Methodenartikel

Peptide Purification: An RP-HPLC-based Technique to Extract Peptides from Digested Protein Lysates

4.9K weergaven

30 april 2023

In dit artikel

Samenvatting

Source: Cheng, L. C. et al. Phosphopeptide Enrichment Coupled with Label-free Quantitative Mass Spectrometry to Investigate the Phosphoproteome in Prostate Cancer. J. Vis. Exp. (2018)

In this video, we demonstrate reversed-phase high-performance liquid chromatography to purify peptides from a digested protein lysate mixture. These purified peptides are then lyophilized and stored for downstream applications. Analysis of peptides can help understand the mechanics involved in cellular signaling and cancer.

Protocol

1. Reverse Phase Extraction

  1. Start with a pre-digested protein lysate in a glass vial. Filter the sample by using a 15 mL 10 kDa cutoff filter. Centrifuge the sample at 3,500 x g using the swing bucket rotor (or 3,500 x g in a fixed angle rotor) at 15 °C until the retentate volume is less than 250 µL (this takes approximately 45 - 60 min). Collect the flow-through and discard the retentate.
    NOTE: The experiment can be paused here. Freeze the samples at -80 °C for later use.
  2. To acidify the sample, add approximately 20 µL of 5% trifluoroacetic acid (TFA) per mL of lysate. Mix them well and measure the sample pH by using pH strips. Adjust pH to 2.5 using 5% TFA.
  3. Connect the shorter end of a C-18 column to a vacuum manifold. Set the vacuum between 17 and 34 kPa (or according to the manufacturer’s instructions). Using glass pipettes, wet the column with 3 mL of 100% acetonitrile (ACN). Do not let the column dry.
  4. Using glass pipettes, equilibrate the column with 6 mL of 0.1% TFA applied as 2x 3 mL. Load the acidified sample into the column. Do not add more than 3 mL at a time. Adjust the vacuum to target about 1 to 2 drops per second.
  5. Using glass pipettes, wash the column with 9 mL of 0.1% TFA applied as 3x 3 mL. Elute the column with 2 mL of 40% ACN, 0.1% TFA. Collect two 2 mL fractions into glass culture tubes. Discard the column.
  6. Cover the eluate tubes with parafilm and punch 3-5 holes on the cover using a 20G needle. Freeze the eluate on dry ice for at least 30 min until it completely solidifies.
  7. Lyophilize the fractions overnight. On the following day, make sure the samples are completely dry before stopping the lyophilizer. Store the tubes in a 50 mL conical tube with delicate wipes at -80 °C.

Toegang beperkt. Log in of start een proefperiode om deze inhoud te bekijken.

Openbaarmakingen

No conflicts of interest declared.

Materialen

Lijst van materialen gebruikt in dit artikel
NaamBedrijfCatalogusnummerOpmerkingen
Ultra-Low Temperature FreezerPanasonicMDF-U76V
Freezer -20 °CVWRscpmf-2020
Swing rotor bucketThermoFisher Scientific75004377
Vacuum manifoldRestek26080
LyophilizerLabconco7420020
CentriVap Benchtop Vacuum ConcentratorLabconco7810010
Amicon Ultra-15 Centrifugal Filter UnitsMillipore SigmaUFC901024
End-over-end rotatorThermoFisher Scientific415110Q
Glass culture tubesFisher Scientific14-961-26
ParafilmFisher Scientific13-374-12
20G needleBDB305175
KimwipesFisher Scientific06-666A
Screw cap cryotubeThermoFisher Scientific379189
Nunc 15 mL conical tubesThermoFisher Scientific12-565-268
Low protein-binding Eppendorf tubesEppendorf22431081
3 mL syringeBD309657
Trifluoracetic Acid (TFA)Fisher ScientificPI-28904
Acetonitrile (ACN)Fisher ScientificA21-1
MilliQ waterDeionized water used to prepare all solutions and bufferes