Method Article

Quantitative and Qualitative Method for Sphingomyelin by LC-MS Using Two Stable Isotopically Labeled Sphingomyelin Species

DOI:

10.3791/57293

May 7th, 2018

In This Article

Summary

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Here, we present a protocol to quantify and qualify each sphingomyelin species using multiple reaction monitoring and MS/MS/MS mode, respectively.

Abstract

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

We present a method of analyzing sphingomyelin (SM) qualitatively and quantitatively by liquid chromatography-electrospray Ionization-tandem mass spectrometry (LC-ESI-MS/MS). SM is a common sphingolipid composed of a phosphorylcholine and a ceramide as the hydrophilic and hydrophobic component, respectively. A number of SM species are present in mammalian cells due to a variety in the sphingoid long chain base (LCB) and an N-acyl moiety in the ceramide. In this report, we show a method of estimating the number of carbon and double bonds in a LCB and an N-acyl moiety based on their corresponding product ions in MS/MS/MS (MS3) experiments. In addition, we present a quantitative analysis method for SM using two stable isotopically labeled SM species, which facilitates determining the range used in SM quantitation. The present method will be useful in characterizing a variety of SM species in biological samples and industrial products such as cosmetics.

Introduction

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Sphingomyelin (SM) is a common sphingolipid in mammalian cells. SM is synthesized intracellularly1 and present as a precursor for other sphingolipids such as a sphingosine-1-phosphate and a ceramide, which have crucial roles in immune cell trafficking and skin barrier homeostasis, respectively2,3. Thus, the precise analysis of the SM metabolism is important for elucidating the physiological and pathological roles of sphingolipids.

SM is composed of a ceramide and a phosphorylcholine that is linked to the 1-hydroxy group of the ceramide, which is further composed of a sphingosine and an N-acyl moiety. A variety in the carbon and double bond numbers in both the sphingosine and N-acyl moiety results in the number of ceramide (and SM) species. Recent advances in LC-ESI-MS/MS has enabled quantitative and qualitative analysis of SM4,5. In the qualitative analysis, the number of carbon and double bonds of a sphingoid LCB of SM was identified by assigning product ion spectra of LCB. However, structural information of the N-acyl moiety was not directly obtained because its corresponding product ions have not been reported, and therefore N-acyl moieties were deduced by differential analysis between precursor ions and product ions corresponding to LCB in both positive and negative ion modes4,5. In this report, we present a method to detect the product ions of both LCB and N-acyl moiety simultaneously in MS3 mode using triple quadrupole and quadrupole linear ion trap mass spectrometry, which facilitates the precise structural speculation of each SM species6.

The ion suppression (or enhancement) effects caused by the matrix in biological samples hamper the accurate quantification in LC-ESI-MS analysis, and therefore, it is desirable to construct calibration curves for all analytes of interest in the identical matrix of the biological sample. However, this strategy is not feasible because it is almost impossible to prepare all SM species in biological samples, especially in comprehensive analysis. Thus, it is practical to construct a calibration curve and determine the quantitative range using a representative SM species spiked in the biological samples. We used two isotopically-labeled SM species to construct a calibration curve; one was used for an internal standard and the other for a standard compound. We detected a small amount of isotopically-labeled SM species as a standard compound spiked in biological samples and successfully obtained a calibration curve and the quantitative range6.

Access restricted. Please log in or start a trial to view this content.

Protocol

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Consult all relevant material safety data sheets (MSDS) before use. Wear gloves to minimize sample contamination by skin-derived SM. The present protocol was applied to HeLa cells grown in Eagle's minimum essential medium supplemented with 10% fetal bovine serum (FBS), 2 mM L-glutamine, 1,000 U/L penicillin, and 100 mg/L streptomycin.

1. Preparation of Lipid Samples

NOTE: It is important that all glassware including test tubes with Teflon-lined screw caps be detergent-free.

  1. Extraction of total lipid fraction from cell homogenate using the Bligh & Dyer method7
    1. Remove the culture (or conditioned) media from the 10-cm tissue culture dish and rinse twice with 6 mL of ice-cold PBS.
    2. Add 1 mL of ice-cold PBS into the 10-cm tissue culture dish by P1000 pipette. Harvest the cells with cell scrapers and collect into 2.0-mL siliconized plastic tubes.
    3. After centrifugation (1,000 × g, 5 min, 4 °C), remove the supernatant by a P1000 pipette.
    4. Add 1 mL of methanol. Vortex tubes briefly and sonicate at 200 W for 5 min in a bath-type sonicator.
      NOTE: Adjust the water level in a bath-type sonicator so that the cell pellet is efficiently homogenized.
    5. Transfer the cell homogenate in methanol by pipette into test tubes (13 mm x 100 mm) with Teflon-lined screw caps.
    6. Add 1 mL of methanol, 1 mL of chloroform, 0.8 mL of double distilled water, and 50 µL of 10 µmol/L of internal standard d18:1/(D31)-16:0 SM into each test tube.
    7. Vortex test tubes vigorously for 5 min at room temperature.
      NOTE: At this point, a single phase (chloroform/methanol/water = 1/2/0.8 (v/v/v)) is formed.
    8. Add 1 mL of chloroform and 1.0 mL of double distilled water.
    9. Vortex tubes vigorously at 2,500 rpm for 5 min at room temperature.
      NOTE: At this point, the aqueous (upper) phase and organic (lower) phase are separated.
    10. After centrifugation (2,150 × g, 5 min, 25 °C), collect and transfer the lower phase into disposable glass tubes (initial lower phase).
      ​NOTE: Collect the lower phase using a Pasteur pipette and a safety pipette filter. Insert the tip of the pipette into the lower phase, squeeze the small bulb next to the 'E' valve to deliver the upper phase and interfacial fluff inside the tip of the pipette, and then siphon the lower phase into the pipette.
    11. Add 2 mL of chloroform into the test tubes, and vortex the tubes vigorously at 2,500 rpm for 5 min at room temperature to sufficiently mix with the upper phase and interfacial fluff.
    12. After centrifugation (2,150 × g, 5 min, 25 °C), collect and transfer the modified lower phase into the disposable glass tubes with the initial lower phase as described in step 1.1.10.
    13. Place the glass tubes under a nitrogen stream and completely remove the organic solvents in the collected lower phase.
    14. Reconstitute the samples with 500 µL of methanol or ethanol, filtrate with a 0.02-µm filter, and store in glass vials at -20 °C.
  2. Sample preparation for constructing the calibration curve and validating the method
    1. Add 50 µL of 0.1, 0.5, 1, 5, 10, or 50 µmol/L of standard compound (d18:1/(D9)-18:1 SM) into test tubes with Teflon-lined screw caps.
    2. Add cell homogenate into each test tube and add methanol to 1 mL.
      NOTE: The amount of cell homogenate as a matrix varies according to each experiment. If the SM species in samples derived from cells in a 10-cm culture dish are routinely analyzed, the amount of cell homogenate in each test tube should be close to that of cells in a 10-cm culture dish.
    3. Extract the total lipid fraction as described in steps 1.1.6-1.1.14.
    4. Prepare quality check (QC) samples for validating the method as descried in steps 1.2.1-1.2.3. Prepare three QC samples with different concentrations of standard compound (d18:1/(D9)-18:1 SM): one within 3x the lower boundary of the standard curve (QC-low, QC-L), one near the center (QC-middle, QC-M), and one near the upper boundary of the standard curve (QC-high, QC-H).

2. SM Analysis by LC-ESI-MS/MS

  1. Preparation of the mobile phase
    1. Mix the solvents (acetonitrile/methanol/ddH2O = 2/2/1 (v/v/v) for the aqueous phase and isopropanol for the organic phase) in glass bottles with Teflon-lined screw caps and sonicate for 5 min in a bath-type sonicator.
    2. Add formic acid (final concentration 26.4 mmol/L) and NH4OH (14.9 mmol/L) into each mobile phase.
  2. Qualitative analysis of SM by LC-ESI-MS3
    1. Activate the high performance liquid chromatography (HPLC) system, put inlet tubes into the glass bottles that contain mobile phases, and purge the HPLC lines. Link a C18 HPLC column (1.5 mm i.d. × 100 mm length, particle size 3.0 µm) to the HPLC system, keep the temperature in the column oven at 50 °C, and condition the column with the aqueous mobile phase at 100 µL/min. Put the samples into a sample rack in the autosampler.
    2. Set the parameters of the triple quadrupole and quadrupole linear ion trap mass spectrometry system for MS3 analysis as listed in Table 1 and Table 2 as well as the first and second precursor ions of the SM species of interest.
      1. Double click the software icon for data acquisition, double click the 'Hardware Configuration', select 'LC+QTRAP4500+Valve', and click 'Activate Profile'.
      2. Create a new subfolder by clicking 'New Sub-Project' and 'OK'.
      3. Click the 'file' tab, select 'New', and select 'Acquisition Method' and 'OK'. Click the 'Mass Spec' icon and 'MS' tab. Select Scan type as 'MS/MS/MS (MS3)', Scan rate as 10,000 Da/s, Polarity as 'Negative', and 'Duration' as entire time to be analyzed (min). Set the m/z of the 'Start (Da)' and 'Stop (Da)' as the range to be scanned. Set the m/z of the 1st and 2nd precursor ions of the target SM species of interest.
        NOTE: To analyze multiple SM species, highlight the '-MS3' icon, right click, select 'Copy this experiment', and then put the m/z of the 1st and 2nd precursor ions of other SM species.
      4. Select the 'Advanced MS' tab, and set each parameter as follows: Scan mode as 'Profile', Step size as 0.12 (Da), Resolution Q1 and Q3 'Unit' and 'LIT', respectively, Intensity threshold as 0, Setting time as 50 (ms), Pause between mass ranges as 1.5 ms, select 'Dynamic fill time', Q3 Entry Barrier as 8 V, and Excitation Time as 25 ms.
      5. Click the 'Edit Parameters' button in the 'MS' tab and select the 'Source/Gas' tab. Set the parameters of the curtain gas, collision gas, ionspray voltage, temperature, ion source gas1 and gas2 as listed in Table 1. Then select the 'Compound' tab. Set the parameters of the declustering potential, entrance potential, collision energy, excitation energy, and collision energy spread as listed in Table 2.
      6. Highlight 'Integrated Valco Valve' and select 'Position Name for Step 0' as A, and set 'B' as position for total time 5.0 min. Also set 'A' as position for total time 70.0 min.
      7. Highlight 'Shimadzu LC system'. Select the 'Pump' tab and set Pumping mode as binary flow, total flow as 0.28 mL/min, and maximum of pressure limits as 20.0 MPa. Select the 'Autosampler' tab and set the rinsing volume as 200 µL, needle Stroke as 52 mm, rinsing speed as 35 µL/s, sampling Speed as 5.0 µL/s, purge time as 25.0 min, rinse dip time as 5 s, and rinse mode as 'Before and after aspiration'.
        1. In addition, enable the cooler unit, set the cooler temperature as 4 °C and the control vial needle stroke as 52 mm. Select the 'Oven' tab, enable the oven and set the oven temperature and maximum temperature as 50 °C and 85 °C, respectively.
        2. Select the 'Controller' tab and check the box 'Power on'. Select the 'Time Program' tab and set the solvent gradients as follows: Solvents A/B at a 100/0 ratio for 5 min, program linear alterations to 80/20 over 4 min, to 35/65 over 50 min, and to 25/75 over 1 min. Afterwards, hold them at 25/75 for 10 min, then linearly to 100/0 over 4 min. Then save the method.
    3. Create batch file
      1. Double click the 'Build Acquisition Batch' icon, click 'Add Set', 'Add Samples', set the number of new samples, and click 'OK'.
      2. Click the 'Method Editor' button and select the method to be used. If multiple methods are used in one batch file, check the box of 'Use Multiple Methods' and select the acquisition method for each sample. Rename 'Sample Name', set the appropriate number for 'Vial Position', and put the amount of injection volume. Then save the batch file.
    4. Obtain the MS3 product ion spectra of the SM species of interest by LC-ESI-MS3 analysis
      1. Select the 'Submit' tab in batch file, highlight the line of samples to be analyzed, and click the 'Submit' button.
      2. Click the ''View Quene' icon, 'Equilibrate' icon, select the acquisition method to be used, set Time as 1 min, and click 'OK'.
      3. Deactivate 'Reserve Instrument for Tuning' by clicking the corresponding icon. Then execute batch sequence by clicking the 'Start Sample' icon.
    5. Assign each MS3 product ion spectra of the SM by comparing the mass to charge ratio (m/z) of the product ion spectra and the exact mass of the sphingoid LCB and N-acyl moiety of interest. Determine the number of carbons and double bonds in the sphingoid LCB and N-acyl moiety according to the corresponding product ions.
      1. Confirm that the appropriate sub-project folder is selected, then double click the 'Open Data File' icon, and select the samples to be analyzed. Drag the peak in chromatogram for each target SM species and then double click. The obtained MS3 product ion spectra within the dragged area will be presented.
  3. Quantitative analysis of the SM by LC-ESI-MS/MS
    1. Set the mobile phases and HPLC column as described in step 2.1 and 2.2.1.
    2. Set the parameters of the triple quadrupole and quadrupole linear ion trap mass spectrometry system for multiple reaction monitoring (MRM) analysis as listed in Table 1 and Table 2.
      1. Conduct the procedures as described in step 2.2.2.1 and 2.2.2.2.
      2. Click the 'file' tab, select 'New', and select 'Acquisition Method' and 'OK'. Click the 'Mass Spec' icon and 'MS' tab. Select Scan type as 'MRM', Polarity as 'Positive'. Set 'Duration' as the entire time to be analyzed. Set the m/z of [M+H]+ and 184 as 'Q1 Mass (Da)' and 'Q3 Mass (Da)' of the target SM species of interest, respectively. Set 'Time' as 10 ms and 'ID' as the name of the target SM species.
      3. Select the 'Advanced MS' tab, and set each parameter as follows: Both of the resolution Q1 and Q3 as 'Unit', Intensity threshold as 0, Setting time as 0 (ms), and Pause between mass ranges as 5 ms.
      4. Click the 'Edit Parameters' button in the 'MS' tab and select the 'Source/Gas' tab. Put the parameter of curtain gas, collision gas, ionspray voltage, temperature, ion source gas1 and gas2 as listed in Table 1. Then select the 'Compound' tab. Put the parameters of declustering potential, entrance potential, collision energy, and collision cell exit potential as listed in Table 2.
      5. Set the valve as described in step 2.2.2.6.
      6. Set the LC condition as described in step 2.2.2.7. Then save the method.
    3. Create the batch file as described in step 2.2.3.
    4. Obtain the MRM data of each SM species by LC-ESI-MS/MS analysis as described in step 2.2.4.
    5. Process the MRM data using the software for data integration and obtain the data of peak area for the extracted ion chromatogram of each SM species.
      1. Double click the software icon for data integration in quantitative analysis, click the 'Edit' tab and select User Integration Defaults. Set Gaussian Smooth Width as 1.0 point, and click 'OK'. Click the 'Edit' tab and select 'New Results Table'. Select the sample to be integrated, click an arrow button, and click 'Next'. Select 'Create New Method', set the name of the method, and click 'Next'. Click 'Next' and check the box of d18:1/(D31)-16:0 SM as IS, click 'Next', and click 'Finish'.
      2. Click 'Displays the peak review' and confirm that the chromatogram peaks are appropriately recognized. It should be noted that the retention time of d18:1/(D31)-16:0 SM is usually smaller by ~ 0.3 min as compared with the one of the 34-1 SM (usually consists of d18:1/16:0 SM).
    6. Quantify each SM species according to the ratio of the peak of each SM species to the d18:1/(D31)-16:0 SM internal standard.
      1. Click the 'File' tab, select 'Export', select 'Results Table', confirm the format as 'MultiQuant', columns as 'Export all columns', rows as 'Export all rows except those explicitly hidden', and then click 'OK'.
      2. Open the exported file in the Excel software. Normalize the area of the target SM species by the area of IS (d18:1/(D31)-16:0 SM). Then multiply by the theoretical amount of the IS in the injected sample to calculate the amount of target SM species in the injected sample.
  4. Constructing the standard curve
    1. Obtain the MRM data of d18:1/(D9)-18:1 SM and d18:1/(D31)-16:0 SM in the samples for constructing the standard curve by LC-ESI-MS/MS analysis as described in step 2.3.1-2.3.4.
    2. Obtain the data of the peak area of d18:1/(D9)-18:1 SM and d18:1/(D31)-16:0 SM as described in step 2.3.5.
    3. Calculate the amount of d18:1/(D9)-18:1 SM in the injected sample as described in step 2.3.6.
    4. Construct the trendline by setting the X-axis and Y-axis as the nominal amount and calculated amount of d18:1/(D9)-18:1 SM and obtain the trendline formula.
  5. Validating the quantitative method using Excel software
    1. For validation of the method, quantify the amount of d18:1/(D9)-18:1 SM and d18:1/(D31)-16:0 SM in the QC samples by analyzing each QC sample at least three times in 1 day, and repeat at least 3 days as described in step 2.3.1-2.3.4.
    2. Obtain the peak area data and calculate the amount of d18:1/(D9)-18:1 SM in the injected sample as described in step 2.3.5 and 2.3.6.
    3. According to the calibration curve obtained, calculate the amount of d18:1/(D9)-18:1 SM in the injected sample.
    4. Evaluation of precision
      1. Calculate the average and the standard deviation of the amount of d18:1/(D9)-18:1 SM obtained in step 2.5.3 on the dataset of the intra-day and inter-day using Excel software.
      2. The average obtained in step 2.5.4.1 is divided by the standard deviation, and expressed as %.
    5. Evaluation of accuracy
      1. Calculate the accuracy of each data as follows:
        [(The calculated amount obtained in step 2.5.3.) / (Nominal amount) - 1] × 100(%)
      2. Calculate the average of the absolute value of the accuracy on the dataset of the intra-day and inter-day.

Access restricted. Please log in or start a trial to view this content.

Results

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

Chemically synthesized d18:1/24:0 SM (Figure 1A) and d18:1/24:0 SM in lipid samples extracted from HeLa cells (Figure 1B) were analyzed by LC-ESI-MS3 employing [M+HCOO]- and [M-CH3]- as first and second precursor ions, respectively. Note that the spectrum intensity of demethylated-sphingosylphosphorylcholine (SPC) (m/z 449) is larger than that ...

Access restricted. Please log in or start a trial to view this content.

Discussion

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

In the present qualitative method, we obtained MS3 product ions of a SPC and an N-acyl moiety. It is critical to properly assign both a SPC and an N-acyl moiety. To this end, it should be noted that other phosphorylcholine-containing molecules can also be detected as MS3 product ions. Diacyl-phosphatidylcholine (PC) and plasmalogen-PC are abundantly present in mammalian cells, and their hydrophobicity is similar to that of SM. Therefore, diacyl-PC and plasmalogen-PC with an isotope...

Access restricted. Please log in or start a trial to view this content.

Disclosures

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

The authors declare that they have no conflict of interest.

Acknowledgements

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,

This work was supported by a research grant from the Ministry of Education, Culture, Sports, Science and Technology of Japan (KAKENHI) to K.H. (#15K01691), Y.F. (#15K08625), K.Y. (#26461532), and a grant for the study of Intractable Disease Project from Ministry of Health, Labour and Welfare (K.Y. #201510032A). We thank the Edanz Group (www.edanzediting.com/ac) for editing a draft of this manuscript.

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
PBSThermoFisher10010023
100 mm tissue culture dishIWAKI3020-100
Cell scraperIWAKI9000-220
Siliconized 2.0 mL tubeFisher Scientific02-681-321
Test tubeIWAKITST SCR 16-100
Teflon-lined screw capIWAKI9998CAP415-15
Disposable glass tubeIWAKI9832-1310
CAPCELL PAK C18 ACR 3 µm 1.5 mm I.D. x 100 mmShiseido92223Guard cartridge is inserted into cartridge holder, and linked to C18 column
CAPCELL C18 MGII S-3 2.0 mm x 10 mm GUARD CARTRIDGEShiseido12197
Cartridge holderShiseido12415
AcetonitrileWako018-19853
2-Propanol (Isopropyl Alcohol)Wako161-09163
MethanolWako134-14523
Formic acidWako066-00466
28% Ammonia waterWako016-03146
Sonicator (bath type)SHARPUT-206H
Vortex mixer for glass test tubeTAITECMix-EVR
1.4 mL glass vialTomsic500-1982Samples are stored in 1.4 mL glass vial sealed with screw cap and 8 mm septum at -20°C
8 mm septumTomsic200-3322When samples are analyzed, screw caps are replaced with screw caps with slit septum
Screw cap for 1.4 mm glass vialTomsic500-2762
Screw cap with slit septumShimadzu GLCGLCTV-803
PVDF 0.22 µm filterMilliporeSLGVR04NL
Triple quadrupole and quadrupole linear ion trap mass spectrometrySCIEXQTRAP4500
The software for data acquisition and analysis of product ion spectraSCIEXAnalyst
The software for data integration in quantitative analysisSCIEXMultiQuant
HPLC systemShimadzuNexera
Glass bottleSansyo85-0002
d18:1/24:0 sphingomyelinAvanti Polar Lipids860592P
SphingosylphosphorylcholineMerck567735
Fetal bovine serumThermoFisher 26140079
L-glutamineThermoFisher 25030081
Penicillin and streptomycinSigmaP4333
Eagle’s minimum essential mediumSigmaM4655

References

Loading...
$$\rightleftharpoonup{xx}$$ $$\longleftharp{xx}$$, $$\longrightharp{xx}$$,
  1. Huitema, K., van den Dikkenberg, J., Brouwers, J. F., Holthuis, J. C. Identification of a family of animal sphingomyelin synthases. EMBO J. 23 (1), 33-44 (2004).
  2. Kihara, Y., Mizuno, H., Chun, J. Lysophospholipid receptors in drug discovery. Exp Cell Res. 333 (2), 171-177 (2015).
  3. Kihara, A. Synthesis and degradation pathways, functions, and pathology of ceramides and epidermal acylceramides. Prog Lipid Res. 63, 50-69 (2016).
  4. Merrill, A. H., Sullards, M. C., Allegood, J. C., Kelly, S., Wang, E. Sphingolipidomics: high-throughput, structure-specific, and quantitative analysis of sphingolipids by liquid chromatography tandem mass spectrometry. Methods. 36 (2), 207-224 (2005).
  5. Houjou, T., et al. Rapid and selective identification of molecular species in phosphatidylcholine and sphingomyelin by conditional neutral loss scanning and MS3. Rapid Commun Mass Spectrom. 18 (24), 3123-3130 (2004).
  6. Hama, K., Fujiwara, Y., Tabata, H., Takahashi, H., Yokoyama, K. Comprehensive Quantitation Using Two Stable Isotopically Labeled Species and Direct Detection of N-Acyl Moiety of Sphingomyelin. Lipids. , (2017).
  7. Bligh, E. G., Dyer, W. J. A rapid method of total lipid extraction and purification. Can. J. Biochem. Physiol. 37, 911-917 (1959).
  8. Gu, H., Liu, G., Wang, J., Aubry, A. F., Arnold, M. E. Selecting the correct weighting factors for linear and quadratic calibration curves with least-squares regression algorithm in bioanalytical LC-MS/MS assays and impacts of using incorrect weighting factors on curve stability, data quality, and assay performance. Anal Chem. 86 (18), 8959-8966 (2014).
  9. Folch, J., Lees, M., Sloane Stanley, G. H. A simple method for the isolation and purification of total lipides from animal tissues. J Biol Chem. 226 (1), 497-509 (1957).
  10. Thomas, M. C., et al. Ozone-induced dissociation: elucidation of double bond position within mass-selected lipid ions. Anal Chem. 80 (1), 303-311 (2008).
  11. Baba, T., Campbell, J. L., Le Blanc, J. C., Baker, P. R. In-depth sphingomyelin characterization using electron impact excitation of ions from organics and mass spectrometry. J Lipid Res. 57 (5), 858-867 (2016).
  12. Ryan, E., Nguyen, C. Q. N., Shiea, C., Reid, G. E. Detailed Structural Characterization of Sphingolipids via 193 nm Ultraviolet Photodissociation and Ultra High Resolution Tandem Mass Spectrometry. J Am Soc Mass Spectrom. , (2017).
  13. Pham, H. T., Ly, T., Trevitt, A. J., Mitchell, T. W., Blanksby, S. J. Differentiation of complex lipid isomers by radical-directed dissociation mass spectrometry. Anal Chem. 84 (17), 7525-7532 (2012).

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Request permission to reuse the text or figures of this JoVE article

Request Permission

Tags

Sphingomyelin AnalysisLC MS MS MethodStable Isotope LabelingLipid ExtractionMobile Phase PreparationTriple Quadrupole MSProduct Ion SpectraStandard Curve ConstructionHeLa Cell SamplesCosmetic Lipid Characterization

Related Articles