Laser-induced breakdown spectroscopy (LIBS) is a simple method of elemental analysis that uses a laser-generated spark as the excitation source. The laser pulse is focused onto a surface that heats, ablates, atomizes and ionizes the surface material resulting in the formation of plasma. The plasma light is spectrally resolved and detected and elements are identified by their spectral signatures. If properly calibrated, LIBS can provide quantitative results. LIBS can analyze solids, gases, and liquids with little or no sample preparation.1 These characteristics make it ideal for analyses that cannot be carried out in the laboratory.
Currently, LIBS is being studied for many different applications especially those that require field-based measurements for quantification.1-8 This requires the development of LIBS instrumentation using rugged and compact components suitable for a field-based system. In most cases, these components will not have the full capabilities of laboratory-based instrumentation, thereby compromising the analysis performance. LIBS results are dependent on laser pulse parameters and other measurement conditions that include sampling geometry, surrounding atmosphere, and the use of gated or non-gated detection.9-12 For field-based LIBS instrumentation, two important factors to consider are the pulse energy and the use of gated versus non-gated detection. These two factors determine to a large extent the cost, size, and complexity of the LIBS instrument. Small, ruggedly built lasers that can generate pulses from 10-50 mJ at repetition rates of 0.3-10 Hz are commercially available and would be highly advantageous to use. Therefore, it is important to know what, if any, loss in detection capabilities will result from the use of these lasers. The pulse energy is a key parameter for LIBS as it determines the amount of material ablated and vaporized and the excitation characteristics of the plasma. In addition, the use of gated detection can increase the cost of the LIBS system; as a result, it is imperative to determine the differences between spectra and detection capabilities using gated and non-gated detection.
Recently, a study was performed comparing gated detection to non-gated detection for minor elements found in steel. The results showed that the detection limits were comparable if not better for non-gated detection.12 One important characteristic of LIBS is that the technique experiences physical and chemical matrix effects. An example of the former is that the laser pulse couples more efficiently with conducting/metal surfaces than non-conducting surfaces.13 For this study, we wanted to determine the effects of pulse energy and timing parameters for non-conducting materials like soil simulants.
Although, field portable LIBS instruments have been developed and used for some applications, a comprehensive study on the detection capabilities has not been performed comparing higher energy and gated systems to lower energy and non-gated systems using soil simulants. This study focuses on laser pulse energy and timing parameters for determination of trace elements in complex matrices. The laser pulse energy ranged from 10 to 100 mJ to obtain a comparison between lower and higher energies. A comparison of the use of gated versus non-gated detection was also performed over the same energy range.