Counting and Determining the Viability of Cultured Cells
Richard Ricardo, Katy Phelan Molecular Pathology Laboratory Network, Inc
You must be subscribed to access this content. Please click here to subscribe or get a FREE trial.
|
0:00 Title0 0:21 Subscription Lock21 0:36 Introduction36 0:57 Preparing the hemacytometer57 1:31 Preparing cell suspension91 2:08 Loading hemacytometer128 2:42 Counting and calculating number of cells162 4:24 Determination of cell viability264 5:25 Conclusion325
|
|
Determining the number of cells in culture is important in standardization of culture conditions and in performing accurate quantitation experiments. A hemacytometer is a thick glass slide with a central area designed as a counting chamber. Cell suspension is applied to a defined area and counted so cell density can be calculated.
|
|
You need a subscription to be able to access files |
|
|
You must be subscribed to access this content. Please click here to subscribe or get a FREE trial.
|
|
Ricardo R, Phelan K (2008). Counting and Determining the Viability of Cultured Cells. JoVE. 16. http://www.jove.com/index/details.stp?id=752, doi: 10.3791/752
|
 |
 |
|  |
 |
Allowed tags: i, b, u, sup, sub 
|
|
 |
| | ![]() |
|
 |
| |  |
|
|  |
| | 06/29/2008 4:32:43 PM
|
|
|  |
gg responded with a statement of type: Neutral
|
|
| |
| |  |
|
| | | | |
| | | | | |
I think it would be helpful if you provided links to related protocols (such as for passaging cells, as referenced in the protocol). For example, have a "Related protocols" section with direct links to them.
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
|  |
| | 08/02/2008 11:55:32 PM
|
|
|  |
Anonymous responded with a statement of type: Neutral
|
|
| |
| |  |
|
| | | | |
| | | | | |
If I take 200microliter.spin it down and resuspended in 20 microliter then should I divide result by 10 ? Is that right?
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
If you are talking about the concentration of cells in suspension, then no, you multiply by ten. When you decrease the volume by a factor of ten, then you are concentrating cells and should increase the concentration by a factor of ten.
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
|  |
| | 08/03/2008 5:32:57 PM
|
|
|  |
Anonymous responded with a statement of type: Neutral
|
|
| |
| |  |
|
| | | | |
| | | | | |
I thought it's the opposite.Let's say if you dilute the cells 10x won't you multiply by 10 and it should be the oppsite in concentrating?
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
|  |
| | 08/03/2008 5:33:11 PM
|
|
|  |
Anonymous responded with a statement of type: Neutral
|
|
| |
| |  |
|
| | | | |
| | | | | |
I thought it's the opposite.Let's say if you dilute the cells 10x won't you multiply by 10 and it should be the oppsite in concentrating?
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
So let's make sure we are talking about the concentration of cells, which is what you get after counting them. concentration of your cells is defined as equal to the # cells/ unit volume.
so, if you spin down your cells, which were at 200 microliters, and then resuspend in 20 (thereby decreasing the volume by a factor of 10) you can multiply the concentration by a factor of 10 (because they are inversely proportional as the definition states). Or if you feel like being fancy multiply 200 microliters, which is in the denominator by 0.1 (because 20 is a 10th of 200).
when you dilute cells you are doing the opposite, increasing the volume that contains a fixed # of cells will lower the concentration, because in your example you are multipling the volume (in the denominator) by a factor of 10. divide the concentration by 10.
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
|  |
| | 10/21/2008 12:12:33 PM
|
|
|  |
C Kolar responded with a statement of type: Disagree
|
|
| |
| |  |
|
| | | | |
| | | | | |
Overflowing the center posts when you are loading the cell suspension into the hemacytometer results in liquid flowing over and cells staying behind, so that the count is artificially raised. The cell suspension loaded onto the stage area (usually about 10-15 uL) should only be allowed to flow (in one swoop, not in jerks as shown) to cover the counting surface, then stopped so that liquid does not overflow into the side wells. This technique as shown is less than desirable.
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
|  |
| | 03/09/2009 9:52:27 AM
|
|
|  |
kavitha responded with a statement of type: Neutral
|
|
| |
| |  |
|
| | | | |
| | | | | |
maintenance of THP-1 monocyte cell line and can they be used for anti-inflammatory studies
|
|
|  |
| | |
| | | ![]() |
|
 |
| |  |
|
I agree with C. Kolar in his/her comment about not letting the liquid overflow; although it requires some manual sensitivity for not pushing too hard the micropipette plunger, it is something everybody should learn how to do.
I would like to make some suggestions to improve the work. First, I would invite authors to consider and discuss subjectivity in this type of cell counting; sometimes it is very subjective to decide how blue a cell should be to be considered dead;. In second place, letting the cells 5 minutes into the trypan blue solution might lead to cell death (and therefore decreased cell viability) if not properly stored; so this time should be standardize for every cell type.
The authors also should consider explaining the remaining cell counting area, not just the central part. Some people consider the remaining 4 zones of 1 square mm for cell counting and use all of them to get an average instead of one single value from the central zone.
When calculating the cell concentration, the authors do not explain why the "10.000" appears in the formula (just the product of 1 square millimeter (area) and 0.1 mm (height of the chamber)), and what is considered as dilution...I believe an example would be very helpful to explain this aspect and is missing from the presentation.
| |  |
|
|
|
|
| |
|
| | |
|
|
|  |
| | |
| | | ![]() |
|
|
Mailer.URL: http://www.jove.com/index/details.stp?id=752 |
|
You must be logged in to track comments. |
|
My Stuff (Please log in to use) |
|
|
|