1. Pressure Transducer Calibration
In this demonstration, a subsonic wind tunnel with a 2.6 ft x 3.7 ft test section and maximum dynamic pressure setting of 25 psf was used. A pre-calibrated pressure transducer was used to set the dynamic pressure in the wind tunnel test section, and a differential U-tube manometer with colored water and scale was used to measure fluid height (Figuur 3). A differential pressure transducer (Figuur 4), standard voltage supply (to power the transducer), and multimeter (to read the output current from the transducer) were also used, shown in Figuur 5.

Figuur 3. Differentiële drukhulpmeetinstrument U-buis met manometer.

Figuur 4. Differentiële drukhulpmeetinstrument.

Figuur 5. Stroomvoorziening (links) en multimeter (rechts).

Figuur 6. Pitot-statische buis.
Tabel 1. Gegevens verzameld voor het drukkalibratie-experiment
| Ptransducer (psf) |
hmanometer (in) |
Itransducer (mA) |
| WT transducer | manometer | multimeter |
Bron: Shreyas Narsipur, Mechanical and Aerospace Engineering, North Carolina State University, Raleigh, NC