Three-Chambered Array-Based Impedance Assay: A Real-Time Analysis Technique to Assess Invasive Potential of Cancer Cells by Measuring Electrical Impedance

July 8th, 2025

In This Article

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Source: Sharif, G. M. et al., Real-Time Detection and Capture of Invasive Cell Subpopulations from Co-Cultures. J. Vis. Exp. (2022).

This video describes an invasion assay using a three-chambered array based on electrical impedance. This technique measures the invasive potential of cancer cells under the influence of soluble factors secreted by the resident stromal cells in tumor microenvironments.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

1. New chamber design (Figure 1)

  1. Open a new dual-chamber cell analyzer plate. Set aside the top chamber with electrodes.
  2. Using a milling machine, shave off 2 mm of the U-shaped bottom wells of the cell analyzer plate.
  3. Attach a 2 cm x 7 cm polyethersulfone (PES) membrane with 0.2 μm pore size to the bottom of the shaved wells using UV-curated adhesive. Allow 30 min curation time to ensure the glue is completely cured and inert.
  4. Using a milling machine, cut out two longitudinal slits (1.5 mm x 5.6 mm) along the sides to snap into the ridges of the newly fabricated third chamber.
  5. Using a milling machine, create a t....

Access restricted. Please log in or start a trial to view this content.

Results

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Microfluidic device components labeled with electrodes, slits; setup for fluid control study.
Figure 1: Images of the array chambers and modifications.

(A) The three chambers used to build the array. No modification was made on the top chamber harboring the electrodes. (B) From the middle ch.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
No conflicts of interest declared.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
0.05% Trypsin-EDTAThermofisher25300-054
AdhesiveNorland Optical Adhesive NOA63
Bovine serum albumin (BSA)SigmaA9418
Cell lifterSarstedt83.1832
Cholera Toxin from Vibrio choleraeThermofisher 12585-014
CIM-plateAgilent 5665817001Cell analyzer plate
Collagenase from Clostridium histolyticumSigmaC0130
DMEMThermofisher 11995-065
DMEM-F12Thermofisher11875-093
Fetal Bovine Serum (FBS), Heat InactivatedOmega Scientific FB-12
HEPES Thermofisher15630106
Horse serum (HS)Gibco16050-122
Human EGFPeprotechAF-100-15
Hydrocortisone Sigma H4001
Insulin Transferrin Selenium Ethanolamine (ITSX) (100x)Thermofisher51500056
Insulin, Human Recombinant, Zinc SolutionSigmaC8052
J2 FibroblastsStemcell (RRID:CVCL_W667) 100-0353
LymphoPrepStemcell7851Density gradient medium for the isolation of mononuclear cells
MatrigelCorning354230Basement membrane matrix
MCFDCIS.com cells ( DCIS)RRID:CVCL_5552
MDA-MB-231 cells RRID:CVCL_0062
Milling machine Bridgeport Series 1 Vertical
Phosphate-buffered saline (1x)Thermofisher10010049
Polyethersulfone (PES) membraneSterlitechPCTF029030
RBC lysis solutionStemcell7800
RTCA DP analyzerAgilent3X16 Dual purpose cell analyzer
TrypsinSigma T4799