Central venous catheter (CVC) tip position has to be confirmed since misplaced catheters can lead to trauma and erosion of endothelium and vessel walls with subsequent potentially fatal pericardial tamponade1. Furthermore, CVC in the upper superior vena cava (SVC) can provoke thrombotic events, especially in cancer patients. The same is true for misplaced CVC in opposite brachiocephalic vein or internal jugular vein (IJV)2.
Several methods have been established to confirm CVC tip position in the lower SVC. Chest radiograph has been performed traditionally as a bedside method to visualize the CVC tip and exclude pneumothorax. However, additionally to exposure to radiation and costs, the accuracy of chest radiograph for tip position and sensitivity for pneumothorax have been questioned, specifically for the patient in a supine position3. Computer tomography is highly accurate and sensitive for both CVC tip position and pneumothorax, respectively. However, the costs and exposure to radiation do not justify this method for the sole indication of CVC tip confirmation. Transesophageal echocardiography (TEE) is highly accurate for tip position, but indication for TEE is limited to patients with cardiac pathology4. Intra-cardiac electrocardiography (ECG) is an accurate method but depends on a p-wave cardiac rhythm5.
Parasternal and subcostal ultrasound view were suggested to verify the CVC tip position, but success depends on deep insertion of the guidewire into the right atrium. A negative result does not exclude misplaced catheters6,7. The right supraclavicular fossa ultrasound view has been solely described for diagnostic purposes in cardiac failure. In a pilot study, we have demonstrated that this view can also be safely used for venipuncture of the right IJV and accurate positioning of the CVC guidewire with subsequent CVC placement in real-time8. The right supraclavicular fossa ultrasound view can be established with a microconvex probe which has the advantage of a small footprint, deep pentetration and, yet a sufficient resolution for venipuncture when compared to a linear probe8-10.
The proposed procedure has the advantage of visualizing venipuncture and guidewire advancement with a single ultrasound probe within the sterile field in real-time. CVCs insertion length does not have to be corrected afterwards and extra manipulations and potential sources of contamination are prevented. In conjunction with lung ultrasound for exclusion of a pneumothorax, the proposed procedure has the potential to replace chest radiograph for CVC tip position confirmation and exclusion of pneumothorax, reducing costs and exposure to radiation and increasing patient comfort in any hospital setting where an ultrasound machine equipped with a microconvex ultrasound probe is available.