Here we present simplified methods for fabricating transparent neurovascular phantoms and characterizing the flow therein. We highlight several important parameters and demonstrate their relationship to field accuracy.
Method Article
* These authors contributed equally
Here we present simplified methods for fabricating transparent neurovascular phantoms and characterizing the flow therein. We highlight several important parameters and demonstrate their relationship to field accuracy.
Particle image velocimetry (PIV) is used in a wide variety of fields, due to the opportunity it provides for precisely visualizing and quantifying flows across a large spatiotemporal range. However, its implementation typically requires the use of expensive and specialized instrumentation, which limits its broader utility. Moreover, within the field of bioengineering, in vitro flow visualization studies are also often further limited by the high cost of commercially sourced tissue phantoms that recapitulate desired anatomical structures, particularly for those that span the mesoscale regime (i.e., submillimeter to millimeter length scales). Herein, we present a simplified experimental protocol developed to address these limitations, the key elements of which include 1) a relatively low-cost method for fabricating mesoscale tissue phantoms using 3-D printing and silicone casting, and 2) an open-source image analysis and processing framework that reduces the demand upon the instrumentation for measuring mesoscale flows (i.e., velocities up to tens of millimeters/second). Collectively, this lowers the barrier to entry for nonexperts, by leveraging resources already at the disposal of many bioengineering researchers. We demonstratethe applicability of this protocol within the context of neurovascular flow characterization; however, it is expected to be relevant to a broader range of mesoscale applications in bioengineering and beyond.
PIV is widely used in experimental fluid mechanics for flow visualization and quantitative investigations of fluid motion that vary in length scale from atmospheric to microcirculatory flows1,2,3. While the specifics of its implementation can vary as widely as its applications, one aspect common to nearly all PIV studies is the use of video imaging of tracer particles seeded within the working fluid, followed by a pair-wise analysis of consecutive image frames to extract desired flow characteristics. Typically, this is accomplished by first subdividing each image frame into smaller regions termed interrogation windows. As a consequence of the random positions of the dispersed particles, each interrogation window contains a unique distribution of pixel intensities. If the window size and data acquisition rate are chosen appropriately, cross-correlation of the intensity signal in each window can be used to estimate the average displacement within that region. Finally, given that the magnification and frame rate are known experimental parameters, an instantaneous velocity vector field can be readily computed.
A major advantage of PIV over single-point measurement techniques is its ability to map vector fields across a two- or three-dimensional domain. Hemodynamic applications, in particular, have benefited from this capability, since it allows a thorough investigation of local flows, which are known to play a significant role in vascular disease or remodeling (e.g., atherosclerosis, angiogenesis)4,5,6. This has also been true for the evaluation of neurovascular flows, and the interactions thereof with endovascular devices (e.g., flow diverters, stents, intrasaccular coils), since the relevant length-scales in such applications can often span one or more orders of magnitude (e.g., from micrometer to millimeter), and device geometry and placement can significantly impact the local fluid mechanics7.
Most groups conducting PIV-based hemodynamic studies have relied on experimental set-ups that closely mimic some of the earliest investigations of stent influence on vascular flow7,8. Typically, these include a) pulsed lasers and high-speed cameras, to capture high-velocity flows; b) synchronizers, to prevent aliasing between the pulse frequency of the laser and the camera acquisition frame rate; c) cylindrical optics, to form a light sheet and, thus, minimize the background fluorescence from tracer particles above and below the interrogation plane; d) in the case of commercial turn-key systems, proprietary software packages, to perform the cross-correlation analyses. However, while some applications require the performance and/or versatility collectively afforded by these components, many others do not. Moreover, the high cost of commercially sourced tissue phantoms that recapitulate desired vascular structures can also prove limiting for many in vitro studies, particularly for phantoms with features that bridge the mesoscale regime (> 500 USD/phantom). Herein, we report the development of a simplified protocol for implementing PIV for the in vitro visualization of neurovascular flows, which typically lie both spatially and temporally within the mesoscale regime (i.e., length scales ranging from submillimeter to millimeter, and velocities up to tens of millimeters/second). The protocol seeks to leverage resources already at the disposal of many bioengineering researchers, thus lowering the barrier to entry for nonexperts.
The first element of this protocol involves the use of an investment casting technique to enable the in-house fabrication of transparent, polydimethylsiloxane (PDMS)-based tissue phantoms from 3-D-printed sacrificial molds. By leveraging the increasing availability of 3-D printers in recent years, particularly those in shared/multi-user facilities (e.g., institutional facilities or public makerspaces), this methodology cuts costs significantly (e.g., < 100 USD/phantom in the case presented here), while enabling a rapid turnaround for the fabrication of a wide variety of designs and geometries. In the current protocol, a fused deposition modeling system is used with acrylonitrile butadiene styrene (ABS) as the building material, and the printed part serves as a sacrificial mold for the subsequent phantom casting. Our experience has shown that ABS is well-suited for such use since it is soluble in common solvents (e.g., acetone), and it has sufficient strength and rigidity to maintain mold integrity after the removal of the support material (e.g., to prevent the deformation or fracture of diminutive mold features). In the current protocol, mold integrity is further ensured using solid printed models, although this comes at the expense of increased dissolution time. The use of hollow models may also be possible in some cases, to enhance solvent access, and thus, reduce dissolution time. However, careful consideration should be given to the effect this may have on mold integrity. Finally, while the phantoms fabricated herein are based upon idealized representations of neurovascular structures generated using a common computer-aided design (CAD) software package, the protocol is expected to be amenable to the fabrication of more complex, patient-specific geometries as well (e.g., via the use of model files generated by the conversion of clinical imaging data to the .STL file format used by most 3-D printers). Further details regarding the phantom fabrication process are provided in section 2 of the protocol.
The second element of the protocol involves the use of an open-source plug-in for ImageJ to conduct the cross-correlation analyses9. This is coupled with the implementation of a simple statistical thresholding scheme (i.e., intensity capping)10 to improve the image signal prior to cross-correlation, as well as a postcorrelation vector validation scheme, the normalized median test (NMT), to eliminate spurious vectors through a comparison of each to its nearest neighbors11. Collectively, this allows imaging to be accomplished using equipment commonly found in many bioengineering laboratories, thus eliminating the need for the acquisition of many of the costly components of typical PIV systems (e.g., pulsed laser, synchronizer, cylindrical optics, and proprietary software). Further details regarding the video collection, image processing, and data analysis are provided in sections 5 and 6 of the protocol.
Figure 1 illustrates the PIV set-up used in this protocol, which relies upon a fluorescence microscope equipped with a high-speed camera for imaging, as well as an external, continuous white-light source (i.e., metal halide lamp) for through-objective volumetric illumination. A variable-speed gear pump is used to impose the recirculating flow of a transparent mock blood solution through the neurovascular tissue phantoms. The solution is composed of a 60:40 mixture of deionized (DI) water and glycerol, which is a common substitute for blood in hemodynamic studies12,13,14, due to a) its similar density and viscosity (i.e., 1,080 kg/m3 and 3.5 cP vs. 1,050 kg/m3 and 3 - 5 cP for blood)15,16; b) its transparency in the visible range; c) its similar refractive index as PDMS (1.38 vs. 1.42 for PDMS)17,18,19,20, which minimizes optical distortion; d) the ease with which non-Newtonian behavior can be introduced, if needed, via the addition of xanthane21. Finally, fluorescent polystyrene beads are used as tracer particles (10.3 µm in diameter; 480 nm/501 nm excitation/emission). While neutrally buoyant beads are desired, sourcing tracer particles with optimal fluid mechanical properties (e.g., density, size, composition) and emission wavelength can prove challenging. For example, the beads used herein are slightly less dense than the glycerol solution (1,050 kg/m3 vs. 1,080 kg/m3). However, the hydrodynamic effects, thereof, are negligible, given that the duration of a typical experiment is far shorter than the time scale associated with buoyancy effects (i.e., 5 min and 20 min, respectively). Further details regarding the mock blood solution formulation and in vitro circulatory system set-up are provided in sections 3 and 4 of the protocol.
Access restricted. Please log in or start a trial to view this content.
1. ABS-based Sacrificial Mold Fabrication
2. PDMS-based Vascular Phantom Fabrication
3. Mock Blood Solution Formulation
4. In Vitro Circulatory System Set-up
5. Video Collection
6. Image Processing and Data Analysis
Access restricted. Please log in or start a trial to view this content.
Figure 2 illustrates the PDMS tissue phantom fabrication process. The phantoms designed herein are intended for the study of flow in idealized wide-necked, saccular, intracranial aneurysms, as well as proximal branching perforator arteries. Important additional design features include 1) a common reservoir that all vessels drain into, to ensure unencumbered fluid egress from the phantom - otherwise, droplet formation may occur at the smaller vessel outlets; 2...
Access restricted. Please log in or start a trial to view this content.
The protocol described herein outlines a simplified method for performing PIV studies to visualize neurovascular flows at physiologically relevant dimensions and flow conditions in vitro. In doing so, it serves to complement protocols reported by others that have also focused on simplifying the quantification of vector fields, but within very different contexts that require the consideration of far larger length scales25 or lower flow rates26,
Access restricted. Please log in or start a trial to view this content.
The authors have nothing to declare.
The authors acknowledge partial support for this project provided by a Collaborative Seed Grant from the Office of Research and Economic Development at UC Riverside.
Access restricted. Please log in or start a trial to view this content.
| Name | Company | Catalog Number | Comments |
|---|---|---|---|
| Solidworks 2015 | Dassault Systems | N/A | CAD Software |
| Dow Corning Sylgard 184 Kit | Ellsworth Adhesive | 184 SIL ELAST KIT 3.9KG | PDMS Kit |
| Stratasys Dimension Elite | Stratasys | 9180-00105 | 3D printer |
| P430 Model Material Cartridge | Stratasys | 340-21202 | ABS build material |
| P400 SR Soluble Support Material Cartridge | Stratasys | 340-30200 | Support material |
| CleanStation DT3 | PM3 Technologies | 00-00300R | Base bath |
| Lindberg Blue M LGO Box Furnace | Thermo Scientific | LB305745M | Oven |
| 21G BD PrecisionGlide Needle | Betcon Dickenson | BD 305167 | Branching perforator mold segment |
| Desiccator (Vacuum) | Polylab | 55205 | Desiccator |
| Branson 1800 Utrasonic Cleaning | Branson | CPX-952-116R | Sonicator |
| Acetone | Fisher Chemical | A9494 | Acetone |
| Isopropol Alcohol | Fisher Chemical | A4514 | Isopropol Alcohol |
| Glycerol | Fisher Chemical | GW33500 | Glycerol |
| 10um Polystyrene Yellow-Green Fluorescent Particles | Magsphere | PSF-010UM | Fluorescent beads |
| Phantom Miro | Vision Research | Miro M310 | High speed camera |
| Micropump | Cole-Parmer | 81101 | Recirculating pump |
| Leica DM2000 | Leica Microsystems | DM2000 | Fluorescent Microscope |
| Leica 10X Objective | Leica Microsystems | 506259 | Objective for perforator |
| Leica 2.5X Objective | Leica Microsystems | 11506083 | Objective aneurysm sac |
| Leica Blue Filter Cube L5 | Leica Microsystems | 513840 | Blue filter cube |
| Leica EL6000 | Leica Microsystems | 11504115 | Light source |
| Alconox | Alconox Inc | 1104-1 | Detergent |
| ImageJ | NIH | N/A | Open source image analysis software https://imagej.nih.gov/ij/ |
| Particle Image Velocimetry PIV Plugin | Qingson Tseng | N/A | https://sites.google.com/site/qingzongtseng/piv |
Access restricted. Please log in or start a trial to view this content.
Request permission to reuse the text or figures of this JoVE article
Request Permission