$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
GUV swelling
With the spontaneous swelling approach described here, GUVs composed of DOPG, eSM, and Chol were grown overnight in 210 mM sucrose or 100 mM NaCl, 10 mM Tris, pH 7.5 forming a visible aggregate. Harvesting the aggregate ensures high vesicle yields. The resuspension in the swelling solution resulted in symmetric trans-membrane solution conditions. To create asymmetric conditions, the aggregate was resuspended in an iso-osmolar sucrose or high-salinity solution, respectively (Figure 2). The resulting dilution corresponds to a quasi-external solution exchange while minimizing the dilution of the number of GUVs.
Phase diagram mapping of GUVs using fluorescence microscopy
The presence of 0.1 mol% of the phase-specific DiIC18 in the GUVs allowed for the observation of their phase states via wide-field fluorescence microscopy. Vesicles exhibiting S+L phase separation were observed through a 63x/1.2NA to be able to resolve finely structured finger-like domains. For all remaining cases, a 40x/0.6 NA objective was used. To avoid artifacts during the visual inspection of GUVs and to maximize reproducibility, certain criteria were set that determined which vesicles to consider for the phase state analysis (protocol step 3.6).
GUVs prepared from ternary DOPG/eSM/Chol mixtures of a broad range of ratios swollen in sucrose or high-salinity solution and observed at room temperature exhibited homogenous Lo or Ld phases, Lo+Ld, and S+L phase separation, see Figure 3. Figure 9 shows exemplarily defective vesicles, which should not be included in the data analysis and also shows how to identify multilamellar vesicles.
Due to their unknown history, GUVs are likely to exhibit within-batch compositional variation35. Hence, the overall phase state of a particular composition was determined in a statistical approach. GUV populations of a certain lipid composition were ascribed to the phase state that was observed to be dominant in a random sample (protocol step 3.6). Yet, compositions close to the Lo+Ld coexistence region often yielded batches where the dominant phase state made up a narrow majority. For such vague cases, the visual inspection was repeated with at least three independent samples. The fraction of vesicles with identical phase state (e.g. exhibiting Lo+Ld phase separation) present within the random samples was averaged over the number of trials which was taken as the final result (Figure 10).
The described protocols resulted in sufficient GUV growth over a broad range of different ratios of DOPG, eSM, and Chol in high-salinity and sucrose solutions. Extensive regions within the ternary phase diagram could be mapped under symmetric as well as asymmetric high-salinity solution conditions (Figure 11). The differences in the vesicle phase behavior observed under different solution conditions were discussed elsewhere in detail9.
Observations of phase behavior after complete buffer exchange using the microfluidic method
The creation of asymmetric solution conditions by dilution results in residues of the swelling solution outside. Our microfluidic approach allows for a complete external solution exchange. Figure 5A shows the microfluidic device fully assembled on the confocal microscope stage together with fluidic outlet and pressure control inlets. Tubes are connected via 90 ° metal pipes to allow space for transmitted light imaging from above.
For observation within the microfluidic device, a confocal microscope with a 63x/1.2NA water immersion objective lens was implemented. As with the previous observations in bulk, DiIC18 was used to stain the membrane. When making observations of the phase state of GUVs trapped in the device, care must be taken not to misinterpret the data. Due to the close proximity of the GUVs to the posts, the excitation and emission light paths may be partially blocked by the PDMS leading to the false appearance of domains (Figure 12). Here, the transmitted light detection is useful to check for the position of the GUVs at the posts. This unwanted effect is particularly prominent for small GUVs of less than 10 µm in diameter. In these cases, the data were rejected. The confocal image in Figure 13A shows a planar vesicle section that crosses an Lo domain present on the GUV vesicle. In this case, planar cross-section scans further above or below the section would have not showed the domain due to its small size compared to that of the GUV, which is why it would have been missed and the vesicle considered to be in a homogeneous phase state. Hence, a confocal z-stack should be used to inspect the entire GUV surface. For wide-field microscopy, this might not be an issue because the whole vesicle can be imaged at once.
Finally, examples are given of GUVs before and after exchange of the outer solution where it is clear what the phase states of the membranes are (Figure 14). Each device has 60 chambers (each with a pair of posts to trap a single GUV), allowing tens of experiments per device (see Figure 4). However, some GUVs may get lost during the external solution exchange. This can be minimized by the following steps: 1) Using bovine serum albumin (BSA) coating to prevent GUV adhesion/rupture at the posts; coating is done by exposing the chamber walls to 20 mg/mL BSA for 60 min and subsequent rinsing with the working buffer. Another adhesive molecule that could be used is poly(L-lysine)-graft-poly(ethylene glycol)39. 2) Careful osmotic matching of the inner and outer solutions to avoid flaccid GUVs, which could pass through the center of the posts or bursting of GUVs. 3) Optimizing the GUV preparation procedure to obtain vesicles bigger than ~8 µm in diameter to prevent passage through the center of the posts.
Design and characterization of temperature-controlled chamber for observation of GUV phase states
We designed a simple flow chamber, which features connections to the external water thermostat (Figure 6). We obtained the chamber by milling of an aluminum block. In general, the chamber material does not need to be heat conductive, as the sample is coupled to the water bath by the lower cover glass as shown in Figure 6B and 6C.
Independently of the exact design, the performance of the chamber should be evaluated. Specifically we checked for the linearity of the sample temperature with the externally-set temperature, any systematic temperature offset, and a temperature gradient within the chamber. The first two points were addressed by direct temperature measurement within the chamber by a fiber optic temperature probe (e.g. FISO FTI-10). We also checked for a temperature gradient within the GUV suspension. Such a gradient could stem from heat flow to the outside of the chamber. Spatially resolved temperature measurements can be obtained by a temperature sensitive fluorophore40, see Figure 7.
Data plotting and fitting to obtain Tmix of phase-separated GUVs
Plotting the fraction of homogenous GUVs over the observed temperature range resulted in a sigmoidally shaped data point trajectory. We fit the data to the Boltzmann model (protocol section 6.8) from which the Tmix could be deduced (Figure 8).

Figure 1: Experimental steps during the spontaneous swelling protocol (top panel) with the corresponding stage of vesicle growth (lower panel). (A) A homogenous lipid film is spread onto a roughened PTFE plate and dried from any solvent. (B) The dried lipid film is then pre-swollen in a water-saturated atmosphere inside a closed container with water to facilitate the bilayer hydration. The void glass vial contains the PTFE plate and remains open during this hydration step. (C) The pre-swollen lipid film finally becomes fully hydrated by the addition of the desired swelling solution onto the lipid-coated PTFE plate inside the glass vial. To avoid evaporation, the glass vial is sealed properly during the overnight incubation. To minimize compositional variations within a batch, all steps need to be performed at a temperature where the lipid mixture is fully miscible. Please click here to view a larger version of this figure.

Figure 2: GUV harvesting. (A) A deposited lipid film consisting of DOPG/eSM/Chol at molar ratios of 20/60/20 with 0.1 mol% DiIC18 was swollen in high-salinity buffer composed of 100 mM NaCl, 10 mM Tris, pH 7.5. The lid of the glass container was additionally sealed with paraffin film. The magnified region of interest contains the resulting GUV aggregate. Its red appearance is a consequence of the presence of DiIC18. (B) The GUVs were harvested with a truncated pipette tip. In this image, the aggregate was pipetted up together with 50 µL of swelling solution, which were transferred into a fresh vial. (C) To create asymmetric trans-membrane solution conditions, the aggregate was diluted in another isotonic external solution. Here, the aggregate together with the 50 µL swelling solution was resuspended in 950 µL sucrose solution resulting in a 20x dilution of the swelling solution. Please click here to view a larger version of this figure.

Figure 3: GUV imaging. Wide field fluorescence images of GUVs doped with 0.1 mol% DiIC18 prepared and observed in symmetric salt or sucrose solutions consisting of different ratios of DOPG, eSM, and Chol (in salt buffer, from left to right: 40/20/40; 50/20/30; 30/60/10; in sucrose, from left to right: 30/30/40; 20/60/20; 40/50/10). The images depict GUVs in different (coexisting) phase states assessed according to the criteria in protocol step 3.6. Here, vesicles were imaged through a 40X/0.6 NA objective. Scale bars = 5 µm. Please click here to view a larger version of this figure.

Figure 4: Layout of microfluidic channel design. GUVs enter the lower fluidic layer (outlined channels) via the inlet below a reservoir. A filter blocks unwanted debris from the device. They then enter an array of 60 chambers (8 rows and 15 columns) each containing posts for single GUV capture (see insert). Each chamber can be isolated within a ring valve actuated by a control layer (filled black channels) above the fluidic layer. Please click here to view a larger version of this figure.

Figure 5: Microfluidic device. (A) Photograph of a microfluidic device used to trap single GUVs and fully exchange the outer solution. Labels indicate the reservoir to add solutions (1), the 8 x pressure inlets connected to the pressure control unit (2), and the fluidic outlet connected to the syringe and pump (3). Panel (B) shows the pressure control unit featuring 8 valves each regulating 1 tube connected to the microfluidic device. Here valves #1 and #2 are open and hence the corresponding ring valves are closed. Please click here to view a larger version of this figure.

Figure 6: Temperature control chamber. (A) Chamber before the assembly. (B) Assembled chamber (facing upwards) with 2 mm cover glasses glued to the top and bottom. The orange rubber spacer measures 0.5 mm in height and is sealed with a 0.17 mm cover slip for observation of the enclosed GUV suspension. In this image a temperature probe is inserted for calibration purposes (brown fiber exiting the chamber on the right). (C) Final assembly on the stage of an inverted microscope without the temperature probe. The orange rubber spacer is now facing downwards. The rubber is adhesive enough to hold the sample in place. Note that light can be transmitted through the sample enabling both epi-fluorescence and bright field observations. Please click here to view a larger version of this figure.

Figure 7: Spatially resolved temperature data inside the observation chamber obtained by FLIM measurements of a temperature sensitive dye (here: 500 µM Rhodamine B)40. Data points obtained at 0, 10, 30, 300, and 400 µm above the bottom cover slip. The red and blue data points were measured for temperature of the water bath set at 30 °C and 12 °C, respectively. The black data points indicate the water bath left to equilibrate to room temperature. Small temperature variations throughout the chamber were observed but stayed below 0.5 °C (gray bars). The total chamber height is about 500 µm according to the spacer thickness (see above). The error bars indicate standard deviations. Please click here to view a larger version of this figure.

Figure 8: Locating the miscibility temperature. The graph shows individual data points of homogeneous (single-liquid state) the fraction of GUVs prepared from DOPG/eSM/Chol in a ratio 30/40/30 doped with 0.1 mol% DiIC18 from three independent random samples (N = 20 - 40). Error bars represent standard error of means. Data points were fitted to the Boltzmann model described in step 6.8 (continuous black line) from which the domain mixing temperature Tmix was deduced (follow red continuous line to abscissa) according to the half maximum of the sigmoidal curve on the ordinate (dashed black line). Initial and final values (A1 and A2) were fixed to 0 and 1, respectively. Please click here to view a larger version of this figure.

Figure 9: Defective vesicles. Examples of giant vesicles prepared from 20/60/20 DOPG/eSM/Chol and doped with 0.1 mol% if DiIC18 that do not meet the criteria set in step 4.2 imaged by wide-field fluorescence microscopy. Intensity display ranges of individual images were optimized for the fluorescence intensity of the depicted vesicle each time. Due to the presence of additional membrane material and encapsulated smaller vesicles, the phase state of the vesicle in (A) cannot be clearly defined. The appearance of this giant vesicle does not allow for any reliable visual inspection for lamellarity. Panel (B) depicts a vesicle of which the interior is crowded with membrane material. As a consequence, the fluorescence signal of the exterior vesicle membrane is superimposed with its internal signal, rendering a visualization of lamellarity and potential domains impossible. (C) The fluorescence intensities of three different giant vesicles are shown in direct comparison within the same display intensity range. This image shows that giant multilamellar vesicles (1, 2) can be identified by their increased fluorescence signal compared to the GUV (3). Here, multilamellar as well as unilamellar giant vesicles exhibit Lo+Ld phase separation. Vesicles in all images were imaged through a 40x/0.6 NA objective. Scale bars = 5 µm. Please click here to view a larger version of this figure.

Figure 10: Fraction of Lo+Ld phase separated GUVs averaged over at least three independent samples. Vesicles were composed of 30/40/30 DOPG/eSM/Chol close to the border of the Lo+Ld coexistence region. The error bars for symmetric sucrose conditions illustrate the batch-to-batch scattering. The label of the ordinate describes solutions inside/outside the vesicles; sucrose: 210 mM sucrose; salt: 100 mM NaCl, 10 mM Tris, pH 7.5; Error bars indicate standard deviations. Please click here to view a larger version of this figure.

Figure 11: Phase diagram of GUVs prepared from DOPG, eSM, and Chol mapped under different solution conditions using 210 mM sucrose (sucrose) and 100 mM NaCl, 10 mM Tris, pH 7.5 (salt). GUV phase states were explored in sucrose/sucrose (A; upper section), sucrose/salt (in/out) (A; lower polygonal section), salt/salt (B; upper section), and salt/sucrose (B; lower polygonal section). The cartoons illustrate the dominant domain pattern within the highlighted sections and the corresponding solution conditions. Vesicle phase states represented in the lower polygonal sections were observed under asymmetric solution conditions upon 20x GUV dilution. Adapted from reference9. Please click here to view a larger version of this figure.

Figure 12: Artifacts in trapped GUVs. Example of a small GUV where a false domain can be seen due to the close proximity with the posts (DOPG/eSM/Chol 60/20/20). The bright-field transmitted light image showing the posts (grey) is overlaid with the confocal fluorescence image of the GUV (orange). The posts are seen to create an interference pattern in the bright-field image. The fluorescence signal close the posts (right) is reduced giving the appearance of phase separation. Scale bar = 5 µm.

Figure 13: Examples of two different GUVs captured by the PDMS posts where the phase states are clearly visible. (A) Lo + Ld phase separated vesicle made of DOPG/eSM/Chol 60/20/20. (B) A vesicle made of 40/30/30 exhibiting Ld or Lo single phase. A confocal z-stack of the whole vesicle was examined to confirm that no (out-of-focus) domains were present. The edges of the posts are seen on the right hand side of the images (grey). Scale bar = 5 µm.

Figure 14: Resulting phase behavior after a complete fluidic exchange using the microfluidic device. The same GUV is shown both before and after solution exchange for (A) symmetric salt/salt (in/out) to salt/sucrose (in/out) (DOPG/eSM/Chol 60/20/20, scale bar is 2 µm) and (B) symmetric sucrose/sucrose (in/out) to sucrose/salt (in/out) (DOPG/eSM/Chol 30/40/30, scale bar = 3 µm). Adapted from reference9.