Method Article

Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models

3.6K views

⸱

DOI:

10.3791/61706

⸱

November 11th, 2020

 ,  , 

In This Article

Summary

We present a high-throughput, in vitro method for quantifying regional pulmonary deposition at the lobe level using CT scan-derived, 3D printed lung models with tunable air flow profiles.

Abstract

Development of targeted therapies for pulmonary diseases is limited by the availability of preclinical testing methods with the ability to predict regional aerosol delivery. Leveraging 3D printing to generate patient-specific lung models, we outline the design of a high-throughput, in vitro experimental setup for quantifying lobular pulmonary deposition. This system is made with a combination of commercially available and 3D printed components and allows the flow rate through each lobe of the lung to be independently controlled. Delivery of fluorescent aerosols to each lobe is measured using fluorescence microscopy. This protocol has the potential to promote the growth of personalized medicine for respiratory diseases through its ability to model a wide range of patient demographics and disease states. Both the geometry of the 3D printed lung model and the air flow profile setting can be easily modulated to reflect clinical data for patients with varying age, race, and gender. Clinically relevant drug delivery devices, such as the endotracheal tube shown here, can be incorporated into the testing setup to more accurately predict a device’s capacity to target therapeutic delivery to a diseased region of the lung. The versatility of this experimental setup allows it to be customized to reflect a multitude of inhalation conditions, enhancing the rigor of preclinical therapeutic testing.

Introduction

Many pulmonary diseases such as lung cancer and chronic obstructive pulmonary disease (COPD) exhibit regional differences in disease characteristics; however, there are a lack of therapeutic techniques available to target drug delivery to only diseased regions of the lung1. Multiple computational fluid dynamic (CFD) models have demonstrated that it is possible to modulate drug deposition profiles by identifying specific streamlines in the lung2,3. Development of both inhalers and endotracheal (ET) tube adaptors with regional targeting capabilities are on-going in our lab to control aero....

Access restricted. Please log in or start a trial to view this content.

Protocol

1. Preparation of 3D printed experimental components

NOTE: All software used in the protocol are indicated in the Table of Materials. Additionally, the slicing software utilized is specific to the 3D printer listed in the Table of Materials; however, this protocol can be extended to a wide range of stereolithography (SLA) 3D printers.

  1. Convert patient CT scans to 3D objects (.stl files).
    NOTE: For a more detailed discussion of the geometric features of the specific lung model used in these studies, refer to Feng et al.5.
    1. Render CT scans into a 3D ob....

Access restricted. Please log in or start a trial to view this content.

Results

Particles in this size range (1-5 μm) and flow conditions (1-10 L/min) follow the fluid stream lines based on both their theoretical Stokes number and in vivo data; therefore, in the absence of a targeted delivery device, particles released into the lung model are expected to deposit according to the percentage of total airflow diverted to each lobe. The relative amounts of particle delivery to each lobe can then be compared to clinical lobe flow rate data obtained through analyzing patient-specific high-resolution compu.......

Access restricted. Please log in or start a trial to view this content.

Discussion

The current state-of-the-art device for pulmonary pharmaceutical testing of a complete inhalation dose is the Next Generator Impactor (NGI), which measures the aerodynamic diameter of an aerosol4. This sizing data is then used to predict the lung generation at which the aerosol will deposit based on a correlation developed for a healthy adult male11. Unfortunately, this method is limited in its ability to assess differences in regional lung deposition, determine the effects.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors have nothing to disclose.

Acknowledgements

The authors thank Professor Yu Feng, Dr. Jenna Briddell, Ian Woodward, and Lucas Attia for their helpful discussions.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
1/4" Plastic Barbed Tube FittingMcMaster Carr5372K111
10 um Filter PaperFisher1093-110
1um Fluorescent Polystyrene ParticlesPolysciences15702-10
1um Non-Fluorescent Polystyrene ParticlesPolysciences8226
2-PropanolFisherA516-4Referred to in protocol as "IPA"
3/8" Plastic Barbed Tube FittingMcMaster Carr5372K117
Air Flow Meter (1 - 280 mL/min)McMaster Carr41695K32Referred to in protocol as "flow meter"
Carbon M1 3D PrinterCarbon 3Dhttps://www.carbon3d.com/, Associated software referred to in protocol as "slicing software"
Collison Jet NebulizerCH TechnologiesARGCNB0008 (CN-25)6 Jet MRE style horizontal collision with glass jar, Referred to in protocol as "nebulizer", http://chtechusa.com/Manuals/MRE_Collison_Manual.pdf
Convection OvenYamatoDKN602
Copley Critical Flow Controller TPK2000 Reve 120VMSP Corp0001-01-9810Referred to in protocol as "flow controller"
Copley High Capacity Pump Model HCP5MSP Corp0001-01-9982Referred to in protocol as "vacuum pump"
CytationBioTekCYT5MPVMultifunctional Spectrophotometer/Fluorescent imager equiped with 4x/20x/40x objectives and DAPI/GFP/TexasRed laser/filter cubes
EPU40 ResinCarbon 3Dhttps://www.carbon3d.com/materials/epu-elastomeric-polyurethane/, Referred to in protocol as "soft resin"
Filter for vacuum pumpWhatman6722-5000
Flow Meter Model DFM 2000MSP Corp0001-01-8764Referred to in protocol as "electronic flow meter"
ImageJ SoftwareImageJhttps://imagej.nih.gov/ij/download.html
Inline Air Flow Control Valve (Push-to-Connect)McMaster Carr62005K333Referred to in protocol as "valve"
Inline Filter DevicesWhatmanWHA67225000
Marine-Grade Plywood SheetMcMaster Carr62005K333Referred to in protocol as "wooden board"
Materialise Mimics SoftwareMaterialisehttps://www.materialise.com/en/medical/mimics-innovation-suite, Referred to in protocol as "CT scan software"
Meshmixer SoftwareAutodeskhttp://www.meshmixer.com/, Referred to in protocol as "mesh editing software"
MethanolFisherA454-4
Opticure LED CubeAPM Technica102843Referred to in protocol as "UV oven"
PR25 ResinCarbon 3Dhttps://www.carbon3d.com/materials/uma-urethanemethacrylate, /Referred to in protocol as "hard resin"
PVC Tube for ChemicalsMcMaster Carr5231K1611/4" ID
Screws
SolidWorks SoftwareDassault Systèmes SolidWorks Corporationhttps://www.solidworks.com/, Referred to in protocol as "3D modeling software"
Straight Flow Rectangular ManifoldMcMaster Carr1125T31
Tubing to Flow ControllerMcMaster Carr5233K653/8" ID
Wire

References

  1. Goel, A., Baboota, S., Sahni, J. K., Ali, J. Exploring targeted pulmonary delivery for treatment of lung cancer. International Journal of Pharmaceutical Investigation. 3 (1), 8-14 (2013).
  2. Kleinstreuer, C., Zhang, Z., Li, Z., Roberts, W. L., Rojas, C.

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Explore More Articles

3D Printed Lung ModelsPatient-Specific Lung ModelsFluorescent Aerosol DeliveryFlow Rate ControlFluorescence MicroscopyEndotracheal Tube TestingLobe Outlet Cap AssemblyNebulizer Aerosol GenerationFilter Paper Deposition