An extra-corporeal robotic ultrasound (US) system refers to the configuration in which a robotic system is utilized to hold and manipulate a US probe for external examinations, including its use in cardiac, vascular, obstetric, and general abdominal imaging1. The use of such a robotic system is motivated by the challenges of manually holding and manipulating a US probe, for instance, the challenge of finding standard US views required by clinical imaging protocols and the risk of repetitive strain injury2,3,4, and also by the needs of US screening programs, for instance, the requirement for experienced sonographers to be on-site5,6. With emphases on different functionalities and target anatomies, several robotic US systems, as reviewed in earlier works1,7,8, have been introduced since the 1990s, to improve different aspects of US examination (e.g., long-distance teleoperation9,10,11,12, as well as robot-operator interaction and automatic control)13,14. In addition to the robotic US systems used for diagnostic purposes, robotic high-intensity focused ultrasound (HIFU) systems for treatment purposes have been widely investigated as summarized by Priester et al.1, with some recent works15,16 reporting the latest progress.
Although several robotic US systems have been developed with relatively reliable technologies for control and clinical operation, only a few of them have been successfully translated into clinical use, such as a commercially available tele-ultrasound system17. One possible reason is the low level of acceptance for large-size industrial-looking robots working in a clinical environment, from the point of view of both patients and sonographers. Additionally, for safety management, the majority of the existing US robots rely on force sensors to monitor and control the applied pressure to the US probe, while more fundamental mechanical safety mechanisms to limit the force passively are usually not available. This may also cause concerns when translating into clinical use as the safety of robot operation would be purely dependent on electrical systems and software logic.
With the recent advancements of 3D printing techniques, specially shaped plastic links with custom-made joint mechanisms could provide a new opportunity for developing bespoke medical robots. Carefully designed lightweight components with a compact appearance could improve clinical acceptance. Specifically for US examination, a bespoke medical robot aimed at being translated into clinical use should be compact, with enough degrees of freedom (DOFs) and range of motion to cover the region of interest of a scan; for example, the abdominal surface, including both the top and sides of the belly. Additionally, the robot should also incorporate the ability to perform fine adjustments of the US probe in a local area, when trying to optimize a US view. This usually includes tilting movements of the probe within a certain range, as suggested by Essomba et al.18 and Bassit19. To further address the safety concerns, it is expected that the system should have passive mechanical safety features which are independent of electrical systems and software logic.
In this paper, we present the detailed design and assembly method of a 5-DOF dexterous robotic manipulator, which is used as the key component of an extra-corporeal robotic US system. The manipulator consists of several lightweight 3D-printable links, custom-made joint mechanisms, and a built-in safety clutch. The specific arrangement of the DOFs provides full flexibility for probe adjustments, allowing easy and safe operations in a small area without colliding with the patient. The proposed multi-DOF manipulator aims to work as the main component that is in contact with patients and it can be simply attached to any conventional 3-DOF global positioning mechanism to form a complete US robot with fully active DOFs to perform a US scan.