A dual raster-scanning photoacoustic imager was designed, which integrated wide-field imaging and real-time imaging.
Imaging of vascular networks on small animals has played an important role in basic biomedical research. Photoacoustic imaging technology has great potential for application in the imageology of small animals. The wide-field photoacoustic imaging of small animals can provide images with high spatiotemporal resolution, deep penetration, and multiple contrasts. Also, the real-time photoacoustic imaging system is desirable to observe the hemodynamic activities of small-animal vasculature, which can be used to research the dynamic monitoring of small-animal physiological features. Here, a dual-raster-scanning photoacoustic imager is presented, featuring a switchable double-mode imaging function. The wide-field imaging is driven by a two-dimensional motorized translation stage, while the real-time imaging is realized with galvanometers. By setting different parameters and imaging modes, in vivo visualization of small-animal vascular network can be performed. The real-time imaging can be used to observe pulse change and blood flow change of drug-induced, etc. The wide-field imaging can be used to track the growth change of tumor vasculature. These are easy to be adopted in various areas of basic biomedicine research.
In the basic biomedical field, small animals can simulate human physiological function. Therefore, small-animal imaging plays an important role in guiding the research of human homologous diseases and seeking effective treatment1. Photoacoustic imaging (PAI) is a non-invasive imaging technique combining the advantages of optical imaging and ultrasound imaging2. Photoacoustic microscopy (PAM) is a valuable imaging method for basic research of small animal3. PAM can easily obtain high-resolution, deep-penetration, high-specificity and high-contrast images based on optical excitation and ultrasound detection4.
A pulse laser with a specific wavelength is absorbed by endogenous chromophores of tissues. Subsequently, the temperature of the tissue rises, which results in the production of photo-induced ultrasonic waves. The ultrasonic waves can be detected by an ultrasonic transducer. After signal acquisition and image reconstruction, the spatial distribution of the absorber can be obtained5. On the one hand, the visualization of whole-organ vascular network requires a wide field of view. The process of wide-field scanning usually takes a long time to ensure high-resolution6,7,8. On the other hand, observing the hemodynamic activities of small animals requires fast real-time imaging. The real-time imaging is beneficial to study the vital signs of small animals in real time9,10,11. The field of view of real-time imaging is usually sufficiently small to ensure a high update rate. Thus, there is often a tradeoff between achieving a wide field of view and real-time imaging. Previously, two different systems were used for wide-field imaging or real-time imaging, separately.
This work reports a dual raster-scanning photoacoustic imager (DRS-PAI), which integrated wide-field imaging based on a two-dimensional motorized translation stage and real-time imaging based on a two-axis galvanometer scanner. The wide-field imaging mode (WIM) is performed to show vascular morphology. For the real-time imaging mode (RIM), there are currently two functions. First, RIM can provide real-time B-scan images. By measuring the displacement of vasculature along the depth direction, the characteristics of respiration or pulse can be revealed. Second, the RIM can quantitatively measure the specific area in the WIM image. By providing comparable images of local WIM regions, the details of the local change can be accurately revealed. The system designs a flexible transition between wide-field imaging of vascular visualization and real-time imaging of the local dynamic. The system is desirable in basic biomedical research where there is a need for small-animal imaging.
All animal experiments were performed in compliance with guidelines provided by the institutional animal care and use committee of South China Normal University, Guangzhou, China.
1. System Setup
2. System alignment
3. Animal experiment
The schematic of the DRS-PAI is shown in Figure 1. The system allows flexible and repeatable switching between WIM with RIM. The acquired PA signal is processed quickly to generate PA B-Scan and MAP images. The CCD camera can provide photographs of samples.
All components of the DRS-PAI are integrated and assembled in an imager setup (Figure 2), making it easy to assemble and operate. In the WIM, continuous raster scanning of a two-dimensional motorized stage is used. The signal of the running stage is recorded. The data acquisition proceeded during uniform translation of the stage. In the RIM, a two-axis galvanometer scanner was used. The data were collected synchronously with the Galva scanning (Figure 3).
Here, the vascular images of samples with each imaging mode were collected. Figure 4A shows the MAP image of the mouse back in WIM. The imaging time was about 33 min. Figure 4B shows B-scan images of mouse back during RIM. The whole process of RIM is shown in Video 1. Then, a focused ultrasonic transducer was used. The vascular networks of the mouse ear and brain are shown in Figure 5. The imaging time was about 16 min. This demonstrates the ability of DRS-PAI to image wide-field vasculature. In addition, Figure 6A shows that the imaging range contains a vessel. The imaging range of RIM is about 100 μm due to the use of the focused ultrasonic transducer. The displacement image along the depth direction of the mouse abdomen versus time is shown in Figure 6B. Video 2 shows the process of vascular displacement and obtaining the current pulse or respiration curve.
Figure 1: The schematic of DRS-PAI system. Please click here to view a larger version of this figure.
Figure 2: The design of DRS-PAI system.
(A) The photograph of DRS-PAI system. (B) The panel shows a photograph of the setup for laser path assembly. (C) The panel shows the 3D model for laser path assembly. (D) The panel shows the two-axis fast galvanometer scanner assembly. (E) The panel shows the probe assembly. (F) The panel shows the CCD optical path assembly. Please click here to view a larger version of this figure.
Figure 3: The setup of scan for different imaging modes.
(A) The scan path of WIM. (B) The scan path of RIM. (C) The trigger setup of two imaging modes. Please click here to view a larger version of this figure.
Figure 4: The photoacoustic WIM and RIM of the mouse back.
(A) The MAP image of mouse back in WIM. (B) The B-Scan images of mouse back in RIM. Please click here to view a larger version of this figure.
Figure 5: The photoacoustic WIM of a mouse.
(A) The MAP image of the mouse ear in WIM. (B) The MAP image of the mouse brain in WIM. Please click here to view a larger version of this figure.
Figure 6: The photoacoustic RIM of the mouse abdomen.
(A) The B-scan images of the mouse abdomen in RIM. (B) The MAP image along the depth direction of the mouse abdomen versus time in RIM. Please click here to view a larger version of this figure.
Video 1: The process of RIM of the mouse back. Please click here to download this video.
Video 2: The process of MAP image along the depth direction of the mouse abdomen versus time. Please click here to download this video.
Supplementary Figure 1: The part of the imaging region in contact with the polyethylene membrane. Please click here to download this figure.
Here we presented a dual raster-scanning photoacoustic small-animal imager for noninvasive vascular visualization which was designed and developed to capture the structure of the vasculature and the related dynamic change of blood. The advantage of DRS-PAI is that it integrates the WIM and the RIM into one system, which makes it easier to study vascular dynamic and vascular network structure of small animals. The system can provide high-resolution wide-field vascular visualization and real-time blood dynamics.
In the current system, the optical excitation was implemented with a single-wavelength light source. A future multi-wavelength system would provide other parameters such as blood oxygen saturation. Further, a special image processing algorithm can be developed for quantitative analysis, including estimating vascular diameter, vascular density, vascular tortuosity, etc. The quantitative analysis can provide valuable information for early diagnosis and treatment of diseases.
In summary, the system enables researchers to obtain high-dimensional physiological and pathological insights into small-animal research with biomedical relevance. The system can be adapted to most small-animal research settings, include but not limited to, imaging of angiogenesis, tumor microenvironments, hemodynamic, functional connections in the brain, microcirculation, drug responses, and therapy responses. Critical steps within the protocol include the design of the dual scanning structure, the confocal adjustment of the optical and acoustic focus in the WIM, and center point adjustment of the sound field in the RIM.
The authors have nothing to disclose.
The authors would like acknowledge the financial support from National Natural Science Foundation of China (61822505; 11774101; 61627827; 81630046), The Science and Technology Planning Project of Guangdong Province, China (2015B020233016), and The Science and Technology Program of Guangzhou (No. 2019020001).
12 bit multi-purpose digitizer | Spectrum | M3i.3221 | Data acquisition card |
A-line collected program | National Instrument | LabVIEW | User-defined program |
Amplifier | RF Bay | LNA-650 | Amplifier |
Depilatory Cream | Veet | 33-II | Animal depilatory |
Fiberport Coupler | Thorlab | PAF-X-7-A | Fiber Coupler |
Field Programmable Gate Array | Altera | Cyclone IV | Trigger Control |
Fixed Focus Collomation Packages | Thorlabs | F240FC-532 | Fiber Collimator |
Foused ultrasonic transducer | Self-made | ||
Graphics Processing Unit | NVIDIA | GeForce GTX 1060 | Processing data |
Holder | Self-made | Animal fixation | |
Laser control program | National Instrument | LabVIEW | User-defined program |
Mice | Guangdong Medical Laboratory Animal Center | BALB/c | Animal Model |
Microscope camera | Mshot | MS60 | CCD camera |
Microscope Objective | Daheng Optics | GCO-2111 | Objective Lens |
Mirror | Daheng Optics | GCC-1011 | Moveable/Fixed Mirror |
Moving Magnet Capacitive Detector Galvanometer Scanner | Century Sunny | S8107 | real-time scanner |
Mshot image analysis system | Mshot | Display software | |
Normal Saline | CR DOUBLE-CRANE | H34023609 | Normal Saline |
Ophthalmic Scissors | SUJIE | Scalp Remove | |
Planar ultrasonic transducer | Self-made | ||
Plastic Wrap | HJSJLSL | Polyethylene Membrane | |
Program Control Software | National Instrument | LabVIEW | User-defined Program |
Pulsed Q-swithched Laser | Laser-export | DTL-314QT | 532-nm pulse Laser |
Real-time imaging program | National Instrument | LabVIEW | User-defined program |
Ring-shaped white LED | Self-made | ||
Shaver | Codos | CP-9200 | Animal Shaver |
Single-Mode Fibers | Nufern | 460-HP | Single-mode fiber |
Surgical Blade | SUJIE | 11 | Blade |
Surgical Scalpel | SUJIE | 7 | Scalp Remove |
Translation Stage | Jiancheng Optics | LS2-25T | wide-field scanning stage |
Ultrasonic Transducer | Self-made | ||
Ultrasound gel | GUANGGONG PAI | ZC4252418 | Acoustic Coupling |
Urethane | Tokyo Chemical Industry | C0028 | Animal Anestheized |
Water tank | Self-made | ||
Wide-field imaging program | National Instrument | LabVIEW | User-defined program |
XY Translator Mount | Self-made |