A Microfluidic-based Hydrodynamic Trap for Single Particles

Published: January 21, 2011 doi: 10.3791/2517

Summary

In this article, we present a microfluidic-based method for particle confinement based on hydrodynamic flow. We demonstrate stable particle trapping at a fluid stagnation point using a feedback control mechanism, thereby enabling confinement and micromanipulation of arbitrary particles in an integrated microdevice.

Abstract

The ability to confine and manipulate single particles in free solution is a key enabling technology for fundamental and applied science. Methods for particle trapping based on optical, magnetic, electrokinetic, and acoustic techniques have led to major advancements in physics and biology ranging from the molecular to cellular level. In this article, we introduce a new microfluidic-based technique for particle trapping and manipulation based solely on hydrodynamic fluid flow. Using this method, we demonstrate trapping of micro- and nano-scale particles in aqueous solutions for long time scales. The hydrodynamic trap consists of an integrated microfluidic device with a cross-slot channel geometry where two opposing laminar streams converge, thereby generating a planar extensional flow with a fluid stagnation point (zero-velocity point). In this device, particles are confined at the trap center by active control of the flow field to maintain particle position at the fluid stagnation point. In this manner, particles are effectively trapped in free solution using a feedback control algorithm implemented with a custom-built LabVIEW code. The control algorithm consists of image acquisition for a particle in the microfluidic device, followed by particle tracking, determination of particle centroid position, and active adjustment of fluid flow by regulating the pressure applied to an on-chip pneumatic valve using a pressure regulator. In this way, the on-chip dynamic metering valve functions to regulate the relative flow rates in the outlet channels, thereby enabling fine-scale control of stagnation point position and particle trapping. The microfluidic-based hydrodynamic trap exhibits several advantages as a method for particle trapping. Hydrodynamic trapping is possible for any arbitrary particle without specific requirements on the physical or chemical properties of the trapped object. In addition, hydrodynamic trapping enables confinement of a "single" target object in concentrated or crowded particle suspensions, which is difficult using alternative force field-based trapping methods. The hydrodynamic trap is user-friendly, straightforward to implement and may be added to existing microfluidic devices to facilitate trapping and long-time analysis of particles. Overall, the hydrodynamic trap is a new platform for confinement, micromanipulation, and observation of particles without surface immobilization and eliminates the need for potentially perturbative optical, magnetic, and electric fields in the free-solution trapping of small particles.

Protocol

The hydrodynamic trap consists of a two-layer hybrid (polydimethylsiloxane (PDMS) /glass) microfluidic device for particle confinement. Steps 1-2 describe fabrication of microfluidic devices, and Steps 3-4 discuss device design and operation.

1. SU-8 Mold Fabrication (not shown in video)

  1. Clean two silicon wafers (3" diameter) with acetone and isopropyl alcohol (IPA).
  2. Dry wafers with N2 and place them on a hotplate at 65°C for 1 min to remove residual moisture.
  3. Spin coat wafer #1 with SU-8 2050 photoresist (PR) for 30 sec at 4000 rpm to create a ~40 μm thick mold for the fluidic layer. Spin coat wafer #2 with PR for 30 sec at 1500 rpm to create a ~150 μm thick mold for the control layer.
  4. Soft bake wafer #1 at 65°C for 3 min and then at 95°C for 6 min. Soft bake wafer #2 at 65°C for 5 min and then at 95°C for 20 min.
  5. Expose wafers to UV with their respective masks (wafer #1: ports and fluidic channels, wafer #2: port and control layer) and appropriate exposure intensity (~150 mJ/cm2, ~260 mJ/cm2 respectively).
  6. Post bake wafer #1 at 65°C for 1 min and then at 95°C for 6 min. Post bake wafer #2 at 65°C for 5 min and then at 95°C for 10 min.
  7. Develop wafers with propylene glycol methyl ether acetate (PGMEA) until uncured PR is removed. Rinse wafers with IPA and dry with N2.

2. Microfluidic Device Fabrication

  1. Silanize the surface of the SU-8 molds by placing the wafers in a desiccator under vacuum for ~10 min with a glass dish containing a few drops of trichlorosilane. Surface silanization assists in peeling the (PDMS) replicas off the SU-8 molds.
  2. Mix and degas PDMS in base:crosslinker ratios of 15:1 and 5:1 for the fluidic and control layers respectively.
  3. Spin coat the 15:1 PDMS mixture onto the fluidic layer mold (wafer #1) for 30 sec at 750 rpm and then place the wafer into a Petri dish. Place the control layer mold into a Petri dish and pour 5:1 PDMS mixture onto the mold to a thickness of ~ 4 mm.
  4. Bake wafers/PDMS for 30 min at 70°C to partially cure the PDMS layers.
  5. After cooling down the wafers/PDMS to room temperature, cut the PDMS replica, which will form the control layer (wafer #2), from the Petri dish with a scalpel and peel it off the SU-8 mold. Hole punch an access port to the microchannel that will act as the on-chip membrane valve with a 21 gauge needle.
  6. Place the PDMS replica with the control layer onto the wafer #1 (which has the spin-coated PDMS fluidic layer). Carefully align and seal the control layer to the fluidic layer using a stereo microscope. Make sure to remove all air pockets between the layers and bake at 70°C overnight to fully cure both layers. This baking step will result in a monolithic PDMS slab with two layers.
  7. After cooling to room temperature, cut and peel the PDMS replica containing both the control and the fluidic layers off the SU-8 mold using a scalpel. Remove excess PDMS and separate each device unit with a razor blade. Hole punch access ports to the microchannels in the fluidic layer with a 21 gauge needle.
  8. Bond the PDMS slab to a coverslip to obtain a complete device. First, clean a coverslip (No: 1.5, 24 x 45 mm) with acetone and IPA. Next, treat both the coverslip and the PDMS replica surfaces with oxygen plasma under 500 mTorr for 30 sec, and immediately bring the two surfaces into contact to form an irreversible seal.
  9. Bake the devices overnight to increase bonding between the PDMS layers and the coverslip.

Steps 3-4 describe implementing the hydrodynamic trap using the microfluidic device described above.

3. Hydrodynamic Trap Experimental Setup

  1. Place the microfluidic device onto the stage of an inverted microscope and secure it with stage clips.
  2. Fill two gas-tight syringes separately with buffer and sample solutions and place them on a Harvard Apparatus Syringe pump (PHD 2000 Programmable). The buffer and the sample solutions are delivered to the microfluidic device through a 1 ml and a 250 μl syringe, respectively. Typically, a 50 mM Tris/Tris HCl buffer solution (pH 8.0) containing 0.02% v/v Triton X-100 is used as the buffer solution. The sample solution consists of a particle suspension (e.g. fluorescent polystyrene beads) in the buffer solution.
  3. Establish the fluidic connections between the syringes (delivering the sample and the buffer) and the microfluidic device. Connect the syringes to 1/16" outer diameter (OD) x 0.020" inner diameter (ID) perfluoroalkoxy (PFA) tubing using luer-lock adapters. Connect the other end of the PFA tubing to the inlet ports of the microfluidic device with 24 gauge metal tubing. A T-valve can be placed between the sample syringe and the sample port on the microfluidic device to control sample delivery.
  4. Establish the fluidic connections for the outlet channels in the microfluidic device. Connect the two outlet channels to PFA tubing (1/16" OD x 0.020" ID) by using 24 gauge metal tubing. The PFA tubing for the outlets should be of equal length. Submerge both outlet tubes into a 1.5 mL centrifuge tube filled with buffer solution, which serves to maintain a constant pressure drop between the syringes and the outlet channels.
  5. Fill the on-chip valve with fluorinated carrier oil using a 3 mL luer-lock plastic syringe to prevent air from leaking into the fluidic layer during operation. The air in the valve chamber is pushed through the PDMS membrane into the microchannel in the fluidic layer and later removed from the device with fluid flow through the outlet ports.
  6. Connect a pressurized inert gas (nitrogen) supply to the port in the control layer for on-chip valve operation. For this purpose, we use a nitrogen tank (2200 psi) and an electronic pressure regulator to supply 0-30 psi to the on-chip valve in the microfluidic device. The nitrogen tank is connected to the pressure regulator using ¼" OD x 0.170" ID tubing. The pressure regulator is connected to the microfluidic device through 1/16" OD x 0.020" ID PFA tubing with 24 gauge metal tubing at its terminus.
  7. Rinse the fluidic connections and the microfluidic device with 0.5 mL of buffer solution to ensure that all the air bubbles are removed from the system including the outlet channels. Typical flow rates used for clearing bubbles range between 2000-5000 μL/hr. After the air bubbles are rinsed out of the microfluidic channels, reduce the flow rate to 50-100 μL/hr, which is a typical volumetric flow rate for particle trapping.
  8. At this point, the fluidic connections are established, the sample and the buffer solutions are delivered to the microfluidic device at a fixed flow rate (50-100 μL/hr), and the device is ready for hydrodynamic trapping.

4. Hydrodynamic Trapping Procedure

  1. Run the custom-built LabVIEW code, which automates particle trapping (see Usage Note for LabVIEW code below).
  2. Using the microscope x-y translation stage, position the trapping region (cross-slot) at the center of the camera view. Bring the trapping region into focus of the objective lens and adjust the camera settings to optimize imaging conditions.
  3. Choose a rectangular region of interest (ROI) within the camera's field of view such that the center of the ROI will be the position of the trap center.
  4. Initialize the offset pressure applied to the on-chip valve. In one of the outlet channels, a 100-200 μm wide constriction is introduced to provide an offset pressure for the on-chip valve. The constant off-set pressure enables the on-chip valve to adjust the stagnation point position in the vicinity of the center of the channel cross-slot. For most experiments, the offset pressure is set between 0-12 psi depending on the channel dimensions (height and width), the constriction width, and the specifications of the on-chip valve (valve size, membrane thickness, etc.).
  5. Initiate the feedback controller and adjust the proportional gain to optimize trap response. The feedback controller will adjust the pressure applied to the on-chip valve in order to move the stagnation point position, which minimizes the error or the distance between the particle position and the set point (trap center). Depending on the flow rate and the on-chip valve position, there is an optimal proportional gain value, which increases trap stability and eliminates unwanted particle oscillations.
  6. Trap a particle. The LabVIEW code will automatically trap one of the particles entering the trapping region. Once a desired particle is trapped, it is possible to shut off the sample flow and isolate the trapped particle in buffer solution, if desired.
  7. Monitor the trapped particle and maintain particle focus within the image plane using manual focus or an automated focus microscope setup. It may be necessary to slightly adjust the proportional gain of the feedback controller in order to ensure trap stability during the course of a long time-scale trapping event (minutes to hours).

LabVIEW Code: Usage Note for Feedback Controller

Automated particle trapping is achieved using a linear feedback control algorithm implemented using a custom LabVIEW code. The LabVIEW code captures images from a CCD camera and transmits an electric potential (voltage) to a pressure regulator, which actively modulates the position (partially open/closed state) of an on-chip dynamic pneumatic valve. As the valve position changes, the hydrodynamic flow rate in one outlet line is adjusted, thereby re-positioning the stagnation point and enabling hydrodynamic trapping. The steps in the feedback loop are sequentially and iteratively executed at the rate of image capturing (10-60 Hz). The LabVIEW code executes the following steps during each feedback loop cycle:

  • Image capture. An image is acquired for a "target" particle in the trapping region of the microfluidic device using fluorescence microscopy with a 10x objective lens (NA: 0.4) and a CCD camera.
  • Particle tracking. Particle centroid position is determined, and the particle tracking algorithm is initiated. Particles are localized by fitting the emission intensity profile of the particle to a point spread function (PSF), from which the centroid position is determined.
  • Flow field control. The updated pressure intended for the on-chip dynamic valve is calculated using a feedback control algorithm with a proportional controller. In this way, the action of the valve is to re-position the stagnation point, which exerts a hydrodynamic force on the particle in order to steer the particle toward the trap center.

The LabVIEW code records the following data for every image captured during particle trapping: 1) time elapsed, 2) centroid (x,y) position of the trapped particle, 3) position of the trap center, 4) distance of the particle from the trap center, 5) pressure applied to the on-chip valve. In addition, the code also records a movie of the trapped particle in AVI file format.

5. Representative Results

We trapped fluorescent polystyrene beads of various size (100, 540, 830 nm, and 2.2 μm diameter) using a hydrodynamic trap. Figure 1(a) shows an image of a particle trapped at the cross-slot junction in a microfluidic device. The trajectory of a trapped particle may be determined directly from the centroid position data recorded by the LabVIEW code during a trapping event or by tracking and localizing the trapped particle from the recorded movie file. Figure 1(b) shows the trajectory of a trapped particle (2.2 μm fluorescent polystyrene bead) along the outlet channel direction. The bead is initially trapped (squares) for 3 min and is then released from the trap and escapes along one of the outlet channels (circles). Particle trajectories along the compressional flow axis (inlet channel direction; data not shown) are similar to particle trajectories along the extensional flow axis (outflow direction) as shown in Figure 1(b). A histogram of particle displacement from the trap center for a trapped bead (2.2 μm diameter) along the outlet channel directions is shown in Figure 1(c). Using the feedback control algorithm described in this work, trapped particles are confined to within ±1 μm of the trap center along the inlet and outlet channel directions.

A schematic of the microfluidic device used for hydrodynamic trapping is shown in Figure 2. The integrated microfluidic device consists of a fluidic layer and a control layer and is fabricated using standard multilayer soft lithography as described in this article. The fluidic layer contains the buffer and sample channels, as well as the cross-slot channel geometry to facilitate hydrodynamic trapping. The control layer consists of a pneumatic valve positioned above one of the outlet channels in the fluidic layer, and the control and fluidic layers are separated by a thin elastomeric membrane. During device operation, the valve in the control layer is pressurized with nitrogen gas, which forces the thin membrane into the fluidic layer, thereby inducing a constriction in the outlet channel. The dynamic pneumatic valve constricts the outlet channel by variable amounts by changing the pressure applied to the control layer, which adjusts the relative flow rates in the outlet channels and enables fine-scale control of the stagnation point.

Figure 1
Figure 1: Particle Trapping. (a) Image of a single bead confined in the hydrodynamic trap. In addition to the bead at the trap center, several untrapped beads are shown in the trapping region. (b) Trajectory of a trapped particle along the outlet channels (squares). When the particle is released from the trap (arrow), it escapes along one of the outlet channels (circles). (c) Histogram of displacements of a trapped bead (2.2 μm diameter) from the trap center along the outlet channels.

Figure 2
Figure 2: Schematic of microfluidic device for hydrodynamic trapping. The hydrodynamic trap is constructed using a two-layer microfluidic device. The fluidic layer consists of a sample inlet, four buffer inlets, and two waste outlets. The control layer consists of a pneumatic membrane valve situated on top of one of the outlet channels in the fluidic layer. A constriction in the opposing outlet channel provides an offset pressure for the pneumatic valve. Typical channel dimensions range between 100-500 μm. In region (A), sample inlet is flow focused by two buffer inlets. In region (B), opposing inlet streams converge at the cross-slot junction where trapping occurs. The pneumatic valve (C) is positioned on top of one of the outlet channels. The stagnation point position is modulated by regulating pressure to this valve.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

Current microfluidic methods for particle manipulation based on hydrodynamic flow can be characterized as contact-based or non-contact methods. Contact-based methods use fluid flow to physically confine and immobilize particles against microfabricated channel walls 9, whereas non-contact methods rely on circulating flow or microeddies 10. In this work, we present a method for free-solution particle trapping using the sole action of fluid flow. The hydrodynamic trap enables confinement and manipulation of small particles at a fluid stagnation point in a microfluidic cross-slot device. In this device, an automated feedback control mechanism is used to confine particles by fine-scale and active adjustment of the stagnation point position in a flowing fluid.

What is the tightness of confinement for particles in the hydrodynamic trap and how can this be optimized? The accuracy of confining a particle to the trap center depends on the precision of centroid determination when localizing particle position. To achieve robust particle trapping, the user should ensure maximum image contrast between the particle and the background for optimal tracking and localization. In addition, special care should be taken to avoid bubbles or debris in the microchannels, which may affect particle tracking. A stable flow source should be used to minimize perturbations in fluid flow, as the stability of the stagnation point position is sensitive to flow fluctuations. Using this approach, hydrodynamic trap stiffness was measured to be ~1E-4 pN/nm for ~2 μm particles1, which is comparable to alternative methods including electrokinetic traps or optical tweezers. Micron-scale particles are confined to within 1 μm of the trap center for extended periods of time, which allows for precise positioning and manipulation of particles in free-solution. With further technology development, trapped particles may be transiently exposed to variable microenvironments when coupling the hydrodynamic trap with chemical gradients generated using laminar flow in microchannels. Finally, hydrodynamic trapping occurs at a stagnation point, where fluid convection tends to zero. In an ideal trap, particles are confined at a location of zero fluid velocity where particle motion is largely dominated by Brownian motion. From this perspective, the hydrodynamic trap is a non-pertubative method of trapping based on continuous fluid flow.

Hydrodynamic trapping and manipulation is readily achieved for any arbitrary "target" particle, given that the particle can be imaged, tracked, and localized using optical microscopy. Therefore, fluorescent and non-fluorescent particles and non-isotropic objects can be trapped without regard to the chemical/physical/optical nature of the trapped particle. In addition, the hydrodynamic trap can be easily integrated into existing soft lithography-based microfluidic systems without the need for complicated fabrication, patterning of electrodes or extensive optical setups. The hydrodynamic trap is a low-cost and user-friendly tool for particle trapping with minimal laboratory equipment requirements, including a microfluidic device, a pressure regulator, and a computer-based feedback controller. Overall, the hydrodynamic trap has the potential to transform fundamental and applied science studies of micro- and nanoscale particles.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

No conflicts of interest declared.

Acknowledgments

We thank the Kenis group at the University of Illinois at Urbana-Champaign for helpful discussions and generously providing use of cleanroom facilities.

This work was funded by an NIH Pathway to Independence PI Award, under Grant No. 4R00HG004183-03 (Charles M. Schroeder and Melikhan Tanyeri).

This work was supported by the National Science Foundation through a Graduate Research Fellowship to Eric M. Johnson-Chavarria.

Materials

Name Company Catalog Number Comments
21 gauge blunt needle Zephyrtronics ZT-5-021-1-L For punching port holes in PDMS
3 ml plastic syringe BD Biosciences 309585 For filling valve with oil
Si wafers University Wafer 3” P(100) single side polished 380 μm test grade
Cover glass VWR international 48404-428 24 x 40 mm #1.5
DAQ card National Instruments PCI 6229
Fluorescent beads Spherotech, Inc. FP-2056-2 2.2 μm Nile red
Fluorinert 3M FC 40 Fluorinated carrier oil
Inverted Microscope Olympus Corporation IX-71
LabVIEW National Instruments Version 9.0f3 (32bit)
Stereo Microscope Leica Microsystems MZ6 For aligning PDMS control layer to fluidic layer.
Mechanical Convection Oven VWR international 1300U For baking devices to create monolithic PDMS slabs with two layers.
Microfluidic tubing and connectors Upchurch Scientific 1/16 x .020 PFA tubing and super flangeless fittings
PDMS GE Healthcare RTV 615 A&B
Plasma Chamber Harrick Scientific Products, Inc. PDC-001
Pressure Transducer Proportion Air DQPV1
Spin Coater Specialty Coating Systems G3P-8 Spin Coat
Photoresist MicroChem Corp. SU 8 2050
Syringe Pump Harvard Apparatus PHD 2000 Programmable
Terminal Block National Instruments BNC 2110 For analog output to pressure regulator and read out.
UV Collimated Light Source and Exposure System OAI Model 30 Enhanced Light Source

DOWNLOAD MATERIALS LIST

References

  1. Tanyeri, M., Johnson-Chavarria, E. M., Schroeder, C. M. Hydrodynamic Trap for Single Particles and Cells. Applied Physics Letters. 96, 224101-224101 (2010).
  2. Ashkin, A., Dziedzic, J. M., Bjorkholm, J. E., Chu, S. Observation of a Single-Beam Gradient Force Optical Trap for Dielectric Particles. Optics Letters. 11, 288-290 (1986).
  3. Neuman, K. C., Block, S. M. Optical trapping. Review of Scientific Instruments. 75, 2787-2809 (2004).
  4. Gosse, C., Croquette, V. Magnetic tweezers: Micromanipulation and force measurement at the molecular level. Biophysical Journal. 82, 3314-3329 (2002).
  5. Chiou, P. Y., Ohta, A. T., Wu, M. C., C, M. Massively parallel manipulation of single cells and microparticles using optical images. Nature. 436, 370-372 (2005).
  6. Cohen, A. E., Moerner, W. E. Method for trapping and manipulating nanoscale objects in solution. Applied Physics Letters. 86, 093109-09 (2005).
  7. Evander, M. Noninvasive acoustic cell trapping in a microfluidic perfusion system for online bioassays", Analytical Chemistry 79. , 2984-2991 (2007).
  8. Unger, M. A., Chou, H. P., Thorsen, T., Scherer, A., Quake, S. R. Monolithic microfabricated valves and pumps by multilayer soft lithography. Science. 288, 113-116 (2000).
  9. Kim, M. C., Wang, Z. H., Lam, R. H. W., Thorsen, T. Building a better cell trap: Applying Lagrangian modeling to the design of microfluidic devices for cell biology. Journal of Applied Physics. 103, (2008).
  10. Lutz, B. R., Chen, J., Schwartz, D. T. Hydrodynamic tweezers: 1. Noncontact trapping of single cells using steady streaming microeddies. Analytical Chemistry. 78, 5429-5435 (2006).

Tags

Microfluidic-based Hydrodynamic Trap Single Particles Particle Trapping Manipulation Fluid Flow Microfluidic Device Cross-slot Channel Geometry Laminar Streams Planar Extensional Flow Stagnation Point Feedback Control Algorithm LabVIEW Code Pressure Regulation
A Microfluidic-based Hydrodynamic Trap for Single Particles
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Johnson-Chavarria, E. M., Tanyeri,More

Johnson-Chavarria, E. M., Tanyeri, M., Schroeder, C. M. A Microfluidic-based Hydrodynamic Trap for Single Particles. J. Vis. Exp. (47), e2517, doi:10.3791/2517 (2011).

Less