$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
All procedures involving animal models have been reviewed by the local institutional animal care committee and the JoVE veterinary review board.
- Intravital imaging in the zebrafish brain
NOTE: To properly locate the fish brain under the objective lens, a secondary charge-coupled device (CCD) camera is used on the same path as the excitation light for widefield imaging.
- Set up the microscope and calibrate the power.
- Place a low magnification (typically 4x) objective on the microscope.
- Place the Petri dish containing the fish and the tubes under the microscope.
- Use a light-emitting diode (LED) light source to illuminate the Petri dish.
- Open the Camera mode of the image acquisition software (Figure 1).
- Click Live.
- Choose channel A on the right side of the screen.
- Adjust the histogram settings to see the image clearly.
NOTE: These need to be updated as needed. - Set the motor setting on the motor controller to Base.
- Lower the objective until the fish is visible.
NOTE: Ensure that the objective lens does not make any physical contact with the head. - Place the center of the fish head at the center of the field of view.
- Move the objective up and away from the fish head.
- Replace the low magnification objective lens with the high numerical aperture (NA) objective lens for 3PM (three-photon microscopy).
NOTE: The low magnification lens and the high NA objective lens do not need to be parfocal but must be close enough to ensure the fish is within the field of view of the high NA objective lens. - Slowly lower the objective lens, ensuring that the objective does not make any physical contact with the head. In the CCD camera software, stop moving the objective when the top of the head is visible. Set the z location to 0 μm.
NOTE: Ensure that there are no air bubbles under the objective lens when using a water-immersion objective. - Turn off the LED light source and close the dark curtain around the system.
- Set the imaging acquisition software to Multiphoton GG mode for 3P imaging and set the power under the objective lens to less than 1 mW (with ~1 MHz pulse repetition rate).
- Change the motor setting button from Base to Objective on the motor controller.
- Turn down the lights in the room.
- Turn on the photomultiplier tubes (PMTs) and open the 3PM excitation source shutter. Ensure that an outline of the bone appears in the fluorescence signal channel due to autofluorescence and in the third harmonic generation (THG) signal channel due to THG of the bone (Figure 1C).
- Perform imaging at different depths by increasing the power levels when imaging deeper.