The harvested cell culture fluid from the automated microscale bioreactor is purified using fast protein liquid chromatography (FPLC), as seen in Figure 1 and the purified proteins' critical quality attributes (CQAs) were characterized by various downstream analytical methods. This is a key benefit of the automated microbioreactor system; differences in CQAs can be rapidly assessed across a wide range of conditions. N-glycan data from CHO-produced mAbs that are processed by mass spectrometry should appear like the chromatograms shown in Figure 2. The figure depicts a comparison between two chromatograms showing that the mannose 5 peak (M5) from one sample is considerably lower. If only a noisy baseline is observed instead of peaks, this may mean that the chromatography setup is faulty or that the procedure is not successful. Using controls, troubleshooting can be simplified. First, assess the FLR peaks from the dextran ladder; these peaks indicate that the chromatographic system is working correctly. Next, compare the experimentally obtained peaks with those obtained from a processed intact mAb standard. If peaks from the standard are visible, but no sample peaks are identified, then the mAb samples were not processed correctly. This may be due to SDS or nucleophile presence in the buffer interfering with N-glycan labeling and purification.
SEC-MALS can be used to assess two more CQAs: the aggregation profile and the molecular weight of the antibody. A representative SEC-MALS chromatogram is comparable to the one shown in Figure 3. The molecular mass distribution and the absolute molecular weight were determined using the required software with an extinction coefficient of 1.37 mL*(mg*cm)-1 and a dn/dc of 0.185 mL/g. As peak calling and setting the baseline in the software is performed manually, results may vary slightly from user to user. The absolute molecular weight of monomeric IgG1 from Figure 3 is 1.504 x 105 Da ± 0.38% (blue) and the higher order complex is 7.799 x 105 Da ± 3.0% (red). The polydispersity of the aggregates is much greater than that of the monomer, as indicated by the red molar mass distribution of Peak 1 (Figure 3). The small quantity of sample and importance of aggregation as a CQA make this technique a highly valuable complementary analytical tool to the automated microbioreactor system.
The result of mCZE is an electropherogram, such as in Figure 4, which shows the charge variant profile for a monoclonal antibody. The profile is a unique signature for the protein being investigated and is highly sensitive to the operating pH. Also visible is a free-dye peak to the left of the charge variant profile. When establishing an operating pH, there is some discretion to the operator to balance resolution and signal; in addition, the operator must ensure good separation from the free-dye peak which migrates at ~30 s. The sample can be desalted after labelling to remove this peak, though this leads to a significant loss in signal. Once an operating pH is established, the sample profiles can be compared. While generally consistent, changes in labeling efficiency or differences in excipients can lead to minor differences in the migration of a sample and the charge variant profile making electropherograms hard to directly compare. Instead, the method of comparison is usually based on the percentages of basic, main, and acidic species. In this case, relative differences as small as 1-2% can be identified using mCZE.
Amino acid consumption can be monitored to determine if depletion is causing changes in CQAs. Chromatogram readouts from the mass spectrometer can be used to evaluate the successful creation of a calibration curve for the absolute quantification of amino acids in crude bioreactor media samples. Figure 5 depicts two total ion chromatograms (TIC) and one extracted ion chromatogram (XIC) as representative results during this process. In Figure 5A, the TIC shown depicts the background signal from the buffer system as only a water blank was injected. Figure 5B depicts a representative TIC of the amino acid standard where, when compared to the water blank, small peaks that correspond to the individual amino acid species can be observed (such as lysine at 7.96 minutes). To integrate the peak and facilitate the quantification of peak area (and therefore the concentration), the XIC is used where only the signal from a defined "chromatogram mass window" is displayed. Depending on the sensitivity of the instrument and the quality of the chromatographic separation, the optimal mass window will have to be determined by the user. In this example (Figure 5C), the XIC of lysine (m/z = 147.1144) with a mass window of 10 ppm is shown where lysine in the amino acid standard elutes off the column at 8.03 minutes.

Figure 1. Representative chromatogram of the purification scheme using the Fast Protein Liquid Chromatography (FPLC) technique. Purification method phases corresponding to volume (mL) are labeled along the x-axis. UV absorbance at 280 nm (mAU y-axis, solid line) is monitored throughout the purification cycle. Non-specifically bound impurities are displaced by increasing conductivity (mS/cm y-axis, dashed line) during the High Salt Wash. Antibody is eluted from Protein A column with the introduction of elution buffer (Conc B, dotted line) when the pH decreases to 4 (not shown). Please click here to view a larger version of this figure.

Figure 2. A representative fluorescence chromatogram obtained from tagged glycans that are mass verified. The x-axis is retention time (minutes) while the y-axis is signal intensity. The peak at 14.94 min represents the Mannose 5 (M5) glycan, where a large difference between the M5 signal strength can be observed between the two samples that are overlaid. Please click here to view a larger version of this figure.

Figure 3. Molecular weight distribution of IgG1 monoclonal antibody. Chromatogram of an intact IgG1 monoclonal antibody separated by size exclusion chromatography in 1x PBS (pH7.4). Absorbance is monitored at 280 nm (black; left axis) and light scattering and refractive index detectors were used to calculate the absolute molecular weight of each peak (red and blue; right axis). High Molecular Weight species are indicated with the peak labeled "HMW". Please click here to view a larger version of this figure.

Figure 4. Charge variant profile of a IgG1 monoclonal antibody. This electropherogram is generated on a mCZE platform. A free-dye peak migrates at ~30 s and is well separated from the IgG1. For quantification, peaks were split into basic, main, and acidic species using instrument data analysis software. The red line outlines the integrated peak areas. Please click here to view a larger version of this figure.

Figure 5. Representative results of the ion chromatograms for mass spectrometry-based amino acid analysis of crude bioreactor media. The x-axis is time (minutes) while the y-axis is signal intensity (A) A water blank serves as the negative control and reveals the background signal observed over the course of the liquid chromatography gradient (B) The 225 pmol/µL amino acid standard is used here as a positive control, as the individual peaks observed in this total ion chromatogram represent the different amino acids of the standard mix being resolved chromatographically (C) A representative extracted ion chromatogram for m/z 147.1144, which is lysine. The 7.96 min peak in B corresponds to the 8.03 peak in C of lysine. Please click here to view a larger version of this figure.