$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
Coil Characterization
Upon successful tuning and matching of a coil, its performance may be characterized by the coil Q-factor, 90° reference pulse, and SNR/mm3. For the 1.5 mm ID susceptibility-matched solenoid coil demonstrated here, the measured Q-factor(unloaded) was 244, compared to 561 for a 5 mm birdcage coil.
The reference 90° pulse was 12 µs at a power level of 0.6 W; cf. 5 µs at 45 W for a 5 mm birdcage coil (Figure 4 and Figure 5). This equates to an RF pulse field strength (B1), using
of 0.53 mT for the microcoil and 1.17 mT for the birdcage coil14 where y is the gyromagnetic ratio, while tau is the pulse duration. Since the pulse power levels (P) differ, coils may be compared in terms of transmit efficiency
: 0.69 mT/W1/2 and 0.18 mT/W1/2 for the microcoil and birdcage respectively14. Comparing by a 90° pulse, the microcoil is found to be a factor ≈ 4 times more sensitive than the birdcage coil.
Effect of susceptibility matching
At ultra-high field strengths, sample and coil susceptibility become a dominant factor for image quality, as seen in Figure 7A,7B. Compared to a coil lacking a susceptibility matching fluid reservoir, the signal is retained longer and more homogeneously in a reference sample. However, due to the susceptibility reservoir, the maximum sample dimensions decrease with respect to the coil without the reservoir.
High-resolution imaging
A high resolution of 13 x 13 x 13 μm3 of a Medicago truncatula root specimen was attained in 20 hours and 23 minutes (Figure 8). Starting from the surface of the root, the root cortex is seen, along with some residual water on the outside of the root. Furthermore, the xylem is observed as a dark band enclosing the phloem. Some air pockets are observed as dark spots with complete signal loss.
Symbiotic root nodules of M. truncatula may also be imaged using this protocol (Figure 9). Using a slightly larger unmatched coil (length circa 3500 µm, inner diameter 1500 μm), images with a resolution of up to 16 x 16 x 16 μm3 were obtained in 33 minutes.

Figure 1: A solenoid microcoil. (A) The solenoid coil design consists of wire looped helically, typically wrapped around a capillary. The geometry of the wire, such as its thickness, diameter, number of windings, and wire spacing, influence the coil characteristics. (B) A home-built solenoid microcoil with a reservoir for susceptibility matching fluid (Fomblin). It consists of a 0.4 mm thick coated copper wire wound six times around capillary with an outer diameter of 1500 µm and a coil length of 3500 µm. The coil is submerged in a reservoir which is made from a syringe. Sample capillaries up to an outer diameter of 1000 µm can be inserted. Two capacitors are used, a 1.5 pF capacitor in series with the inductor and a second variable 1.5-6 pF capacitor is placed in parallel to the inductor. All components are soldered to a fiberglass board (yellow). It is mounted on a commercial holder (grey polymer) that is modified to support the reservoir. (C) Solenoid coil design components: 1. solenoid coil, 2. sample capillary, 3. 1.5 pF tuning capacitor, 4. variable matching capacitor, 5. fiberglass base plate, 6. copper wire leads. Please click here to view a larger version of this figure.

Figure 2: Sample preparation under a stereomicroscope. (A) Items needed for the preparation of microcoils. From left to right: 1. CuSO4 reference solution, 2. perfluorodecalin, 3. microcoil, 4. scalpel, 5. positive tension tweezers, 6. tweezers, 7. capillaries outer diameter = 1000 μm, 8. wax pen, 9. capillary wax, 10. nitrile gloves, 11. stereomicroscope, 12. watch glass with Petri dish cover, 13. plant material in growth substrate. Not shown: 2 mL syringe with ø 0.8 x 40 mm needle and fine tissue paper. (B) Close up of sample insertion into a capillary using tweezers, while both are kept submerged. (C) Sealing of the capillary using molten wax. (D) Insertion of the prepared capillary into the microcoil. Please click here to view a larger version of this figure.

Figure 3: The component of a micro-imaging probe. (A) Micro5 probe base, containing all necessary connections for water cooling, heating, temperature sensors, gradient power, RF (co-axial connector visible) and optionally probe identification (PICS). Underneath the probe base are knobs that allow for adjusting the variable tuning and matching capacitors, as well as retaining screws to hold the probe in place inside the spectrometer. (B) The home-built microcoil mounting atop the probe-base. Note the variable capacitors (white ceramic) mounted on the probe-base that allow for tuning and matching. (C) Integrated 3-axial gradient mounted on the probe base with water-cooling receptacles and gold-plated contacts for grounding the gradient. Please click here to view a larger version of this figure.

Figure 4: Nutation curve. A nutation curve is acquired to determine the reference pulse power. The reference pulse power (90° pulse) is defined as the combination of power and pulse length needed to generate a B1 field that flips all available magnetization in the z-direction to the transverse plane. A series of a pulse is recorded in the absence of gradient encoding. With each pulse, either pulse length or pulse power is incremented. Here the pulse power is set to 0.6 W, while the pulse length is incremented by 1 µs each time. The maximum signal intensity indicates the 90° pulse, around 12 µs. The 180° pulse may also be determined in this way using the minimum intensity. Please click here to view a larger version of this figure.

Figure 5: Visual determination of 90° pulse power. Once an approximate reference pulse power has been found using a nutation curve, it may be checked visually by varying the pulse length. Depending on the coil, the B1 field may be more or less sensitive to changes. (A) 11 µs pulse length. (B) 12 µs pulse length, optimal for this coil. (C) 13 µs pulse length. (D) 20 µs pulse length. If the pulse power is set too high, over-tipping may occur, thereby reducing image intensity in the center of the coil (arrowhead). The increased B1 field also increases the range of the coil, as can be observed in the width of the image. Please click here to view a larger version of this figure.

Figure 6: Region of Interest placement. The regions of interest (ROI) for the volume normalized SNR calculation can be seen. The mean sample intensity is taken from an ROI that falls within the reference solution sample. The mean noise and standard deviation are calculated from one or more ROI located in the corners of the image. Please click here to view a larger version of this figure.

Figure 7: RF homogeneity evaluated by gradient echo imaging. A multiple gradient echo (MGE) sequence is used to evaluate RF (B1 -Field) homogeneity using a series of gradient echoes. Basic parameters were: repetition time 200 ms, echo time 3.5 ms with the number of echoes 48, echo spacing 3.5 ms, 64 averages, acquisition time 27 m 18 s, flip angle 30°. Field of view was 5 x 5 mm, matrix 128 x 128, resolution 39 x 39 x 200 µm. (A) Susceptibility-matched coil. The susceptibility matching fluid (Fomblin) surrounding the RF coil reduces susceptibility effects due to the coil wire. Small air bubbles cause loss of signal as the echo time increases. (B) A coil (not susceptibility matched) with equal coil diameter. At longer echo times, increasing artifacts caused by B0 field inhomogeneity are observed. Please click here to view a larger version of this figure.

Figure 8: 3D imaging of a Medicago truncatula root section. (Top) FLASH image. Several features of the root section can be distinguished, including the epidermis (e), cortex (c), phloem (ph) and xylem (xy). Air pockets (a) in the root cause complete signal loss. Basic parameters were as follows: Repetition time 70 ms, echo time 2.5 ms, 256 averages, acquisition time 20 h 23 m. Resolution 13 x 13 x 13 µm3. Matrix size was 128 x 64 x 64 and field of view 1.6 x 0.8 x 0.8 mm. Receiver bandwidth 50 kHz. (Bottom) MSME image. Basic parameters were as follows: Repetition time 500 ms, echo time 5.2 ms, 28 averages, acquisition time 15 h 55 m. Resolution 13 x 13 x 13 µm3. Matrix size was 128 x 64 x 64 and field of view 1.6 x 0.8 x 0.8 mm. Receiver bandwidth 70 kHz. Please click here to view a larger version of this figure.

Figure 9: 3D imaging of a Medicago truncatula root nodule. (Top) Low-resolution image. Basic parameters were as follows: Repetition time 60 ms, echo time 2.3 ms, 4 averages, acquisition time 4 m. Resolution 31 x 31 x 31 µm3. Matrix size was 64 x 32 x 32 and field of view 2 x 1 x 1 mm. Receiver bandwidth 50 kHz. (Bottom) High-resolution image. Basic parameters were as follows: Repetition time 60 ms, echo time 2.3 ms, 8 averages, acquisition time 33 m. Resolution 16 x 16 x 16 µm3. Matrix size was 128 x 64 x 64 and field of view 2 x 1 x 1 mm. Receiver bandwidth 50 kHz. Please click here to view a larger version of this figure.