The atomic emission spectrometer comprises a sample nebulizer, a plasma source, a wavelength-isolating device, transducers, and a microprocessor controller.
The sample is drawn using a peristaltic pump and transported to a nebulizer, which converts the sample solution into fine droplets.
Larger droplets are directed to a drain, while the fine droplets are carried by argon gas to the plasma—commonly, direct current plasma or inductively coupled plasma.
In the plasma, the samples are atomized to gas-phase atoms and ions and, on further collision, get excited.
The atomic or ionic emission from the plasma is then separated by a monochromator, polychromator, or spectrograph.
Monochromators separate individual wavelengths sequentially through a single exit slit, whereas polychromators with several exit slits isolate multiple wavelengths concurrently.
Alternatively, a spectrograph enables a range of wavelengths to be monitored simultaneously.
The isolated radiation is converted into electrical signals by single or multiple photomultiplier transducers or array detectors and is further analyzed.
The instrumentation of atomic emission spectrometry (AES) involves various components, including atomization devices that convert samples into gas-pha…