Here, we present a protocol for the detection and quantification of Plasmodium falciparum in infected aqueous red blood cells using an attenuated total reflection infrared spectrometer and multivariate data analysis.
A subscription to JoVE is required to view this content. Sign in or start your free trial.
Method Article
Here, we present a protocol for the detection and quantification of Plasmodium falciparum in infected aqueous red blood cells using an attenuated total reflection infrared spectrometer and multivariate data analysis.
We demonstrate a method of quantification and detection of parasites in aqueous red blood cells (RBCs) by using a simple benchtop Attenuated Total Reflection Fourier Transform Infrared (ATR-FTIR) spectrometer in conjunction with Multivariate Data Analysis (MVDA). 3D7 P. falciparum were cultured to 10% parasitemia ring stage parasites and used to spike fresh donor isolated RBCs to create a dilution series between 0–1%. 10 µL of each sample were placed onto the center of the ATR diamond window to acquire the spectrum. The sample data was treated to improve the signal to noise ratio and to remove the contribution of water, and then the second derivative was applied to resolve spectral features. The data were then analyzed using two types of MVDA: first Principal Component Analysis (PCA) to determine any outliers and then Partial Least Squares Regression (PLS-R) to build the quantification model.
Malaria is among the most devastating diseases of our time; over half the population lives at risk in endemic regions and it disproportionately burdens the poor1,2,3,4. A large part of the issue is the asymptomatic carriers and early stage patients that act as reservoirs for mosquito vectors5, causing spikes of infection during wet seasons and allowing it to persist in communities. Malaria is caused by five Plasmodium parasites, the most deadly of which is P. falciparum which causes the most severe f....
Access restricted. Please log in or start a trial to view this content.
Please consult appropriate Material Safety Data Sheets (MSDS) and seek appropriate Biosafety Level 2 (BSL-2) training. All culturing steps must be done in a BSL-2 cabinet using aseptic technique, meaning there is a risk of exposure to harmful UV radiation from decontamination steps, needle stick injuries, and potential biological exposure and infection if the parasite culture enters any injuries. Furthermore, stock blood from blood banks is only screened for certain diseases and the potential for spreading blood borne diseases is a potential risk. Seek immediate medical aid in the occurrence of injury.
1. Preparation and Measurement of 3D7
Access restricted. Please log in or start a trial to view this content.
Partial Least Squares (PLS-R) plot and its associated regression vector, Figure 1a and 1b respectively, show that the signal from parasites are distinct enough from the RBCs that they can be used to form a linear regression model to be used for the prediction of parasites in future data sets.
A robust, linear PLS-R model was generated with an R-squared value of 0.87 and a root mean .......
Access restricted. Please log in or start a trial to view this content.
PLS-R model is a supervised multivariate method that finds a linear relationship, Y=bX+E, between the predictive variables X (here, the absorbance at each wavenumber) and a continuous variable Y (here, the parasitemia level). In short, the model combines the variables in X to create a new set of latent variables (LVs) that capture the variance on X correlated with Y, and computes a regression vector (b) that multiplied by a new spectrum results in the estimation of the y value. The strength of the model can be taken from.......
Access restricted. Please log in or start a trial to view this content.
The authors have nothing to disclose.
Funding to the authors was provided by the Australian Research Council (Future Fellowship FT120100926 to BRW), National Health and Medical Research Council of Australia (Program grant and Senior Research Fellowship to JGB; Early Career Fellowships JSR; Infrastructure for Research Institutes Support Scheme Grant to the Burnet Institute), and the Victorian State Government Operational Infrastructure Support Grant to the Burnet Institute. We acknowledge Mr. Finlay Shanks for instrumental support.
....Access restricted. Please log in or start a trial to view this content.
| Name | Company | Catalog Number | Comments |
|---|---|---|---|
| Donor blood | - | - | Must be collected by a trained medical practitioner |
| Stock blood | Australian Red Cross` | - | |
| 3D7 Plasmodium falciparum | The Burnet Institute | - | Laboratory strain wild type equivalent |
| RPMI-1640 media with L-hepes without Sodium Bicarbonate | Sigma Aldrich | R6504 | |
| Albumax (Gibco) | Thermofisher | E003000PJ | Aliquot into 50 mL falcon tubes and store in freezer |
| Sodium Bicarbonate | Sigma Aldrich | S5761 | |
| Giemsa Stain | Sigma Aldrich | 48900 | |
| Sorbitol | Sigma Aldrich | S1876 | Must be filtered before use in culture |
| Blood collection tubes (Vacutainers) | Becton, Dickonson and Company | 367671 | |
| Immersion Oil | Thermofisher | M3004 | |
| 50 mL culture dishes | Falcon | 353025 | |
| 25mL pipette tips | Falcon | 357515 | |
| 10mL pipette tips | Falcon | 357530 | |
| Microscope Slides | Sigma Aldrich | S8902 | |
| Centrifuge tubes 50 mL | Sigma Aldrich | T2318 | |
| Automated pipette controller | Integra-biosciences | 155 015 | |
| Sorvall Legend X1R Centrifuge | Thermofisher | 75004260 | |
| Forma™ 310 Direct Heat CO2 Incubators | Thermofisher | 310TS | |
| Nikon Eclipse E-100 Binocular Microscope | Nikon Instruments | E100_2CE-MRTK-1 | When purchasing ensure that the 100x lens is an oil immersion lens |
| Bruker Alpha Ft-IR Spectrometer with ATR Quick Snap Attachment | Bruker | 9308-3700 | Be sure to request "Eco-ATR" attachment when purchasing |
| Matlab | Mathworks Inc | Multivariate data analysis software | |
| The Unscrambler X | CAMO | Multivariate data analysis software | |
| PLS-Toolbox | Mathworks, Inc. | GUI for Matlab |