Method Article

A Microfluidic Chip-Based Method for Rapid Neutrophil Chemotaxis Analysis Using Whole Blood

July 8th, 2025

In This Article

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Source: Yang, K., et al. An All-on-chip Method for Rapid Neutrophil Chemotaxis Analysis Directly from a Drop of Blood. J. Vis. Exp. (124), (2017)

This video demonstrates neutrophil chemotaxis assay using an all-on-chip method. The method allows direct isolation of neutrophils from the whole blood by immunomagnetic negative selection followed by chemotaxis assay on a microfluidic chip.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

All procedures involving sample collection have been performed in accordance with the institute's IRB guidelines.

1. Microfluidic device preparation.

  1. Microfluidic Cell Migration Assay Preparation
    1. Prepare 50 µg/mL fibronectin solution by diluting 50 µL of stock fibronectin solution (1 mg/mL) to 950 µL Dulbecco's phosphate-buffered saline (DPBS) inside a biosafety cabinet.
    2. Prepare the migration medium by mixing 9 mL of Roswell Park Memorial Institute medium (RPMI-1640) and 1 mL of RPMI-1640 with 4% bovine serum albumin (BSA).
    3. Remove....

Access restricted. Please log in or start a trial to view this content.

Results

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Microfluidic chip diagram; chemotaxis test; cell isolation; magnetic neutrophil selection process.
Figure 1: Illustration of the all-on-chip method for neutrophil chemotaxis analysis. (A) Illustration of the microfluidic device. The device includes two layers. The first layer (4 µm high) defines the cell docking barrier channel to.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
No conflicts of interest declared.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Device fabrication
Mask alignerABMN/A
SpinnerSolitec5000
HotplateVWR11301-022
Plasma cleanerHarrick PlasmaPDC-001
Vacuum dessicatorFisher Scientific08-594-15A
SU-8 2000 thinnerMicrochemSU-8 2000
SU-8 2025 photoresistMicrochemSU-8 2025
(tridecafluoro-1,1,2,2- tetrahydrooctyl) trichlorosilaneGelest78560-45-9
Si waferSilicon, IncLG2065
Glass slidesFisher Scientific12-544-4
Cutting padN/AN/ACustom-made
Punchers
On-chip cell isolation and chemotaxis assay
RPMI 1640Fisher ScientificSH3025502
DPBSFisher ScientificSH3002802
Bovine serum albumin (BSA)Sigma-AldrichSH3057402
FibronectinVWRCACB356008
fMLPSigma-AldrichF3506-10MG
Magnetic disksIndigo Instruments44202-15 mm in diameter, 1 mm thick
FITC-DextranSigma-AldrichFD10S
RhodamineSigma-AldrichR4127-5G
Giemsa stain solutionRowley Biochemical IncG-472-1-8OZ
EasySep Direct Human Neutrophil Isolation KitSTEMCELL Technologies Inc19666
DithiothreitolSigma-AldrichD0632
Nikon Ti-U inverted fluorescent microscopeNikonTi-U
Microscope environmental chamber.InVivo ScientificN/A
CCD cameraNikonDS-Fi1

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Immunomagnetic SelectionChemoattractant GradientFluorescence MicroscopyCell DockingFibronectin CoatingMigration MediumTemperature Control