Gas Chromatography (GC) is a core instrumental analysis technique of Analytical Chemistry and is arguably as ubiquitous as a hot plate or balance in a chemistry laboratory. GC instrumentation can be used for the preparation, identification, and quantitation of a multitude of chemical compounds and can be coupled to a variety of detectors, such as flame ionization detectors (FIDs), photo-ionization detectors (PIDs), thermal conductivity detectors (TCDs), electron capture detectors (ECDs), and mass spectrometers (MS), depending on the analytes, methodology, and application. Samples can be introduced through a standard split/splitless inlet when working with small sample solutions, specialized headspace analysis inlets, solid phase micro-extraction (SPME) syringes, or thermal desorption systems. GC-MS is often the standard technique used in validation and verification applications of alternative or emerging, detection techniques because of its utility, flexibility, and identification power with established chemical databases and libraries.1–7 GC and its related sampling and detecting components is ideal for routine chemical analysis and more specialized, challenging analytical applications.
An analytical application of increasing interest to military, homeland security, and commercial enterprises is trace explosive vapor detection, with detection including identification and quantitation. Trace explosive vapor detection is an unique analytical chemistry challenge because the analytes, such as 2,4,6-trinitrotoluene (TNT) and cyclotrimethylenetrinitramine (RDX) have physical properties that make them especially difficult to handle and separate using broader, more generic chemical analysis methodologies. The relatively low vapor pressures and sub parts-per-million by volume (ppmv) saturated vapor concentration, combined with relatively high sticking coefficients, necessitate special sampling protocols, instrumentation, and quantitation methods.8–12 A GC coupled to an Electron Capture Detector (ECD) or mass spectrometer (MS) is an effective method for quantitating explosive analytes, specifically dinitrotoluene (DNT), TNT, and RDX.6,13–17 GC-ECD is particularly useful for nitro-energetic compounds because of their relatively high electron affinity. The U.S. Environmental Protection Agency (EPA) has created standard methods for explosive analyte detection using GC-ECD and GC-MS, but these methods have focused on samples in solution, such as ground water, and not samples collected in the vapor phase.2,18–23 In order to detect explosive vapors, alternative sampling protocols must be used, such as vapor collection with sorbent-filled thermal desorption sample tubes, but quantitative detection remains difficult due to lack of vapor standards and calibration methods that do not account for sample tube and instrumentation losses.
Recently, quantitation methods using thermal desorption systems with a cooled inlet system (TDS-CIS), coupled to a GC-ECD have been developed for TNT and RDX vapors.24,25 The losses associated with the TDS-CIS-GC-ECD instrumentation for trace explosive vapors were characterized and accounted for in example calibration curves using a direct liquid deposition method onto sorbent-filled thermal desorption sample tubes. However, the literature focused on instrumentation characterization and method development but never actually sampled, analyzed, or quantitated explosive vapors, only solution standards. Herein, the focus is on the protocol for sampling and quantitating explosive vapors. The protocol and methodology can be expanded to other analytes and trace explosive vapors, such as Pentaerythritol tetranitrate (PETN).