14.13
Plasma emission spectroscopy uses three basic instrument types.
Sequential plasma emission spectrometers measure one emission line at a time, whereas multichannel and Fourier-transform instruments simultaneously measure multiple emission lines.
Sequential spectrometers typically use a grating monochromator, where the grating is rotated to focus different wavelengths on the exit slit.
For example, in slew scan spectrometers, the monochromator rapidly rotates to a wavelength close to one emission line and then rotates in small steps across that line.
In contrast, scanning echelle spectrometers have a moving photomultiplier tube behind an aperture plate, with the echelle grating scattering the wavelengths in a broad array that can be detected one slit at a time.
Some multichannel instruments use a grating polychromator to measure multiple emission lines simultaneously with photomultiplier tubes positioned behind fixed exit slits.
Other multichannel instruments use spectrographs for simultaneous measurements, employing two-dimensional charge-injection devices or charge-coupled devices as transducers.
Although Fourier transform spectrometers offer wide wavelength coverage, they are not
Inductief gekoppeld plasma (ICP) is de gebruikelijke plasmabron die wordt gebruikt in atomaire-emissiespectroscopie (AES), een techniek die verschille…
Copyright © 2026 MyJoVE Corporation. Alle rechten voorbehouden.