1. Pulmonary Infection by Intratracheal Intubation
- Weigh mice and, optionally, marks can be made on the ears for easy identification.
- Anesthetize the mice with ketamine (100 μg per g of mouse weight) and xylazine (10 μg per g of mouse weight) by intraperitoneal inoculation.
- Place mice in cages until fully anesthetized. Squeeze the pads of their feet to check pedal reflex. Mice should display reduced or no reflex reaction.
- Place the mouse on the intubation stand lying on its back.
- Fix a rubber band to intubation stand and then place it under the upper incisors of the mouse. Use tape to fix legs and arms onto the intubation stand.
- Gently move tougue out of the mouth using forceps.
- Insert speculum with octoscope inside of the mouth and look into oropharynx until larynx opening is visible.
- Advance guide wire covered with cathether into trachea until catether hub is at the incisors. Remove guide wire. Put air into cathether to observe chest expansion and confirm correct intubation.
- Inject 50 μL of bacterial (in the images shown we used bacillus Calmette Guerin (BCG) expressing click beetle red luciferase (CBRLuc), constructed in our laboratory, but any luminescent bacterial strain can be used) solution through cathether using a 1 mL syringe.
- Place mouse in a cage and observe it during recovery from anesthesia.
2. Animal Anesthetization and Preparation for Bioluminescence Imaging
Mice are anesthetized with isoflurane using the XGI-8 Gas Aneshesia System.
- Before operating the XGI-8 anesthesia system, weigh each charcoal filter canister and write the weight and date on it. If the weight is 50 gram over of the initial weight, replace it with a new canister.
- Check the isoflurane level in the vaporizer and fill it if necessary.
- Turn on the evacuation pump in the front of the XGI-8 anesthesia system and set to 8 lpm (liters per minuite).
- Turn on the oxygen supply from the high pressure cylinder and set it to 55 psig.
- Turn on the oxygen toggle (green) on the front of the XGI-8 anesthesia system.
- Turn on the gas flow to the imaging chamber to set the level of gas flow. Set to 0.25 lpm and then turn off the gas flow.
- Turn on the gas flow to the anesthesia induction chamber to set the level of gas flow. Set to 1.5 lpm and then turn off gas flow.
- Turn on the isoflurane with the vaporizer and set to 2-2.5 percent. The isoflurane level can be adjusted depending upon the number of animals being used and weight.
- Place mice into the induction chamber and close the lid. Turn on the gas flow to induction chamber. Place mice in the chamber for 5 - 10 minuites until they are anesthetize completely.
- Apply optical ointment to the eyes to protect the mouse eyes during imaging and place the mice in the imaging chamber in one of the nose cones on the anesthesia manifold. Use light baffle between mice to prevent the reflection of light onto adjacent animal subjects.
- Inject luciferin (150 μg/gram of body weight) by intraperitoneal inoculation.
3. Bioluminescence Imaging
- Start the Living Imaging Software.
- If the system is not initialized, initialize the IVIS Imaging system.
- Set the parameters for imaging by clicking sequence setup in IVIS acquisition control panel.
Select luminescence and photograph in Imaging Mode. If the option to carry out DLIT 3D reconstruction is desired, select photographic, luminescent and structural light images as part of the image sequence.
Set exposure time from 0.5 sec to 10 minutes.
Set Binning and F/stop based on expected brightness of sample.
Set excitation filter to block and emission filter to open, unless planning to acquire only specific wavelengths of light. In the case of DLIT 3D constitution, set multiple emission filters of different wavelengths to allow accurate localization of source.
Set FOV from A to D depending on the number of mice or region of animal to be imaged. Select A and B for 1 mouse, C for 2-3 mice, and D for 4-5 mice.
- Click add in the Image Wizard in the Acquisition Control Panel to add the sequence setup.
- Start imaging sequence by clicking acquire.
In the case of DLIT 3D construction, luminescent images will be acquired for multiple emission filters at different wavelengths. Photographic and structured light images will also be acquired.
- During acquisition the image and edit image levels panels will be visible. Fill in as detailed information as possible for each experiment in edit image to ensure easy tracking of images at later times and save your images.
- ReRemove the mice from IVIS imaging chamber and return them to their cages. Observe recovery from anesthesia to ensure mice were not significantly affected by the procedure. Animals should be monitored constantly until they fully recover from anesthesia (usually 1-2 min).
4. ex vivo Imaging and CFU Analysis for the Quantification of Bacteria in the Lungs
- Inject mice with luciferin intraperitoneally in the same manner as for imaging, just prior to euthanasia.
- Euthanize the mice by intraperitoneal injection of 100 mg/kg pentobarbitol 5 min after luciferin injection.
- Explant the lungs from the mice and place in sterile petri-dishes using sterile forceps and scissors.
- Place petri-dish containing explanted organ in imaging chamber and acquire bioluminescence images in the same manner as for whole mouse.
- After imaging, remove explanted organ from petri dish using sterile forceps and homogenize in 1 mL PBS.
- Make dilutions and plate tissue on the appropriate selective media. Homogenized tissue can be stored -80°C for re-plating should it be necessary. Alternatively, homogenized tissue can be used for RNA or DNA extraction to quantify bacterial numbers by qPCR.
5. Analysis of Luminescence Imaging
- Start "Living Imaging Software" and browse the image files.
- Use "Tool Palette' to adjust images. Click Image Adjust to modify correction/filtering settings and min/max for individual color scales.
- Use ROI tools in pulldown list for the quantification of light intensity. Select ROI shape, number and size. Drag ROI frame to region of interest on image. Ensure all ROI are the same size and shape. Click ROI measurement and save, copy and/or export quantitative data.
- In the case of DLIT 3D reconstructions, load image sequence including the structured light image. Click Surface topography in the Tool Platte. In pulldown list, click Reconstruct tab to generate the surface topography. A surface mess will then appear in the window.
- Select DLIT 3D reconstruction in the Tool Platte. In pulldown list, click the Properties tab to set tissue properties and source spectrum. Muscle is recommended under most circumstances. In the pulldown list, click the Analyze tab to select the wavelength for analysis. Click the Parameter tab and confirm default values or adjust DLIT parameters if necessary. Click Reconstruct tab to start 3D analysis.
- When 3D reconstruction is finished, the 3D view will display the results of 3D reconstruction. Click results tab to see analysis data for photon density, voxels and DLIT algorithm parameters.
- Save and/or export the results of the 3D reconstruction analyses into figures and data files.
6. Representative Results:
The bioluminescence images of mice infected with bioluminescent bacteria along with an uninfected mouse control are shown in Figure 1. The mice pulmonary infected with bioluminescent bacteria produce significant signal from the lungs (Figure 1). Luminescence intensity is quantified as total flux within a R.O.I (region of interested) (Figure 2). The quantitative data for light intensity is normalized for colony forming units (CFU) of bacteria obtained from lungs to confirm that the signal is produced from the bioluminescent bacteria and can be compared to negative control. The location and intensity of signal can be further analyzed by DLIT 3D reconstruction of the luminescence source based on tomography of the mouse surface11. These analyses allow quantification and localization of the bioluminescent signal produced. The results of 3D reconstruction of a luminescent source in infected mice demonstrates that light is produced from lungs of the mouse (Figure 3). The ex vivo images of the resulting mouse lungs confirm that luminescence is emitted from the lungs, rather than some other closely juxtaposed tissue or organ (Figure 2C).

Figure 1. Luminescence imaging of pulmonary infected mice with bioluminescent bacteria tagged with CBRluc. Uninfected control mouse is on the left and two infected mice are on the right. Mice were infected with bacteria expressing CBRluc (n=2) via the intratracheal route. 10 minutes after luciferin injection, luminescence images were acquired for 10 min at 4 positions: dorsal, ventral, left side and right side.

Figure 2. The quantitative light intensity from mice infected with bacteria expressing CBRluc. (A, B) Luminescence images of ROI analysis and the quantified total flux of light from R.O.I. in dorsal and ventral positions, respectively. Uninfected mice are on the left and two infected mice are on the right. Luminescence images were acquired from lungs of mice uninfected and infected with CBRlux (n=2). The light intensity around lungs was quantified by ROI analysis. C) ex vivo images of lungs from uninfected (top) and infected (bottom two sets of lungs) mice. Luciferin was injected 5 miniuts before of euthanasia and images were then acquired. The quantified values are normalized to CFU.

Figure 3. The 3D reconstruction of bioluminescence source(s) from a pulmonary infected mouse. The mouse was infected with bacteria expressing CBRluc by intratracheal injection. The image sequence was acquired using different emission filters of serial wavelengths from 540 nm to 700 nm. An image sequence was used for 3D reconstitution for luminescence source in animal subjects that contained a structured imag. A) The tomography for mouse is shown in different positions: front, back, left and right. Light sources are shown as voxels (red boxes within 3D mouse) that are located within lung as determined by reconstruction. B) Photon densiy map of measured and simulated data. Comparing the measured and simulated photon density curves provide the information of quality of reconstruction. Good quality reconstructions result in similar measured and simulated photon densities.