Results presented below have been obtained using a Wohlwend Compact 02 for HPF (Figure 1A). One major advantage of this instrument is the ease of use of the specimen carriers and its holders. When using other instruments, McDonald recommends that two users should carry out the sample preparation and HPF, one preparing the samples while the other does the freezing and transfer to the FS cryovials 9. However, the Wohlwend specimen carriers and holder are easy enough for a single user to manipulate independently (Figure 2A, B, E, F and G). It should be mentioned however that one should perform a few trial runs to become familiar with the HPF instrument before working with valuable samples. An experienced user should be able to fix several (10 or more) samples in an hour. However, the principles presented in this protocol can be used with any of the commercially available HPF machines.
Perhaps the most critical part of the HPF-QFS procedure is the sample preparation for loading into the specimen carrier (Figure 2C). Although this may not be intuitive, it is the step in the protocol over which the researcher has the most control. The HPF machine is robust, and with proper use and maintenance should produce the expected changes in pressure and temperature with little variation between samples. If a sample is poorly handled during preparation then no other step will redeem the damage so caused. Plant tissues are fairly easy to manipulate during this stage of the process as they are not cultured in medium and their cells walls give the cells good strength. It is nonetheless important that samples are handled quickly to avoid ultrastructural changes that result from removal from the parent plant and to limit stress and wounding responses. The smallest specimen carrier that can accommodate a sample should be used to ensure efficient and consistent HPF (Figure 2C). Finally, the specimen carrier should be filled but not overflowing (Figure 2D). The filled specimen carrier is easily inserted into the HPF machine for freezing (Figure 2 H and I).
The first steps in the QFS protocol can be challenging for a beginner, so it is recommended that two people should work together until users are comfortable with the procedure (Figure 3A-E). Care should be taken at all times when handling liquid nitrogen and the fixatives osmium tetroxide and uranyl acetate. A typical curve for temperature changes during QFS is shown in Figure 2J. The temperature increases rapidly from -196 °C, the temperature of liquid nitrogen, to about -80 °C. It is at temperatures of around -78 °C to -90 °C that freeze substitution is believed to occur 9. One challenging step in the QFS protocol is the removal of dry ice after 2 hours of FS. At this step, mishandling of samples can produce a spike in temperatures. For this, the heater block should be swiftly lifted out of the QFS chamber using a cryoglove and the dry ice quickly poured out into a secondary container.
HPF has been used to fix various plant tissues including Nicotiana benthamiana leaves and Arabidopsis leaves and embryos. It is challenging to fix samples from mature leaves due to the large central vacuoles of most cells. Younger leaves contain smaller vacuoles but the trichomes are usually quite densely packed. The presence of the trichomes can make it difficult to pack the samples with yeast paste but care must be taken to ensure that this is properly done to minimize the amount of air trapped between the leaf surface and the paste. The trapped air will impede heat transfer during HPF and reduce the quality of fixation. This is true for any sample.
After HPF and QFS samples may be prepared for viewing under the TEM by infiltrating and embedding with resin. Thin sections of 65-100 nm may then be prepared by sectioning. Typical results are shown in Figure 4. The images shown are all from Arabidopsis leaf samples. Plasma membranes are typically smooth and pressed against the cell wall, a sign of good fixation (Figure 4A, C and E). Other organelles including chloroplasts (Figure 4A, D, F and H) and thylakoids (Figure 4B), mitochondria (Figure 4D and F), Golgi (Figure 4 G), microtubules (Figure 4E) and ribosomes (especially Figure 4C) are also clearly visible and the large central vacuoles remain intact (Figure 4A). Poor handling during HPF-QFS results in artifacts including ice crystal-induced damage (Figure 4D) and plasmolysis (Figure 4H). Lead precipitate may also form during staining of sections (Figure 4F).

Figure 1: The Wohlwend Compact 02 High-Pressure Freezer. (A) The Wohlwend Compact 02 HPF machine used for cryofixation with the attached computer terminal. Samples are inserted into the front of the machine (small circle) for freezing. A temperature curve can be generated on the computer screen for each run, as desired by the user. (B) A typical temperature- pressure curve for a HPF run. The yellow and purple lines represent the temperature and pressure, respectively. Note the steep slopes of both curves. The high pressure is maintained for about 400 msec. Each interval on the x-axis represents 50 msec. Data was collected with EasyScopeII for DSIM12 software. (C) The insulated box and cover used for storage of frozen samples immediately after freezing can be conveniently placed on top the HPF machine. It is filled with liquid nitrogen. The round aluminum containers hold the cryovials.

Figure 2: Preparing tissue sample for HPF. (A) The specimen holder for the Compact 02 HPF machine in its closed configuration. (B) The specimen holder is open. (C) A leaf sample in the 0.2 mm well of a Type A specimen carrier. The other side of this carrier is 0.1 mm deep. (D) The leaf sample covered in yeast paste. Note that the carrier is full but not overflowing. (E) The specimen carrier in the specimen holder. (F) The sample is covered with the Type B carrier. This carrier has one flat surface and on the other side a well that is 0.3 mm deep. Here the flat surface is used to sandwich the sample. (G) The sample holder is closed and ready for insertion into the HPF machine. (H) The orifice on the front of the HPF instrument into which the specimen holder will be inserted for freezing. (I) The temperature and pressure probe/holder inserted into the HPF machine. (J) A typical temperature curve for a QFS run (data collected with EasyLog software). Temperature was recorded each second. The spike at the beginning of the run is due to the electronics of the probe used and does not reflect a real temperature measurement.

Figure 3: Equipment used in QFS. QFS can be carried out in any insulated container such as a Styrofoam box or an ice bucket. (A) A layer of dry ice 1-2 cm deep covers the bottom of the QFS chamber, here a styrofoam box. (B) A heater block with 13 mm holes is used to house samples during QFS. The temperature probe with digital data logger is placed in the heater block. (C) After cooling the heater block in liquid nitrogen the samples are placed in the block, the liquid nitrogen is poured out, and the entire assembly is placed in the QFS chamber with the dry ice. (D) The QFS chamber is filled with dry ice so that the samples are covered; there is no disadvantage to covering the entire heater block with dry ice. (E) The box is covered and then moved onto the rotary shaker in a fume hood where the samples are agitated throughout the QFS procedure.

Figure 4: Arabidopsis leaves prepared by HPF-QFS. Results obtained for samples prepared by HPF-QFS and imaged by TEM on a Zeiss Libra 200 HT FE MC. (A) A medium high magnification image showing a chloroplast (CH) with cell walls and cytoplasm of adjacent cells. The vacuoles (V) contain electron dense material. Scale bar = 0.5 μm. (B) A high magnification image of a chloroplast. Scale bar = 100 nm. (C) A high magnification image of the cell wall between adjacent cells. A branched plasmodesma is visible (arrow). Note the smooth plasma membrane pressed against the cell wall by the vacuole. The cytoplasm is densely packed with ribosomes. V is vacuole. The tonoplast is smooth and continuous. Scale bar = 200 nm. (D) A low magnification image showing areas of good fixation in the chloroplast (CH) near regions of ice damage (asterisk). Scale bar = 0.5 μm. (E) A high magnification view of a cell wall and microtubules. Scale bar = 100 nm. (F) Image showing good preservation of chloroplasts (CH) and mitochondria (M). Large black precipitate is from staining sections with lead citrate. Scale bar = 1 μm. (G) High magnification view of Golgi (GA). Scale bar = 200 nm. (H) Poorly preserved sample showing lack of details in chloroplasts and plasmolysis (black arrow). Scale bar = 1 μm.