Method Article

Microfluidic Chip Fabrication and Method to Detect Influenza

DOI:

10.3791/50325

March 26th, 2013

In This Article

Summary

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

An integrated microfluidic thermoplastic chip has been developed for use as a molecular diagnostic. The chip performs nucleic acid extraction, reverse transcriptase, and PCR. Methods for fabricating and running the chip are described.

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Fast and effective diagnostics play an important role in controlling infectious disease by enabling effective patient management and treatment. Here, we present an integrated microfluidic thermoplastic chip with the ability to amplify influenza A virus in patient nasopharyngeal (NP) swabs and aspirates. Upon loading the patient sample, the microfluidic device sequentially carries out on-chip cell lysis, RNA purification and concentration steps within the solid phase extraction (SPE), reverse transcription (RT) and polymerase chain reaction (PCR) in RT-PCR chambers, respectively. End-point detection is performed using an off-chip Bioanalyzer (Agilent Technologies, Santa Clara, CA). For peripherals, we used a single syringe pump to drive reagent and samples, while two thin film heaters were used as the heat sources for the RT and PCR chambers. The chip is designed to be single layer and suitable for high throughput manufacturing to reduce the fabrication time and cost. The microfluidic chip provides a platform to analyze a wide variety of virus and bacteria, limited only by changes in reagent design needed to detect new pathogens of interest.

Introduction

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Millions of deaths have been reported during the three influenza pandemics of the 20th century1. Moreover, the most recent influenza pandemic was declared by World Health Organization (WHO) 2 in 2009, and as of August 1, 2010, 18,449 deaths were reported by WHO3. This pandemic demonstrated again the high burden of infectious disease, and the need for rapid and accurate detection of influenza to enable fast disease confirmation, appropriate public health response and effective treatment4.

There are several methods widely used for diagnosing influenza, these include rapid immunoassays, direct fluorescent antigen testing (DFA) and viral culture methods. Rapid immunoassays dramatically lack sensitivity5-8, while the other two methods are labor-intensive and time consuming9. Molecular tests offer multiple advantages including a short turn-around time, high sensitivity, and higher specificity. Several commercial entities have been working towards fast molecular tests (also called nucleic acid tests or NATs) for infectious diseases, and several have influenza assays in their pipelines. However most of them require off-chip sample preparation. None of the Clinical Laboratory Improvement Amendments (CLIA) waived molecular tests incorporate sample preparation into the assay cartridge or module.

Lab-on-a-chip technology plays an important role in the development of point-of-care testing devices. After the introduction of the first PCR chip in 199310, numerous efforts have been put into developing nucleic acid test chips. However, only a few of these have integrated crude sample preparation with downstream amplification.

We have previously demonstrated the miniaturization of a solid phase extraction column (SPE) into a plastic microfluidic chip11 and the development and optimization of a continuous flow PCR chip12. Here, we extend the previous work to integrate the SPE with RT and PCR steps into a single chip for clinical diagnostics and demonstrate its capability to amplify nucleic acids from patient nasopharyngeal (NP) swabs and aspirates.

Access restricted. Please log in or start a trial to view this content.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

1. Chip Fabrication12

  1. Make two plaques from Zeonex 690R pellets: distribute 8-9 grams Zeonex pellets evenly in the center of a metal plate, preheat on the heated press at 198 °C for 5 min, and then apply pressure slowly to 2,500 psi for another 5 min. To complete this step, we used a Carver hot press.
  2. Emboss the microfluidic channel in the plaque with an epoxy mold. Details on the mold fabrication are outlined elsewhere12 (channel design in Figure 2b): put one plaque onto the epoxy mold; preheat them on the heated press at 157 °C for 10 min and then apply pressure slowly to 1,000 psi for another 10 min. Remove the plaque from the mold by hand or using a tongs before it cools down. (Note: wear thermal gloves during this step.)
  3. Drill holes in the embossed chip at the inlet, waste port, and the outlet of the microfluidic channel. We used a 1.25 mm diameter drill bit to make the three holes.
  4. Bond the embossed chip with another plaque (on top): Wash the chips with IPA, RNAse Away and deionized water in sequence; Air dry and preheat them on the heated press at 131 °C for 10 min and then press at 350 psi for another 10 min.
  5. Attach Nanoports at the inlet, waste, and the outlet ports separately using J-B Weld Epoxy; cure for 15 to 24 hr as per the manufacturer's instructions.
  6. Rinse the SPE channel with 50 μl RNAse Away, followed with 100 μl nuclease free water.
  7. Load the SPE channel with 4 μl of the grafting solution (methyl methacrylate with 3% w/v benzophenone) and then crosslink the methacrylate by incubating for 10 min in a UV oven under 365 nm UV wavelength and 2,000 mJ/cm2. Remove the residual grafting solution with vacuum. We used the wall vacuum. For the best results, a fresh solution should be used. It is usually acceptable, however, to store a prepared solution at 4 °C for up to one week.
  8. Make the SPE column solution (all v/v %): 16% ethylene dimethacrylate, 24% butyl methacrylate, 42% 1-dodecanol, 18% cyclohexanol, and add 1% 2-dimethylamino-4-(methyl-phenylamino)-phenol as the photo-initiator.
  9. Take out the same volume of silica microspheres solution as the SPE column solution (to make a 1:1 mixture), and completely dry it at 70 °C overnight in a vacuum oven. Break up the pellet by tapping on the tube with the lid closed. Add the SPE column solution into the dried silica microsphere tube and vortex to mix.
  10. Load 4 μl of the SPE solution into the same channel and then crosslink it by UV irradiation for 2.5 min. Then, flip the chip over and irradiate for another 2.5 min. The polymerized SPE column should be an opaque white solid.
  11. Wash the channel with 500 μl of 100% methanol to rinse away excess reactants. This wash also removes intra-polymer porogenic solvents, thereby creating the open pore structure of the SPE column when dried in ambient conditions.
  12. Attach two thin-film heaters to the bottom of the chip with thermally conductive tape. The sides of the heaters need to be aligned with the edge of the wide channels for denaturation and annealing separately (see Figure 1).
  13. Place five thermocouples into the chip through the dead ended open side channels of each chip (see Figure 2b). The thermocouples stay in these channels via press fit. Tape was used to further fix the location of thermocouples.

2. Solid Phase Extraction Method

  1. Prepare 96 μl of 70% ethanol with DEPC-treated and autoclaved water (deionized or Millipore-filtered).
  2. Add 4 μl of 25X RNA Secure to 96 μl of 70% ethanol in one tube, and duplicate the same mixture with 100% ethanol in a separate tube.
  3. Prepare 300 μl of channel buffer by mixing the following: 75 μl of 6 M guanidine thiocyanate (GuSCN),150 μl of 2-propanol, 63 μl of nuclease free water, and 12 μl of 25X RNA Secure.
  4. Prepare 312 μl of lysis buffer by mixing: 100 μl of 6 M GuSCN, 200 μl of 2-propanol, and 12 μl 25X RNA Secure.
  5. Heat the 70% ethanol, 100% ethanol, channel buffer and lysis buffers at 60 °C for 10-20 min to activate the RNA Secure.
  6. To equilibrate the microfluidic channel, run the channel buffer through the SPE channel at a flow rate of 0.8 ml/hr.
  7. Quick-thaw the test sample in VTM at 37 °C in water bath, and remove the sample immediately from water bath as soon as it is thawed.
  8. Centrifuge sample at 13,000 rpm for 10 min, and transfer 100 μl of the supernatant into 300 μl of the lysis buffer. Complete the mixture by adding 6 μl of 1 μg/μl carrier RNA.
  9. Vortex to mix, and after spin for 5 sec, load the entire lysate into a Luer-lok 1 cc syringe.
  10. Run lysate through the SPE channel at a flow rate of 0.8 ml/hr .
  11. Wash the SPE channel with 100 μl of 70% ethanol , followed by 100 μl of100% ethanol at 1 ml/hr. (There is a 50 μl dead volume in the syringe tip that will not get pushed out, so actually only 50 μl is going through the channel.)
  12. Position an empty syringe at the 0.5 ml mark and push air through the channel to dry it at 1 ml/hr.
  13. Run 67.5 μl of nuclease-free water through the channel to elute the bound nucleic acids at 0.5 ml/hr, and collect 13.5 μl from the nanoport at the waste port. (The fluid loss is due to the 54 μl dead volume in the syringe tip and the SPE channel.)

3. RT-PCR

  1. Prepare 36.5 μl of the RT-PCR master mix during the air dry step (2.12) using the reagents in a Qiagen OneStep RT-PCR kits.
  2. Load RT-PCR reagent in the nanoport at the waste port, and mix with RNA to get the following 50 μl RT-PCR reaction: 13.5 μl eluted RNA in nuclease free water, 4 μl RT-PCR enzyme mix, 10 μl Q solution, 10 μl1X one step buffer, 2 μl 25 mM MgCl2, 1 μl 50 μM forward primer 5'-GAC CRA TCC TGT CAC CTC TGA C-3', 1 μl 50 μM reverse primer 5'-AGG GCA TTY TGG ACA AAK CGT CTA-3', 2 μl dNTP, 0.75 μl 1.0% w/v BSA, 0.73 μl PEG8000 and 5.02 μl nuclease free water.
  3. Load the RT-PCR mixture into the RT channel by gentle vacuum (we used the wall vacuum) and seal the Nanoport with a closed fitting.
  4. Apply 35 V to heater 1. Keep the reagents in the RT channel for 30 min after the heater 1 equilibrates at 50 °C. Keep the reagents in the RT channel for 15 min after the heater 1 equilibrates at 95 °C.
  5. Apply 27 V to heater 1 and 50 V to heater 2, wait about three minutes or until heater 1 equilibrates at 60 °C and heater 2 equilibrates at 95 °C.
  6. Push the reagents into the PCR channel at 0.5 μl /min. It should take about 20 min for the reagents to flow through the serpentine channel.
  7. As the sample makes it to the end of the serpentine channel, collect the PCR products at the outlet using an appropriately sized pipettor and tip.

4. PCR Products Detection

  1. We use an Agilent High Sensitivity DNA test to detect the products. To run this test take out the Agilent High Sensitivity DNA Kit from refrigerator 30 min before testing.
  2. Turn on the Bioanalyzer and open the 2100 Expert software.
  3. Load the 1 μl of the PCR products into the testing and follow the Agilent protocol (http://gcf.pbrc.edu/docs/Agilent/Agilent%20Manual.pdf).
  4. Analyze and record the data.

Access restricted. Please log in or start a trial to view this content.

Results

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

A typical result is shown in Figure 3 for an influenza A infected nasopharyngeal wash specimen. Due to the different amounts of influenza virus in each patient specimen, the final concentration of PCR product will vary. A good result should have low noise, two clear ladder peaks (35 and 10380 bp) and a single product peak at the designed product size (107 bp) for the positive sample. While the product peak should theoretically be absent for negative controls, we did observe spurious PCR peaks near...

Access restricted. Please log in or start a trial to view this content.

Discussion

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The diagnostic method presented here demonstrated the ability of an integrated microfluidic plastic chip to amplify influenza A RNA from patient specimens with high specificity and a low detection limit.13 We designed this chip for potential point of care testing: (a) the temperature and fluidic control were simplified, (b) the chip is low cost and suitable for high throughput fabrication using injection molding, and (c) the chip is disposable and intended for one time use, thus reducing the concern of s...

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The authors have declared no competing financial interests.

Acknowledgements

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

This research was supported by National Institutes of Health (NIH) grant R01 EB008268.

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
1-dodecanolSigma-Aldrich, St. Louis, MO443816-500G
2,2-Dimethoxy-2-phenylacetophenoneSigma-Aldrich, St. Louis, MO196118-50G
2100 BioanalyzerAgilent Technologies, Santa Clara, CAG2943CA
2-PropanolSigma-Aldrich, St. Louis, MO19516
BenzophenoneSigma-Aldrich, St. Louis, MO239852-50G
BSAThermo Fisher Scientific,pittsburge, PAA7979-50ML
Butyl methacrylateSigma-Aldrich, St. Louis, MO235865-100 ml
Carrier RNAQiagen, Valencia, CA1017647
CyclohexanolSigma-Aldrich, St. Louis, MO105899-1L
EthanolSigma-Aldrich, St. Louis, MOE7023
Ethylene dimethacrylateSigma-Aldrich, St. Louis, MO335861
Ethylene glycol dimethacrylateSigma-Aldrich, St. Louis, MO335681-100ML
Glass syringe 250 μlHamilton, Reno, NV81127
Guanidine thiocyanateSigma-Aldrich, St. Louis, MO50981
High Sensitivity DNA KitAgilent Technologies, Santa Clara, CA5067-4626
Hot pressCarver,Wabash, IN4386
J-B Weld EpoxiesMcmaster-Carr,Elmhurst, IL7605A11
Luer-Lok syringesBD-Medical, Franklin Lakes, NJ309628
Magnesium ChlorideThermo Fisher Scientific,pittsburge, PAAB-0359
MethanolSigma-Aldrich, St. Louis, MO494437
Methyl methacrylateSigma-Aldrich, St. Louis, MOM55909
NanoportUpchurch ScientificN-333-01
Nanoport FittingUpchurch ScientificF-120x
Nuclease free waterThermo Fisher Scientific,pittsburge, PAPR-P1193
OneStep RT-PCR kitQiagen, Valencia, CA210210
PEG8000Sigma-Aldrich, St. Louis, MO41009
Power supplyVWR,Radnor, PA300V
RNAse AwaySigma-Aldrich, St. Louis, MO83931-250ML
RNASecureApplied Biosystems, Foster City, CAAM7005
Silica microspheresPolysciences,Warrington, PA24324-15
Syringe pump Harvard Apparatus,Holliston, MAHA2000P/10
Thermally Conductive TapeMcmaster-Carr,Elmhurst, IL6838A11
ThermocoupleOmega Engineering, Stamford, CT5SRTC-TT-J-40-36
Thin-film HeatersMinco,Minneapolis, MNHK5166R529L12A
Ultraviolet CrosslinkerUPV, Upland, CACL-1000
ZeonexZeon Chemicals, Louisville, KY690R

References

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
  1. Nicholls, H. Pandemic Influenza: The Inside Story. PLoS Biol. 4, (2006).
  2. WHO | World now at the start of 2009 influenza pandemic. , World Health Organization. Geneva, Switzerland. Available from: http://www.who.int/mediacentre/news/statements/2009/h1n1_pandemic_phase6_20090611/en/ (2009).
  3. WHO | Pandemic (H1N1) 2009 - update 112. , World Health Organization. Geneva, Switzerland. Available from: http://www.who.int/csr/don/2010_08_06/en/ (2009).
  4. Wenzel, J. J., et al. Analytical performance determination and clinical validation of the novel roche realtime ready influenza A/H1N1 detection set. Journal of Clinical Microbiology. 48, 3088-3094 (2010).
  5. Chan, K. H., et al. Analytical sensitivity of rapid influenza antigen detection tests for swine-origin influenza virus (H1N1. Journal of Clinical Virology: the. 45, 205-207 (2009).
  6. Ginocchio, C. C., et al. Evaluation of multiple test methods for the detection of the novel 2009 influenza A (H1N1) during the New York City outbreak. Journal of. 45, 191-195 (2009).
  7. Aeron, C. H., et al. Performance of influenza rapid point-of-care tests in the detection of swine lineage A(H1N1) influenza viruses. Influenza and Other Respiratory Viruses. 3, 171-176 (2009).
  8. Takahashi, H., Otsuka, Y., Patterson, B. Diagnostic tests for influenza and other respiratory viruses: determining performance specifications based on clinical setting. Journal of Infection and Chemotherapy. 16, 155-161 (2010).
  9. Selvaraju, S. B., Selvarangan, R. Evaluation of Three Influenza A and B Real-Time Reverse Transcription-PCR Assays and a New 2009 H1N1 Assay for Detection of Influenza Viruses. Journal of Clinical Microbiology. 48, 3870-3875 (2009).
  10. Northrup, M. A., Ching, M. T., White, R. M., Watson, R. T. Transducer'93, seventh international conference on solid state sensors and actuators. , 924-926 (1993).
  11. Bhattacharyya, A., Klapperich, C. M. Thermoplastic microfluidic device for on-chip purification of nucleic acids for disposable diagnostics. Analytical Chemistry. 78, 788-792 (2006).
  12. Cao, Q., Kim, M. -C., Klapperich, C. Plastic microfluidic chip for continuous-flow polymerase chain reaction: Simulations and experiments. Biotechnology Journal. 6, 177-184 (1002).
  13. Cao, Q., et al. Microfluidic Chip for Molecular Amplification of Influenza A RNA in Human Respiratory Specimens. PLoS ONE. 7, e33176(2012).
  14. Rådström, P. Purification and Characterization of PCR-Inhibitory Components in Blood Cells. J. Clin. Microbiol. 39, 485-493 (2001).
  15. Boddinghaus, B., Wichelhaus, T. A., Brade, V., Bittner, T. Removal of PCR Inhibitors by Silica Membranes: Evaluating the Amplicor Mycobacterium tuberculosis Kit. J. Clin. Microbiol. 39, 3750-3752 (2001).
  16. Andreasen, D., et al. Improved microRNA quantification in total RNA from clinical samples. Methods. 50, S6-S9 (2010).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Microfluidic ChipInfluenza DetectionRNA PurificationReverse TranscriptionPolymerase Chain ReactionSolid Phase ExtractionCapillary ElectrophoresisThin Film HeatersSyringe PumpBioanalyzer