JoVE   
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Biology

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Neuroscience

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Immunology and Infection

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Clinical and Translational Medicine

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Bioengineering

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Applied Physics

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Chemistry

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Behavior

  
You do not have subscription access to articles in this section. Learn more about access.

  JoVE Environment

|   

JoVE Science Education

General Laboratory Techniques

You do not have subscription access to videos in this collection. Learn more about access.

Basic Methods in Cellular and Molecular Biology

You do not have subscription access to videos in this collection. Learn more about access.

Model Organisms I

You do not have subscription access to videos in this collection. Learn more about access.

Model Organisms II

You do not have subscription access to videos in this collection. Learn more about access.

Essentials of
Neuroscience

You do not have subscription access to videos in this collection. Learn more about access.

Essentials of Developmental Biology

You have subscription access to videos in this collection through your user account.

Simplified LC/MS/MS Bioanalytical Method Development with Radar Technology - ADVERTISEMENT

Robert S. Plumb1

1Pharmaceutical Business Operations, Waters Corporation

Article
    Downloads
    Comments Metrics
     

    Sign in to register to receive emails with product information.

    Abstract

    The development of a bioanalytical assay requires the development of a MRM method as well as a chromatography method, which requires positioning of the analyte peak of interest away from the endogenous background signal. The Xevo™ TQ-S tandem quadrupole mass spectrometer with RADAR™ Technology allows the simultaneous collection of full scan and MS/MS MRM data, vastly simplifying the method development process.

    For a bioanalysis method to be effective it must be transferable, reproducible, and robust. The development of a reliable LC/MS/MS bioanalysis assay involves the thorough optimization of sample preparation, MS detection, and chromatography conditions. This can often be a time-consuming process, requiring the analyte peaks(s) be resolved from endogenous matrix components that cause ion suppression and assay irreproducibility. This often requires multiple analytical runs to acquire the necessary MRM, product ion, and full scan data.

    Described here is the use of Xevo™ TQ-S Mass Spectrometer with RADAR™ Technology - a novel dual-scan data collection capability to simplify bioanalysis method development. The Xevo TQ-S employs a novel collision cell design which allows for the simultaneous collection of full scan MS and MRM data.

    Protocol

    I.Simplified LC/MS/MS Bioanalytical Method Development With RADAR Technology

    1. To demonstrate simplified LC/MS/MS bioanalytical method development with RADAR technology, chromatography was performed on a Waters ACQUITY UPLC® System, using an ACQUITY UPLC Bridged Ethyl Hybrid C18 column.

      LC Conditions:
      ACQUITY UPLC BEH C18 column 2.1 x 50 mm, 1.7 μm
      Mobile Phase:
      1. 0.1% formic acid or 0.1% aqueous ammonium hydroxide
      2. Methanol or acetonitrile

      Gradient: 5-95% organic over 0-2 minutes at a flow rate of 600 μL/min.
    2. The analytes were eluted at a flow rate of 600 μL/min with a 5 to 95% organic gradient over 2 minutes, where 0.1% aqueous formic acid or 0.1% aqueous ammonium hydroxide was exchanged for methanol or acetonitrile.
    3. Mass spectrometry detection was performed on a Waters Xevo™ TQ-S Mass Spectrometer operated in a positive ion electrospray MRM mode, with simultaneous acquisition of full scan data.

      MS Conditions:
      Positive ion electrospray:
           MRM:     315 > 129
           Full scan MS:     50-500 m/z at 5000 amu/sec.
           Product ion scan: Precursors of m/z = 184 from 200-600
    4. To illustrate the use of the RADAR technology, the common H2 receptor antagonist ranitidine hydrochloride was spiked into rat plasma, precipitated with acetonitrile (2:1), and analyzed using a UPLC/MS/MS system.

    II.Representative Results Bioanalysis Using RADAR™ Technology

    Chromatograms obtained using a conventional acidic aqueous buffer and acetonitrile organic modifier gradient were compared with chromatograms obtained using a basic aqueous buffer (Fig. 1). In the LC/MS/MS analysis of ranitidine hydrochloride in rat plasma, the RADAR MRM data shows that with an acidic aqueous modifier, ranitidine is unretained, eluting with the void of the chromatographic system. However, when a basic aqueous modifier is employed, the compound is retained, eluting at 0.9 min. The full scan and parents of m/z = 184 data show that with both the acidic and basic separation, the analyte is resolved from phospholipid and other endogenous materials in the matrix.

    The use of a methanol organic modifier with a basic aqueous buffer increased the analyte retention to 1.30 min. The analyte signal response was also increased by a factor of 4 (Fig. 2). However, with a simple 5-95% basic-methanol gradient, the full scan data revealed that the analyte peak co-eluted with the endogenous material in the sample. This co-elution could cause ion suppression and assay irreproducibility.

    To resolve the analyte peak from the endogenous material, the gradient steepness was adjusted. The Xevo TQ-S RADAR functionality was used to quickly select the best LC conditions with the best throughput. The final chromatographic conditions were a gradient of 15-60% methanol over 2 minutes with a 95% organic wash from 2-2.5 minutes (Fig. 3). The analyte peak is well resolved from the endogenous material in the sample as seen from the full scan trace, in green. This approach can also be used during sample analysis to check for drug-related metabolites, co-administered therapies, and variations in matrix which could affect the veracity of the results.

    Figure 1
    Figure 1: Chromatograms obtained using a conventional acidic aqueous buffer and acetonitrile organic modifier gradient compared with chromatograms obtained using a basic aqueous buffer. With an acidic aqueous modifier, ranitidine is unretained, eluting with the void. When a basic aqueous modifier is employed, the compound is retained, eluting at 0.9 min. The full scan and parents of m/z = 184 data show that with both the acidic and basic separation, the analyte is resolved from phospholipid and other endogenous materials in the matrix.

    Figure 2
    Figure 2: The use of a methanol organic modifier with a basic aqueous buffer increased the ranitidine hydrochloride retention to 1.30 min. The analyte signal response was also increased by a factor of 4. However, the full scan data reveals the analyte peak co-eluted with the endogenous material in the sample when a simple 5-95% basic-methanol gradient is employed.

    Figure 3
    Figure 3: The final chromatographic conditions for ranitidine hydrochloride included a gradient of 15-60% methanol over 2 minutes with a 95% organic wash from 2-2.5 minutes. The analyte peak is well resolved from the endogenous material in the sample as seen from the full scan trace, in green.

    Discussion

    Bioanalytical method development is significantly simplified using Xevo TQ-S RADAR technology. The ability to simultaneously acquire full scan data as well as MS/MS and MRM data allows the endogenous sample matrix to be monitored at the same time as the analyte peak, facilitating:

    • Faster method development
    • Elimination of the need for repeat injections to obtain background signal
    • Matrix monitoring during sample analysis
    • Easy troubleshooting
    • Detection of in vivo metabolites during the pre-clinical and clinical development

    Disclosures

    Acknowledgements

    Paul Rainville, Waters Corporation, Milford, MA USA
    David Little, Waters Corporation, Manchester, UK
    Gary Harland, Waters Corporation, Manchester, UK

    Materials

    Name Company Catalog Number Comments
    Xevo TQ-S Mass Spectrometer Waters More information available at http://www.waters.com Mass Spectrometer
    ACQUITY UPLC System Waters More information available at http://www.waters.com Chromatography System
    ACQUITY UPLC BEH C18, 2.1 x 50 mm, 1.7 μm Column Waters More information available at http://www.waters.com Chromatography Column