The vaginal wall is composed of four layers, the epithelium, lamina propria, muscularis, and adventitia. The epithelium is primarily composed of epithelial cells. The lamina propria has a large amount of elastic and fibrillar collagen fibers. The muscularis is also composed of elastin and collagen fibers but has an increased amount of smooth muscle cells. The adventitia is comprised of elastin, collagen, and fibroblasts, albeit in reduced concentrations compared to the previous layers. The smooth muscle cells are of interest to biomechanically motivated research groups as they play a role in the contractile nature of the organs. As such, quantifying the smooth muscle cell area fraction and organization is key to understanding the mechanical function. Previous investigations suggest that the smooth muscle content within the vaginal wall is primarily organized in the circumferential and longitudinal axis. Histological analysis suggests that the smooth muscle area fraction is approximately 35% for both the proximal and distal sections of the wall1.
The cervix is a highly collagenous structure, that until recently, was thought to have minimal smooth muscle cell content2,3. Recent studies, however, have suggested that smooth muscle cells may have a greater abundance and role in the cervix4,5. The cervix exhibits a gradient of smooth muscle cells. The internal os contains 50-60% smooth muscle cells where the external os only contains 10%. Mouse studies, however, report the cervix to be composed of 10-15% smooth muscle cells and 85-90% fibrous connective tissue with no mention of regional differences6,7,8. Given that the mouse model differs from the frequently reported human model, further investigations concerning the mouse cervix are needed.
The purpose of this protocol was to elucidate the mechanical properties of the murine vagina and cervix. This was accomplished by using a pressure myograph device that enables assessment of mechanical properties in the circumferential and axial directions simultaneously while maintaining native cell-matrix interactions and organ geometry. The organs were mounted on two custom cannulas and secured with silk 6-0 sutures. Pressure-diameter tests were performed around the estimated physiological axial stretch to determine the compliance and tangent moduli9. Force-length tests were conducted to confirm the estimated axial stretch and to ensure that mechanical properties were quantified in the physiological range. The experimental protocol was performed on the nonpregnant murine vagina and cervix at 4-6 months of age in estrus.
The protocol is divided into two main mechanical testing sections: basal tone and passive testing. A basal tone is defined as the baseline partial constriction of smooth muscle cells, even in the absences of external local, hormonal, and neural stimulation10. This baseline contractile nature of the vagina and cervix yields characteristic mechanical behaviors which are then measured by the pressure myograph system. The passive properties are assessed by removing the intercellular calcium that maintains the baseline state of contraction, resulting in relaxation of the smooth muscle cells. In the passive state, collagen and elastin fibers provide the dominant contributions for the mechanical characteristics of the organs.
The murine model is used extensively to study pathologies in women’s reproductive health. The mouse offers several advantages for quantifying the evolving relationships between ECM and mechanical properties within the reproductive system11,12,13,14. These advantages include short and well-characterized estrous cycles, relatively low cost, ease of handling, and a relatively short gestational time15. Additionally, the genome of laboratory mice is well-mapped and genetically-modified mice are valuable tools to test mechanistic hypotheses16,17,18.
Commercially available pressure myograph systems are used extensively to quantify the mechanical responses of various tissues and organs. Some notable structures analyzed on the pressure myograph system include elastic arteries19,20,21,22, veins and tissue engineered vascular grafts23,24, the esophagus25, and the large intestines26. The pressure myograph technology permits simultaneous assessment of properties in the axial and circumferential directions while maintaining the native cell-ECM interactions and in vivo geometry. Despite the extensive use of myograph systems in soft tissue and organ mechanics, a protocol utilizing the pressure myograph technology had not previously been developed for the vagina and cervix. Prior investigations into the mechanical properties of the vagina and cervix were assessed uniaxially27,28. These organs, however, experience multiaxial loading within the body29,30, thus quantifying their biaxial mechanical response is important.
Moreover, recent work suggests smooth muscle cells may play a potential role in soft tissue pathologies5,28,31,32. This provides another attraction of utilizing the pressure myograph technology, as it preserves the native cell-matrix interactions, thus permitting delineation of the contribution that smooth muscle cells play in physiological and pathophysiological conditions. Herein, we propose a protocol to quantify the multiaxial mechanical properties of the vagina and cervix under both basal tone and passive conditions.