10.2
电化学测量是在由各种控制和测量电流和电位的组件构成的电化学电池中进行的。其中一个基本组件是电极,即能够在其表面进行电子转移反应的导电材料。
电化学电池中有两种主要类型的电极。第一种被称为工作电极或指示电极,其电位对分析物的浓度敏感,并对电活性物质的活性变化作出响应。常见的指示电极包括惰性铂电极、对…
通常使用参比电极与工作电极或指示电极的组合来测量双电极电池中的电位和电流。
在电位测定中,指示电极通过氧化还原反应,其响应与被测分析物的浓度成正比。
常见的指示电极类型包括惰性金属电极、可逆反应金属电极以及离子选择性膜电极。
参比电极(如饱和甘汞电极和银-氯化银电极)在给定温度下具有恒定的电位,不受被测物组成变化的影响。当流经参比电极的电流较小时,其性能最佳,可为指示电极提供稳定的参考电位。
相对于参比电极提供的固定参考电位,测量指示电极的相对电位。
然而,对于需要电流通过的方法,可以添加第三个电极——称为辅助电极或对电极。这些辅助电极允许电流通过电解池,以完成电路。
View the full transcript and gain access to JoVE Core videos
Q1: What is the role of a reference electrode in electrochemical measurements?
A reference electrode provides a constant potential at a given temperature, independent of analyte composition changes. It establishes a fixed reference point against which the working electrode's potential is measured. Reference electrodes like the saturated calomel electrode and silver-silver chloride electrode work best when minimal current flows through them, ensuring their potential remains stable during potentiometric analysis.
Q2: How do indicator electrodes respond to analyte concentration?
Indicator electrodes respond proportionally to analyte concentration through redox reactions occurring at their surfaces. Common types include inert metal electrodes like platinum, reversibly reactive metal electrodes like silver, and ion-selective membrane electrodes. The potential change of the indicator electrode reflects the activity of electroactive species in solution, enabling concentration determination in potentiometric tests.
Q3: Why is a third electrode needed in some electrochemical methods?
A third electrode, called an auxiliary or counter electrode, is added when current flow is required through the electrochemical cell. While reference and working electrodes measure potential, the auxiliary electrode completes the electrical circuit and facilitates current passage. This enables dynamic methods where current alters the concentration of species in the cell, such as controlled potential coulometry electrolytic methods.
Q4: What distinguishes inert metal electrodes from reactive metal electrodes?
Inert metal electrodes like platinum do not participate in redox reactions; they only facilitate electron transfer. Reversibly reactive metal electrodes like silver can participate in redox reactions and respond to specific ions such as Ag+, halides, and other species that react with the electrode material. This reactivity makes them suitable as indicator electrodes for detecting particular analytes.
Q5: How do ion-selective membrane electrodes function in potentiometric analysis?
Ion-selective membrane electrodes contain specialized membranes that selectively respond to specific ions in solution. These electrodes generate a potential proportional to the concentration of target ions, enabling selective analyte detection. They represent an important class of indicator electrodes used in potentiometric measurements where specificity for particular ionic species is required.
Q6: What happens to reference electrode potential when current flows through it?
Reference electrodes are designed to maintain constant potential only when minimal current flows through them. If significant current passes through a reference electrode, its potential becomes unstable and unreliable. This is why reference electrodes work best in potentiometric methods with low current, and why auxiliary electrodes are used in high-current applications to protect the reference electrode's stability.
Q7: How are working and reference electrodes used together in a two-electrode cell?
In a two-electrode cell, the working electrode's potential is measured relative to the reference electrode's fixed potential. The working electrode responds to analyte concentration changes through redox reactions, while the reference electrode provides the stable baseline for comparison. This configuration allows potentiometric tests to quantify analyte concentration by monitoring the potential difference between the two electrodes.