by: Mahesh JoshiAugust 31, 2009, 12:53 PM

We have used the Kent and Visitech machines. The bottomline is that no system is perfect and they are highly operator dependent. We have found the Kent machines require more training (for humans and mice!) but the data is more better quality and more similar to telemetry.

1.1

Reply

Posted by: Alan DaughertySeptember 5, 2009, 1:58 PM

why most of the time reading in this experiment shows fail tail volume?

2

Reply

Posted by: SushantNovember 16, 2009, 4:34 AM

is this Available in the philippines

3

Reply

Posted by: aaronFebruary 9, 2012, 7:45 AM

Yes it is, please contact sales@kentscientific.com to obtain a quote.

3.1

Reply

Posted by: David FitzMillerFebruary 9, 2012, 8:43 AM

1) We do stroke surgery. Can we measure the blood pressure continuosly during 1 hour surgery.If not can we measure blood pressure at different time point during the surgery.If i have to measure blood pressure at different time point during the surgery, then do i have to set up the cuff placement on the base of the tail again everytime.

2)In the protocol, occlusion cuff has to placed as close to the base of the tail as possible.During the mice stroke surgery, we insert the rectal probe into the anus. would the placement of the occlusion cuff interfere with the placement of the rectal probe into the anus. we would not use the animal holder because our animals are anesthesized for the stroke surgery.

4

Reply

Posted by: Varun k.May 8, 2012, 3:21 PM

1)What is the purpose of thermal regulation of the animal?During the stroke surgery, our animals are maintianed at 37 celsius. In our case, do we need to thermoregulate the animals at 32 celsius for measuring blood pressure.Will the maintainance of body temperature at 37 cel interfere the accuracy of the measurement of blood pressure.

5

Reply

Posted by: AnonymousMay 8, 2012, 4:20 PM

Good Day sir/mam,
We are having our research problem about the feasibility of okra extract as therapy to hypertension. and our experimental subject is mice. therefore,i would like to ask if where can we avail this equipment in the Philippines? thank you for more suggestions just contact me at this email dvanirish@yahoo.com thank you.!

10

Reply

Posted by: van d.July 3, 2012, 7:26 AM

I am a pharmacy student one of University in Indonesia.
I'm doing research studies (to pass on degree) about antihypertensive using CODA system.
I've followed the procedure properly.
But, why the time reading in this experiment always shows fail tail volume?
Thank you for more suggestions please contact me at this email nadyanovyani@gmail.com.
Thanks a lot..

11

Reply

Posted by: Nadya N.March 27, 2013, 8:14 PM

I am a pharmacy student from Padjadjaran University in Indonesia.
I'm doing research studies (to pass on degree) about antihypertensive using CODA system. the first time I used this device, it could work very well. the device could read the blood pressure and said "OK". but after a couple of days, it couldn't work very well. everytime I used this device, it always said "fail tail volume" even though I have followed the procedure properly and the rats I used in this experiment are in good condition (no smell and no noise). So, I would like to ask you how to solve this problem.This is the first time our faculty use this device and I would really appreciate it if you could give me the reply as soon as possible. Thanking you in advance.
Sincerenly
Nadya

13

Reply

Posted by: Nadya N.March 28, 2013, 2:31 AM

Post a Question / Comment / Request

You must be signed in to post a comment. Please or create an account.

Waiting
simple hit counter