Immunology and Infection

User-friendly, High-throughput, and Fully Automated Data Acquisition Software for Single-particle Cryo-electron Microscopy

Published: July 29, 2021 doi: 10.3791/62832

Summary

Single-particle cryo-electron microscopy demands a suitable software package and user-friendly pipeline for high-throughput automatic data acquisition. Here, we present the application of a fully automated image acquisition software package, Latitude-S, and a practical pipeline for data collection of vitrified biomolecules under low-dose conditions.

Abstract

In the past several years, technological and methodological advancements in single-particle cryo-electron microscopy (cryo-EM) have paved a new avenue for the high-resolution structure determination of biological macromolecules. Despite the remarkable advances in cryo-EM, there is still scope for improvement in various aspects of the single-particle analysis workflow. Single-particle analysis demands a suitable software package for high-throughput automatic data acquisition. Several automatic data acquisition software packages were developed for automatic imaging for single-particle cryo-EM in the last eight years. This paper presents an application of a fully automated image acquisition pipeline for vitrified biomolecules under low-dose conditions.

It demonstrates a software package, which can collect cryo-EM data fully, automatically, and precisely. Additionally, various microscopic parameters are easily controlled by this software package. This protocol demonstrates the potential of this software package in automated imaging of the severe acute respiratory syndrome-coronavirus 2 (SARS-CoV-2) spike protein with a 200 keV cryo-electron microscope equipped with a direct electron detector (DED). Around 3,000 cryo-EM movie images were acquired in a single session (48 h) of data collection, yielding an atomic-resolution structure of the spike protein of SARS-CoV-2. Furthermore, this structural study indicates that the spike protein adopts two major conformations, 1-RBD (receptor-binding domain) up open and all RBD down closed conformations.

Introduction

Single-particle cryo-EM has become a mainstream structural biology technique for high-resolution structure determination of biological macromolecules1. Single-particle reconstruction depends on acquiring a vast number of micrographs of vitrified samples to extract two-dimensional (2D) particle images, which are then used to reconstruct a three-dimensional (3D) structure of a biological macromolecule2,3. Before the development of DEDs, the resolution achieved from single-particle reconstruction ranged between 4 and 30 Å4,5. Recently, the achievable resolution from single-particle cryo-EM has reached beyond 1.8 Å6. DED and automated data acquisition software have been major contributors to this resolution revolution7, where human intervention for data collection is minimal. Generally, cryo-EM imaging is performed at low electron dose rates (20-100 e/Å2) to minimize electron beam-induced radiation damage of biological samples, which contributes to the low signal-to-noise ratio (SNR) in the image. This low SNR impedes the characterization of the high-resolution structures of biological macromolecules using single-particle analysis.

The new generation electron detectors are CMOS (complementary metal-oxide-semiconductor)-based detectors, which can overcome these low SNR-related obstacles. These direct detection CMOS cameras allow fast readout of the signal, due to which the camera contributes better point spread function, suitable SNR, and excellent detective quantum efficiency (DQE) for biological macromolecules. Direct detection cameras offer high SNR8 and low noise in the recorded images, resulting in a quantitative increase in the detective quantum efficiency (DQE)-a measure of how much noise a detector adds to an image. These cameras also record movies at the speed of hundreds of frames per second, which enables fast data acquisition9,10. All these characteristics make fast direct detection cameras suitable for low-dose applications.

Motion-corrected stack images are used for data processing to calculate 2D classification and reconstruct a 3D density map of macromolecules by using various software packages such as RELION11, FREALIGN12, cryoSPARC13, cisTEM14, and EMAN215. However, for single-particle analysis, an enormous dataset is required to achieve a high-resolution structure. Therefore, automatic data acquisition tolls are highly essential for data collection. To record large cryo-EM data sets, several software packages have been used over the past decade. Dedicated software packages, such as AutoEM16, AutoEMation17, Leginon18, SerialEM19, UCSF-Image420, TOM221, SAM22, JAMES23, JADAS24, EM-TOOLS, and EPU, have been developed for automated data acquisition.

These software packages use routine tasks to find hole positions automatically by correlating the low-magnification images to high-magnification images, which assists in identifying holes with vitreous ice of appropriative ice thickness for image acquisition under low-dose conditions. These software packages have reduced the number of repetitive tasks and increased the throughput of the cryo-EM data collection by acquiring a vast amount of good-quality data for several days continuously, without any interruption and the physical presence of the operator. Latitude-S is a similar software package, which is used for automatic data acquisition for single-particle analysis. However, this software package is only suitable for K2/K3 DEDs and is provided with these detectors.

This protocol demonstrates the potential of Latitude-S in the automated image acquisition of SARS-CoV-2 spike protein with a direct electron detector equipped with a 200 keV cryo-EM (see the Table of Materials). Using this data collection tool, 3,000 movie files of SARS-CoV-2 spike protein are automatically acquired, and further data processing is carried out to obtain a 3.9-4.4 Å resolution spike protein structure.

Subscription Required. Please recommend JoVE to your librarian.

Protocol

NOTE: Three important steps are required for cryo-EM data collection: 1. cryo-EM grid preparation, 2. calibration and alignment of the microscope, 3. automatic data collection (Figure 1). Furthermore, automated data collection is subdivided into a. suitable area selection, b. optimization of Latitude-S, c. start automatic hole selection, and d. start automatic data acquisition (Figure 1).

1. Cryo-EM grid preparation and sample loading for automatic data acquisition

  1. Clean the grids using a glow discharger and vary the glow discharge parameters based on experimental requirements (here, 60 s at 20 mA).
  2. Add a freshly prepared protein sample (3 µL) to the glow-discharged grid and incubate for 10 s.
  3. Blot the grids for 3-5 s at 100% humidity and quickly plunge them into liquid ethane using a cryo-plunger.
  4. Clamp the grids manually into a clip ring to form the cartridge using a flexible C-clip ring.
  5. Load the frozen cartridge-mounted grids into the autoloader cassette and transfer the cassette by the nano cap to the precooled autoloader of the microscope for data collection.

2. Microscope tuning and basic alignment before automatic data acquisition

  1. Beam shift
    1. Click on Beam shift from the Direct Alignment tab.
    2. Reduce the magnification, and center the beam to the optical axis using the Multifunction X and Y knob.
  2. Pivot point alignment
    1. Click on the Beam tilt option in Direct Alignment pp X from the Direct Alignment tab.
    2. Condense the beam to a spot and minimize the movement by using the Multifunction X and Y knob.
  3. C2 aperture centering
    1. Select the condenser aperture from the Alignment tab.
    2. Condense the beam to a spot, center the beam to the optical axis, and then expand the beam to cover the circle evenly.
    3. Repeat these steps until the Condenser 2 aperture is adjusted.
  4. Coma-free alignment
    1. Click on Coma-free Alignment X from the Direct Alignment tab to align the beam to the optical axis.
    2. Use the Multifunction knob to minimize the shape and the movement of the FFT (ensure it is stable).
    3. Repeat the same procedure for Coma - free alignment Y.
  5. Set parallel illumination before data collection in the cryo-EM because of the C twin lens.
    1. Insert the objective aperture (70 µm) in diffraction mode.
    2. Focus the objective aperture on the front focal plane of the diffraction lens by controlling the defocus-intensity knob (objective lens and C2 lens current).
    3. Ensure that the crisp edge of the objective aperture is seen after proper defocusing.
    4. Insert the beam stopper and spread the intensity until the gold powder diffraction rings are minimized.
      NOTE: If the beam is spread properly, a clear diffraction ring of the gold powder is visible on the screen, which indicates that the beam is parallel.
    5. Retract the beam stopper after setting the parallel illumination and change the microscope mode to Nano probe.
      ​NOTE: Check the microscope tuning before starting data collection to ensure the optimum performance of the microscope. All these settings would be performed in the microscope Direct Alignment GUI Tab. All the microscope tuning is performed using a test grid before data collection.

3. Data acquisition with Latitude-S

  1. Start Latitude-S automated data acquisition software.
    NOTE: Latitude-S installation also requires microscope calibration, which will be performed before data collection, and the settings will be stored permanently. Five different states for data collection are calibrated with four different magnifications (Figure 1 and Figure 2). Atlas state and grid state are in two different magnifications in LM mode (low-magnification ranges). The hole state is in SA mode (high-magnification ranges) but with a moderate magnification. Focus and data state use high-magnification SA mode.
    1. Click on DigitalMicrograph from the Start menu, or double-click on the DigitalMicrograph icon on the desktop.
    2. Select the Technique manager icon from DigitalMicrograph.
      NOTE: This system will show TEM and Latitude-S icons (Figure 2 and Supplemental Figure S1).
    3. Select the Latitude-S icon for single-particle automated data collection.
      NOTE: The K2 camera operates in three modes: linear/integrated, counted, and super-resolution. The user could select any mode in the interface of DigitalMicrograph. Data images could be saved as either dose-fractionated image stacks or as summed images in MRC, TIF, or .dm4 files with different bit depths. Furthermore, data could be saved as motion-corrected images for the K3 camera. On a K2 camera, an unprocessed image stack could be saved as 4-bit MRC, 8-bit TIF, or 8-bit .dm4 files.
  2. Create a new session based on the settings from a previous session.
    1. Check the Based on prior session checkbox in the palette.
    2. Select the New button.
    3. Choose the folder containing the session on which the new session's settings are based. Go to the previous session directory to create the new session. Choose the folder to save the new session and associated data.
    4. Select and choose the folder where the new session and associated data will be saved.
      NOTE: Each state and its underlying settings (magnification, illumination conditions, image, or projector) and beam shift and camera parameters (total exposure, single-frame exposure, and binning) will be exported from the existing session to the new session. The path of the folder is shown as a text string at the bottom of the palette. Each of the states and configuration palettes has an asterisk (*) appended to the title to show that it has already been set up and is ready to use.
  3. Continue an existing session.
    1. Press the Continue button in the palette to continue an existing session.
      NOTE: The atlas montage cannot be modified.
    2. Choose and navigate to the folder that contains the session that needs to be continued.
  4. Start an entirely new session.
    1. Click on New tab in the palette. Choose the folder that contains the session to be continued. Select a folder to save the data.
      NOTE: The default folder name is built by using the date and time.
    2. Click on the Setting icon. In the Manage state explore box that appears, add state, set the TEM condition, the camera condition, and image/stack, and then name the state.
      NOTE: The automated data acquisition workflow uses 5 different states for automated data collection. These states are configured and stored in their respective state palettes. The state summary is given in Table 1.
  5. Configure the atlas state.
    1. Click on Atlas state palette.
    2. Configure the atlas state with the following parameters: magnification 115x LM mode in nano probe, illumination conditions-spot size 8 and brightness 934400, binning: 1 and camera exposure time: 1.0 s for imaging at low magnification. Refer to the state summary given in Table 1.
    3. Click on Next to move to the next state.
      NOTE: Atlas state is the lowest magnification state, which provides the survey of the entire grid (Supplemental Figure S2). Generally, this state helps us visualize the entire grids at low magnification and judge the ice thickness, flatness, and broken square of the grids. It is recommended to generate the atlas at different areas of the grid to observe the optimal ice thickness and suboptimal ice thickness of the grids (Supplemental Figure S3). The mentioned parameters could be varied according to the user's needs.
  6. Configure the grid state.
    1. Click on the Grid state palette.
    2. Configure the Grid state with the following microscope imaging optics (magnification 380x LM mode in Nano probe), illumination conditions (spot size: 8 and brightness 626,200), binning: 1 and camera exposure time: 1.0 s.
    3. See the Latitude-S state summary provided in Table 1.
    4. Click on Next to move to the next state.
      NOTE: The grid state is set at a magnification higher than the atlas state such that the field of view is one grid square (Figure 2). In this particular magnification, one grid square is observed. Therefore, holes are observed correctly in this magnification, which helps check the ice thickness of the holes (Supplemental Figure S4). A simple bandpass filter is used in the grid state to locate the holes in the patent grid. The mentioned parameters could be varied according to the user's needs.
  7. Configure the hole state.
    1. Click on the hole palette.
    2. Configure the Hole state with the following microscope settings: imaging optics (magnification 4500x SM mode in Nano probe), illumination conditions (spot size: 7 and Beam diameter 8.81 µm), binning: 1 and camera exposure time: 1.0 s.
    3. Change the parameters if required based on the grid type. See the state summary provided in Table 1.
    4. Click on Next to move to the next state.
      NOTE: The SA mode indicates a high-magnification range in the electron microscope. The hole state is in the SA magnification range with a field of view of a few micrometers (10-20 µm) (Figure 2 and Supplemental Figure S4). This magnification range is higher than the Atlas or Grid state but much smaller than Focus/Data state. In this magnification, individual holes will be visible. The hole size is appropriate to observe high degrees of contaminations, empty holes, and the proper ice thickness of the holes. The holes for imaging are selected based on these assumptions. Two filters are used in the hole state: one for cross-correlating a hole reference image with a new hole image and another for adjusting the stage height to the eucentric height.
  8. Configure the focus state.
    1. Click on Focus palette.
    2. Configure the focus state with the following microscope settings: imaging optics (magnification 45,000x SA mode in nano probe), illumination conditions (spot size: 8 and brightness 934400), binning: 1 and camera exposure time: 1.0 s.
    3. Focus on the amorphous carbon area near the hole. Refer to the Latitude-S state summary provided in Table 1.
    4. Click on Next to move to the next state.
      NOTE: The SA mode indicates a high-magnification range in the electron microscope. The Focus state is the higher SA range magnification. In focus mode, the beam is shifted to a nearby carbon area of the target hole and performs focus automatically to collect the data in the data state. A bandpass filter combined with a hanning or soft rectangular filter is used in the focus state to measure the offset between two focus state images of the same area (Figure 2). The mentioned parameters could be varied according to the user's needs.
  9. Configure the data state.
    1. Click on Data palette.
    2. Configure the data state with the following microscope settings: imaging optics (e.g., magnification 28,000x, 45,000x, 54,000x at SA mode in nano probe), illumination conditions (spot size: 8 and brightness 934400), binning: 1 and camera exposure time: 1.0 s.
    3. Refer to the Latitude-S state summary given in Table 1.
    4. Click on Next to move to the next state.
      ​NOTE: Data state is the highest magnification selected based on pixel size requirements and target resolution (Figure 2). Generally, after focusing, the beam is automatically shifted to the target area to collect the data. The above-mentioned parameters could be changed based on the user's requirements.

4. Focus configuration

  1. Click on Focus configuration palette. Specify the range of defocus values and the step size in the given tab.
  2. Press Next button to move to the next step.
    ​NOTE: Lower defocus values can be used for high-resolution data acquisition. Generally, -0.5 to -3.0 µm defocus values with 0.25 or 0.5 defocus step sizes are used for image acquisition. Users can skip the focus setup step if they only want to screen the sample. Simply press the Next button on the palette to skip the focus configuration step.

5. Fine alignment

  1. Focus on some features on the grid (e.g., ice contamination hexagonal ice); see Figure 3.
    NOTE: Features should not be too big or too small. They should be visible at both Atlas state magnification 115x (LA mode) and data magnification.
  2. Click on the Capture button. Position the red cross mark on the same feature on each image of different states.
  3. Start with focus, data, and hole states because their field of view is much bigger than the atlas and grid states. Zoom on atlas and grid states to position the red cross mark on the same feature in the atlas and grid states.
  4. Click on the Calculate button to calculate the positions of five different states, which will calculate the offsets between each of the states and reflect these to the output window.
    ​NOTE: The offset values are integrated into the states for further use (Figure 3). Fine alignment is performed to provide high accuracy of the position of each state (Figure 3). This fine alignment helps pinpoint the exact position in all the five states. Fine Alignment is critical for single-particle data acquisition. Therefore, it is highly recommended to perform Fine Alignment before imaging.

6. Data acquisition procedure using Latitude-S

  1. Click on the Capture palette.
    NOTE: Generally, atlas data is collected at low magnification (115x) to visualize most of the grid squares.
  2. Choose the size of the atlas to cover the entire grid or part of the grid based on the requirement (e.g., 6 x 6, 8 x 8, 12 x 12, 6 x 8, 8 x 6, 12 x 8, or 8 x 12).
    NOTE: 16 by 16 atlas size covers the entire grid.
  3. Click on the Capture button to capture the atlas.
    NOTE: The main Latitude-S navigation window opens and fills the available space in DigitalMicrograph (Supplemental Figure S5). Three picture panes in the main navigation window show images of the system states at three different magnifications. The overall atlas is currently displayed in its current state of acquisition in the leftmost pane. Tiles in the atlas will fill up as each capture occurs.
  4. Select the grid square based on the ice thickness by navigating on the atlas (Supplemental Figure S5). Once the desired grid squares are selected, click on the Schedule button and observe the tiles in the grid square fill up as each grid square is captured.
  5. Click on the Schedule button once the grid squares are selected.
  6. Select a representative hole in the grid square by adding its position. Once a hole image is acquired, define the data and focus positions and save the layout as a template (Supplemental Figure S6).
  7. Click on Auto find, give the hole size (e.g., R1.2/1.3), and click on the Find button in the program, which will cause the Find program to automatically find the holes based on the hole diameter. Next, click on the Mark button to add the template (Figure 4) and add red circle marks in all the holes in one grid or partial grid square.
  8. Set up the intensity to remove the holes from the grid square and ice contamination (Figure 4).
    NOTE: Finally, the selected holes will be marked in yellow for scheduling the data collection.
  9. Click on the Schedule button in Latitude tasks after adding the holes through Auto find.
    ​NOTE: Before scheduling the automated data collection, ensure that the liquid nitrogen tank level is sufficient, the autoloader turbo pump is off, and the RAID drive space is free. Latitude-S task manager shows the number of atlas, grid square, hole, and data states scheduled for data collection (Figure 5). In Latitude-S GUI, various color schemes will be visible, and the meaning of the different color schemes will be displayed: 1. Yellow indicates unscheduled; 2. Green indicates scheduled; 3. Blue indicates Acquired; 4. Red indicates failed.

7. Cryo-EM data processing

NOTE: Cryo-EM image processing of spike protein is described in detail in recent literature25.

  1. Perform image processing of spike protein of SARS-CoV2 using RELION 3.011.
  2. Screen the movie images collected using Latitude-S manually, and perform beam-induced motion correction of the individual movies using MotionCor2 software9. Perform the initial screening of the motion-corrected micrographs manually with the help of cisTEM software package14.
    NOTE: Almost 85% of the automatically acquired micrographs were of good quality, and data had signal within 3.7-5.2 Å, which is calculated using cisTEM software14 (Supplemental Figure S7A,B).
  3. Process the data using the RELION 3.0 software package11.
    1. Pick spike particles manually and subject them to 2D class classification (Supplemental Figure S7C). Use the best 2D class as a reference to autopick 3,99,842 single-spike particles from the micrographs using the RELION autopick tool11.
      NOTE: Three rounds of 2D classification were carried out before subjecting the particles to 3D classification (Supplemental Figure S8). Approximately 2,55,982 single particles were selected for 3D classification, and the data set was classified into six classes. The final 3D auto-refinement was performed with the best class; 85,227 spike particles were obtained from the 3D classification.
    2. After auto refinement, perform per particle defocus refinement with proper beam tilt parameters for resolution improvement. Next, subject the particles to Bayesian polishing using the RELION 3.0 software package11. Finally, use the polished particle set for another round of 3D auto-refinement using RELION 3.011.

Subscription Required. Please recommend JoVE to your librarian.

Representative Results

In the current pandemic situation, cryo-EM plays a key role in characterizing the structures of various proteins from SARS-CoV-226,27,28,29, which may help develop vaccines and drugs against the virus. There is an urgent need for fast-paced research efforts with limited human resources to combat the coronavirus disease of 2019. Data acquisition in single-particle cryo-EM is a time-consuming but crucial step in the structure determination of macromolecules. Recent developments in cryo-EM automatic data acquisition have enabled limited human interference in data collection. The Latitude-S software is an important automatic data acquisition software package used here for the automated data collection of purified SARS-CoV2 spike protein.

Cryo-EM data acquisition of SARS-CoV-2 spike protein was performed with a 200 keV cryo-EM equipped with a K2 Summit DED. The locations for data acquisition on the grid with desirable ice thickness and particle distribution were marked manually. The positions were marked in parallel during data acquisition occurring in the background. At the marked positions, Latitude-S software carried out automated data acquisition at a nominal magnification of 42,200x at the pixel size of 1.17 Å at the specimen level. The configuration for data collection at 42,200x magnification was pre-set and tested already. A total of 40 frames were recorded for 8 s with the electron dose of 2 e2 per frame; thus, a total dose of 80 e2 was used for data collection (Supplemental Figure S9). The data were acquired at a defocus range of −0.75 µm and −2.25 µm, with3,000 movie files collected in two days. Every 4 h, periodic checks and adjustments were performed by the software to ensure that all the movie files collected over 48 h were of good quality and there was no beam shift or alignment shift. The data were collected independently without any human intervention. Additionally, Latitude-S automatically stops imaging at the time of liquid nitrogen filling, which reduces unnecessary vibration or mechanical drift in the images.

As mentioned in the protocol section, the initial screening of motion-corrected micrographs was performed manually using cisTEM software14. Based on the screening, most of the data were found to be within the signal range of 3.7-5.2 Å (Supplemental Figure S7A). This suggests that the automatic data collection using Latitude-S is good, and most of the data are suitable for high-resolution 3D reconstruction. Additionally, images were collected at defocus range (-0.75 to -2.25 µm), and various defocus ranges were manually checked by cisTEM14. The acquired data were very close to the setup defocus range in Latitude-S (Supplemental Figure S7A,B).

Data processing was performed using the RELION 3.0 software package11. Spike particles were manually picked to calculate the 2D class averages. Various structural details (helix and β-sheet) are visible in the 2D class averages (Supplemental Figure S7C), which strongly suggests that high-resolution structural characterization is possible using this data set. However, 3D classification also indicates that spike protein has 1-receptor-binding domain (RBD) up open and all RBD down close conformation (Supplemental Figure S8). The 3D classification indicates that class-1 has the maximum number of particles, which appear as a 1-RBD up open conformation. Furthermore, class-3 and class-4 have a similar number of particles, and both models appeared to have all RBD down close conformation. However, class-5 shows an intermediate conformation, where 1-RBD is in an intermediate position. However, the 1-RBD up open conformations of the spike protein were reconstructed using C1 symmetry, and the overall resolution is 4.4 Å (Figure 6 and Supplemental Figure S10). Similarly, all RBD down close conformation (class-3 and class-4) were refined together with C3 symmetry, and the overall resolution at 0.143 FSC is ~3.9 Å (Figure 7).

The overall image processing indicates that the spike protein adopts all RBDs down close and 1-RBD up open conformation. Additionally, identify an intermediate conformation of the spike protein was identified. The high-resolution cryo-EM structure of the S2 subdomain of the spike protein indicates the side chains of individual amino acid residues (Figure 6B and Figure 7C). All the 3D reconstructions and cryo-EM results are highly similar to findings in recently published literature25. However, the high-resolution cryo-EM structure of the spike protein was characterized within 15 days, which is only possible because of automatic cryo-EM data acquisition protocols and automatic particle-picking software. Therefore, automatic data acquisition software packages, including Latitude-S, can significantly contribute to the characterization of several high-resolution cryo-EM structures of biological macromolecules.

Figure 1
Figure 1: Workflow of automated data acquisition using Latitude-S: General steps to be followed before data collection (grid preparation, sample loading, and microscope tuning). Data acquisition is the main part of this manuscript, and the pipeline to be followed during data acquisition is highlighted. Abbreviations: cryo-EM = cryo-electron microscopy. Please click here to view a larger version of this figure.

Figure 2
Figure 2: Setup of different states for single-particle data collection using Latitude-S GUI. (A) Latitude-S software package for data acquisition in DM3 software suit. (B) Workflow of the data acquisition procedure. (C) Expansion of each panel. Abbreviation: GUI = graphical user interface. Please click here to view a larger version of this figure.

Figure 3
Figure 3: Fine alignment to set up high accuracy for focusing and image acquisition. Fine alignment is performed on five states a. Atlas, b. Hole, c. Data, d. Grid, e. Focus. Focus each state by placing the red mark on the same position. It is highly recommended to perform Fine Alignment before starting any new session, which will help perform imaging in a particular position. Image acquisition in a particular position (without any major shift) completely depends on the accuracy of the fine alignment. Please click here to view a larger version of this figure.

Figure 4
Figure 4: Automatic hole selection using Latitude-S. Auto find of holes for data acquisition is carried out automatically based on hole size. (A) Shows the position of the hole-finding vector for auto-finding the holes. Scale bar = 20 µm. (B) Shows the marking of holes using the hole finding vector and adjusting the intensity to remove the marker from the border area and ice contamination. Scale bars = 20 µm (left) and 10 µm (middle). (C) Shows the automatic adding of the holes for imaging (yellow). Scale bar = 20 µm. Please click here to view a larger version of this figure.

Figure 5
Figure 5: Live view of data acquisition. (A) The positions are marked in yellow, green, blue, and red based on the status of data acquisition in each position. (B) Color code for monitoring the status of data acquisition. Green: Scheduled, Yellow: Unscheduled, Blue: Acquire, Red: Failed. The left panel shows several holes colored green (scheduled) and a few holes marked blue (acquired); scale bar = 10 µm. The middle panel shows 4,300x magnification of an individual hole. In this image of a hole (middle panel), the blue square box shows the focus area, and the green square box shows the imaging region; scale bar = 1000 nm. The right panel shows the acquired image. The extreme right panel shows the scheduled image numbers, total time required for imaging, and how many images are scheduled for imaging. Please click here to view a larger version of this figure.

Figure 6
Figure 6: Spike 3D mode with 1RBD open. (A) Auto-refined and sharpened spike protein map of 1-RBD up open is represented in side, top, and bottom views. (B) EM map is fitted with crystal structure for better visualization of the side chains. The highlighted regions in the map have densities for side chains. Abbreviations: 3D = three-dimensional; RBD = receptor-binding domain; EM = electron microscopy; NTD = N- terminal domain; S1 = subunit 1; S2 = subunit 2. Please click here to view a larger version of this figure.

Figure 7
Figure 7: All RBD down close conformation of spike protein. (A) Auto-refined and sharpened spike protein map of all RBD down close conformation represented in side and top views. (B) EM map is fitted with crystal structure for better visualization of the side chains. The arrows show that the regions in the map have densities for side chains (C). Abbreviations: RBD = receptor-binding domain; EM = electron microscopy; NTD = N- terminal domain; S1 = subunit 1; S2 = subunit 2. Please click here to view a larger version of this figure.

Supplemental Figure S1: Latitude-S image acquisition GUI: Various microscope controllers (e.g., column valve open/close, screen insert/retract are controlled by the Latitude-S GUI. Column valve, camera, screen status, and liquid nitrogen filling could be controlled in the left panel. At the bottom of this panel, various calibration parameters appear green (e.g., Magnification, beam tilt, objective focus). If any parameter appears black, it indicates that the parameter is not optimized correctly. Therefore, all parameters should be optimized before starting any new session. Abbreviation: GUI = graphical user interface. Please click here to download this File.

Supplemental Figure S2: Representation of Atlas state. Different Atlas states showing grid squares and the type of ice formed. (A-F) Various Atlas sizes are also highlighted. (A, B, D, F) A thick ice pattern is highlighted. (C, D) Broken squares are highlighted. Thick ice and broken regions (marked in the figure) are not suitable for imaging. (E, F) Good grid squares for imaging; A, B, C, D, and F show the grid squares suitable for high-resolution imaging. However, thick ice grid squares and broken grid squares must be excluded. Scale bars = 100 µm (A-C), 50 µm (D, E), and 25 µm (F). Please click here to download this File.

Supplemental Figure S3: Representation of Grid state and Hole state. Grid state and corresponding Hole state are shown in the image. (A, E) Empty hole, (F, G) thick ice, (E) ice contamination, and (A, B, C, D, E, and G) suitable ice thickness is marked in the image. Suitable ice thickness holes are selected for the image acquisition (A, B, C, D, E, and G). Scale bars = 10 µm (A, E, F, G), 2 µm (B), 50 µm (C), 5 µm (D). Please click here to download this File.

Supplemental Figure S4: Hole reference for automatic imaging. Hole image (QUANTIFOIL-Holey Carbon grid QUANTIFOIL R 1.2/1.3) is captured and saved for future reference. The size of the hole reference could be varied based on the different types of grids. It is recommended to always capture the hole reference before starting any new session. Please click here to download this File.

Supplemental Figure S5: Image-capturing panel at various magnifications and hole selection. (A) The capture panel shows the settings for data acquisition. (B) The Latitude-S main navigation window shows three consecutive magnifications. In Atlas state (150x), the grid squares are selected for data acquisition (left-hand panel, scale bar = 50 µm). In higher magnification (380x), a single square is focused (middle panel, scale bar = 20 µm). Further higher magnification (4,300x), holes inside each square are focused (right-hand panel, scale bar = 5 µm). However, these magnifications would change according to the size and shape of the grids. Please click here to download this File.

Supplemental Figure S6: Creating a template for hole selection and imaging. Template generation is performed by adding a position on the hole for data, and the focus is positioned adjacent to the hole on the carbon surface. The focus should position on the carbon area so that the beam diameter should not touch any adjacent hole. Scale bars = 20 µm (left panel), 10 µm (middle panel), 1000 nm (right panel). Please click here to download this File.

Supplemental Figure S7: Cryo-EM imaging of SARS-CoV2 using Latitude-S and image screening. (A) Screening of acquired micrographs: 1D CTF fit, 2D CTF fit, and CTF parameters; estimation of the SARS-CoV2 spike protein data using cisTEM. 1D CTF fit and Thon ring shows that the overall signal is 4.8 Å. (B) Micrographs at two different defocus value of the spike protein are acquired using Latitude-S. Scale bars = 50 nm. (C) Final 2D class average. 2D class average shows top, bottom, and side views of the spike protein. All the high-resolution details are visible in 2D class averages. Abbreviations: cryo-EM = cryo-electron microscopy; 1D = one-dimensional; 2D = two-dimensional; CTF = contrast transfer function; SARS-CoV2 = severe acute respiratory syndrome coronavirus 2. Please click here to download this File.

Supplemental Figure S8: Data processing of SARS-CoV2 spike protein data acquired using Latitude-S software. The image shows the workflow followed for processing the cryo-EM data of spike protein. 3D classification of spike protein is performed using Relion 3.0. Class-1 shows the 1-RBD up open conformation. Class-3 and Class-4 show all RBD down close conformation of spike-protein. Class-5 shows the intermediate conformation of the spike protein. Abbreviations: cryo-EM = cryo-electron microscopy; 3D = three-dimensional; RBD = receptor-binding domain; SARS-CoV2 = severe acute respiratory syndrome coronavirus 2. Please click here to download this File.

Supplemental Figure S9: Dose-fractionated image captured as a reference for final image acquisition. Dose fractionation images are captured with an 8.0 s exposure and 0.2 s/frame exposure. Click on the Auto save button near the camera to save the movie files automatically. After image acquisition, click on the image and save the dose-fractionated image parameters using the image updated button in the data collection state. Please click here to download this File.

Supplemental Figure S10: Angular distribution and Fourier shell correlation of the generated SARS-CoV-2 1 RBD up open conformation spike protein map. (A) Angular distribution of the final 3D model of 1-RBD up open conformation of the spike protein. Blue represents lower values, and red represents higher values of the normalized particle distribution. (B) Fourier shell correlation curve showing 4.4 Å resolution of the 1-RBD up open conformation of spike protein, estimated at the cut-off. Abbreviations: 3D = three-dimensional; RBD = receptor-binding domain; SARS-CoV2 = severe acute respiratory syndrome coronavirus 2; FSC = Fourier shell correlation. Please click here to download this File.

Atlas Image State Atlas 115
Magnification 115, Pixel size 34.7 nm
Indicated Defocus 4.38 mm
Spot Size 8
Brightness 934400
Mode IMAGINGILM
Binning 1
Exposure Time 1.0 s
Grid Survey State Grid 380
Magnification 380, Pixel size 11.2 nm
Indicated Defocus 2.37 mm
Spot Size 8
Brightness 626200
Mode IMAGINGILM
Binning 1
Exposure Time 1.0 s
Hole Survey State Hole 3400
Magnification 3,400, Pixel size 1.20 nm
Indicated Defocus -0.75 µm
Spot Size 7
Beam Diameter 8.81 µm
Mode IMAGINGILM, SA, Mh
Binning 1
Exposure Time 1.0 s
Focus State Focus 45k
Magnification 45,000, Pixel size 0.0924 nm
Indicated Defocus 4.51 µm
Spot Size 6
Beam Diameter 0.716 µm
Mode IMAGINGILM, SA, Mh
Binning 1
Exposure Time 1.0 s
Acqisition State Data 45k
Magnification 45,000, Pixel size 0.0924 nm
Defocus Setup Min: -4,500 nm, Max: -1,500 nm, Steps: 250 nm
Spot Size 6
Beam Diameter 0.752 µm
Mode IMAGINGILM,SA,Mh
Binning 1
Exposure Time Total 8 s exposure for 20 frames 
Camera setup Counted, Gain Normalized, Defect Corrected
Data save setup MRC

Table 1: Latitude-S state setup summary.

Specification K2 Base K2 Summit
TEM operating voltage 200-400 kV
Sensor active area 19.2 mm × 18.6 mm
Sensor size in pixels 3838 × 3710 7676 × 7420 Super-
Resolution
Physic al pixel size 5 µm
Binning 1–8x
Sensor read-out Any arbitrary area
Magnification relative to film 1.3–1.5x
Sensor read-out speed 50 full fps 400 full fps
Transfer speed to computer 8 full fps 40 full fps
Image display 8 full fps 10 full fps
DQE performance (300 kV) >0.30 (peak) >0.25 at 0.5 of physical Nyquist >0.7 (peak) >0.50 at 0.5 of physical Nyquist >0.06 at 1.25 of physical Nyquist
Software Gatan Microscopy Suite including DigitalMicrograph

Table 2: Camera specifications.

Configuration Option Value
Microscope Type Talos Arctica G2
High Tension 200 kV
Source XFEG
Lens Cryo Twin
Vacuum System Talos TMP IGP
Sample Loader AutoLoader

Table 3: Microscope configuration.

Subscription Required. Please recommend JoVE to your librarian.

Discussion

Latitude-S is an intuitive user interface, which provides an environment to automatically set up and collect thousands of high-resolution micrographs or movie files in two days. It provides easy navigation across the grids and maintains the position of the microscope stage while it moves from low magnification to high magnification. Each step of data acquisition with Latitude-S is time-efficient, with features such as a simple user interface, fast streaming of data at up to 4.5 GB/s speed, and simultaneous display of data during acquisition. Additionally, precalibrated dose fractionation, dose rate, focus, and magnification parameters were easily reloaded to start a new session of automatic data acquisition and save time.

Acquiring data automatically in the absence of constant monitoring and without compromising the quality of the data set is a challenging and time-consuming task. Automated data collection using Latitude-S software is convenient when time and resources are major constraints, particularly during this pandemic. However, several cryo-EM structures of various proteins of SARS-CoV-2 have been resolved in the past few months, which will help pharmaceutical companies develop vaccines. Different laboratories use different types of automatic data collection software packages to collect data. We used Latitude-S with a K2 Summit DED for cryo-EM automatic data collection on holey carbon grids or homemade GO coated grids30.

That study was performed using the above-mentioned parameters in the protocol section, which provides strong evidence that Latitude-S is a suitable and ideal software package for automatic data collection for single-particle cryo-EM. However, it is highly recommended to follow certain protocols before starting the imaging using a K2 camera. The K2 direct detection camera requires basic maintenance routines to achieve the highest performance. Regular annealing of the camera to 50° for 24 h helps the sensor perform optimally by reducing the background noise and contamination at the surface level. However, after camera annealing, updating the gain and dark references of the camera is a mandatory step (takes ~45 min) before performing any image acquisition.

Although Latitude-S is a stable and user-friendly software package for cryo-EM automatic data collection, it is essential to optimize various parameters (magnifications, spot size, brightness, and dose rate) in 5 different states of Latitude-S at the beginning. The magnification of holes or grid states depends on the hole sizes of the holey grids or holey grid types (e.g., R2/2 or R1.2/1.3 or R 0.6/1). For example, the R0.6/1 type grid hole size is 0.6 µm, and the hole size of the R2/2 grid is 2 µm. Thus, two different types of magnification are required to visualize the holes properly for grid types R0.6/1 and R2/2 in the grid and hole states in Latitude-S.

Therefore, the magnification settings for various types of grids in 5 different states will be variable. The spot size and the brightness highly depend on the magnification. Hence, these values may change at different magnification values. Therefore, it is recommended to optimize the various parameters of the 5 different states of Latitude-S using different types of cryo-EM grids before starting automatic data acquisition. However, once all the parameters are optimized and saved, it is easy to reload all the parameters based on the user's requirement and use Latitude-S at different magnifications or grid types.

An important benefit of using K2 with Latitude-S is that users can easily regulate the open/close the beam valve, insert/retract the camera, insert/retract the phosphor screen of the microscope, and regulate liquid nitrogen filling using the GUI of Latitude-S. However, any other options (such as gun tilt, gun shift, beam shift, pivot points, C2 aperture centering, rotation center, coma-free alignment) are not accessible through the K2 Latitude-S GUI Tab (Supplemental Figure S1 and Figure 2). During the long hours of data collection, the position of the beam may shift.

Latitude-S can execute automated periodic checks and corrections to keep track of the system's stability throughout the data acquisition period. The stability of the system is maintained by centering the beam and updating the dark references. The constant check ensures the high quality of the acquired data. In Latitude-S, eucentric height (Z-height) is corrected only once before data collection, and the eucentric height is calculated automatically by Latitude-S when it changes the grid square. The focus is automatically measured and adjusted based on the user-defined focus range. The program will reset the stage Z position if it exceeds the given threshold value. This stability is controlled through the System Stability palette. However, like other automatic data acquisition packages, Latitude-S also has some limitations.

Latitude-S cannot calculate the eucentric height (Z-height) if the grid is uneven. In this scenario, it cannot collect any data, or the defocus values will be completely out of range. Therefore, users should be extremely cautious to prepare their grids without any bend and image only flat-surface grids using Latitude-S. Furthermore, unlike Leginon, SerialEM, and UCSF-Image, Latitude-S is not a freely available software package. Latitude-S is compatible with Gatan cameras, including filtered or standalone K2, K3, and K3 base direct detection cameras, as well as Rio and OneView cameras. Another important disadvantage for users is that it is not compatible with other popular DEDs such as Falcon DED. However, this is also true for EPU, another automatic data acquisition software package, which is available with cryo microscopes and only compatible with the Falcon camera. However, EPU is also functional with K2/K3 with an energy filter (BioQuantum K3 Imaging Filter) but not with a standalone K2/K3 camera.

Latitude-S is quite similar to EPU, SerialEM, AutoEM, AutoEMation, and Leginon, which are software packages used for automatic data acquisition for single-particle cryo-EM. However, Latitude-S is only compatible with K2 DED, K3 DED, or BioQuantum K3 Imaging Filter. Additionally, continuous technical support is provided by the company for Latitude-S users. This technical support is beneficial for small user groups, who need to use the K2 DED, K3 DED, or BioQuantum K3 Imaging Filter devices for data acquisition and have no prior knowledge on how to set up or use free software packages such as SerialEM and Leginon.

There are many other features, such as microcrystal electron diffraction (microED), tomography, and energy-dispersive X-ray spectrometry (EDS), which are available in various other versions of Latitude. Therefore, users can use the same software package for data collection in other modes. To our knowledge, data collection for microED, tomography, and EDS is not available in EPU or any other software packages. Therefore, this Latitude software package could be useful for different purposes in addition to automatic data acquisition in single-particle cryoEM. However, SerialEM and Leginon, both free software packages, are suitable for Falcon or K2/K3 cameras and are extremely useful for new users. However, Latitude-S is not freely available, which might be a disadvantage of this software package.

In summary, the Latitude-S automatic data acquisition tool is as good as other automatic data acquisition software packages (e.g., EPU, Leginon, SerialEM, UCSF-Image). Latitude-S is an extremely stable and user-friendly data acquisition software package, which is available with filtered or standalone K2, K3, and K3 base direct detection cameras, as well as Rio and OneView cameras.

Subscription Required. Please recommend JoVE to your librarian.

Disclosures

The authors have no competing or financial conflicts of interest to declare.

Acknowledgments

We acknowledge Department of Biotechnology, Department of Science and Technology (DST) and Science, and Ministry of Human Resource Development (MHRD), India, for funding and the cryo-EM facility at IISc-Bangalore. We acknowledge DBT-BUILDER Program (BT/INF/22/SP22844/2017) and DST-FIST (SR/FST/LSII-039/2015) for the National Cryo-EM facility at IISc, Bangalore. We acknowledge financial support from the Science and Engineering Research Board (SERB) (Grant No.-SB/S2/RJN-145/2015, SERB-EMR/2016/000608 and SERB-IPA/2020/000094), DBT (Grant No. BT/PR25580/BRB/10/1619/2017). We thank Ms. Ishika Pramanick for preparing cryo-EM grids, cryo-EM data collection, and preparing the Table of Materials. We also thank Mr. Suman Mishra for cryo-EM image processing and for helping us to prepare the figures. We thank Prof. Raghavan Varadarajan for helping us to obtain the purified spike protein sample for this study.

Materials

Name Company Catalog Number Comments
Blotting paper Ted Pella, INC. 47000-100 EM specimen preparation item
Capsule Thermo Fisher Scientific 9432 909 97591 EM specimen preparation unit
Cassette Thermo Fisher Scientific 1020863 EM specimen preparation unit
C-Clip Thermo Fisher Scientific 1036171 EM specimen preparation item
C-Clip Insertion Tool Thermo Fisher Scientific 9432 909 97571 EM specimen preparation tool
C-Clip Ring Thermo Fisher Scientific 1036173 EM specimen preparation item
EM grid (Quantifoil) Electron Microscopy Sciences Q3100AR1.3 R 1.2/1.3 300 Mesh, Gold
Glow discharge Machine Quorum N/A Quorum GlowQube glow discharge machine
K2 DED Gatan Inc. N/A Cryo-EM data collection device (Camera)
Latitude S Software Gatan Inc. Imaging software
Loading station Thermo Fisher Scientific 1130698 EM specimen preparation unit
Talos 200 kV Arctica Thermo Scientific™ N/A Cryo-Electron Microscope
Vitrobot Mark IV Thermo Fisher Scientific N/A EM specimen preparation unit

DOWNLOAD MATERIALS LIST

References

  1. Li, Y., Cash, J. N., Tesmer, J. J. G., Cianfrocco, M. A. High-throughput cryo-EM enabled by user-free preprocessing routines. Structure. 28 (7), 858-869 (2020).
  2. Carragher, B., et al. Leginon: An automated system for acquisition of images from vitreous ice specimens. Journal of Structural Biology. 132 (1), 33-45 (2000).
  3. Stagg, S. M., et al. Automated cryoEM data acquisition and analysis of 284 742 particles of GroEL. Journal of Structural Biology. 155 (3), 470-481 (2006).
  4. Frank, J. Single-particle reconstruction of biological macromolecules in electron microscopy-30 years. Quarterly Reviews of Biophysics. 42 (3), 139-158 (2009).
  5. Biyani, N., et al. Focus: The interface between data collection and data processing in cryo-EM. Journal of Structural Biology. 198 (2), 124-133 (2017).
  6. Nakane, T., et al. Single-particle cryo-EM at atomic resolution. Nature. 587 (7832), 152-156 (2020).
  7. Kühlbrandt, W. The resolution revolution. Science. 343 (6178), 1443-1444 (2014).
  8. McMullan, G., Chen, S., Henderson, R., Faruqi, A. R. Detective quantum efficiency of electron area detectors in electron microscopy. Ultramicroscopy. 109 (9), 1126-1143 (2009).
  9. Zheng, S. Q., Palovcak, E., Armache, J. P., Verba, K. A., Cheng, Y., Agard, D. A. MotionCor2: Anisotropic correction of beam-induced motion for improved cryo-electron microscopy. Nature Methods. 14 (4), 331-332 (2017).
  10. Grant, T., Grigorieff, N. Measuring the optimal exposure for single particle cryo-EM using a 2.6 Å reconstruction of rotavirus VP6. eLife. 4, 06980 (2015).
  11. Scheres, S. H. W. RELION: Implementation of a Bayesian approach to cryo-EM structure determination. Journal of Structural Biology. 180 (3), 519-530 (2012).
  12. Grigorieff, N. FREALIGN: High-resolution refinement of single particle structures. Journal of Structural Biology. 157 (1), 117-125 (2007).
  13. Punjani, A., Rubinstein, J. L., Fleet, D. J., Brubaker, M. A. CryoSPARC: Algorithms for rapid unsupervised cryo-EM structure determination. Nature Methods. 14 (3), 290-296 (2017).
  14. Grant, T., Rohou, A., Grigorieff, N. CisTEM, user-friendly software for single-particle image processing. eLife. 7, 35383 (2018).
  15. Tang, G., et al. EMAN2: An extensible image processing suite for electron microscopy. Journal of Structural Biology. 157 (1), 38-46 (2007).
  16. Zhang, P., Beatty, A., Milne, J. L. S., Subramaniam, S. Automated data collection with a Tecnai 12 electron microscope: Applications for molecular imaging by cryomicroscopy. Journal of Structural Biology. 135 (3), 251-261 (2001).
  17. Lei, J., Frank, J. Automated acquisition of cryo-electron micrographs for single particle reconstruction on an FEI Tecnai electron microscope. Journal of Structural Biology. 150 (1), 69-80 (2005).
  18. Potter, C. S., et al. Leginon: A system for fully automated acquisition of 1000 electron micrographs a day. Ultramicroscopy. 77 (3-4), 153-161 (1999).
  19. Mastronarde, D. N. Automated electron microscope tomography using robust prediction of specimen movements. Journal of Structural Biology. 152 (1), 36-51 (2005).
  20. Suloway, C., et al. Automated molecular microscopy: The new Leginon system. Journal of Structural Biology. 151 (1), 41-60 (2005).
  21. Korinek, A., Beck, F., Baumeister, W., Nickell, S., Plitzko, J. M. Computer controlled cryo-electron microscopy - TOM2 a software package for high-throughput applications. Journal of Structural Biology. 175 (3), 394-405 (2011).
  22. Shi, J., Williams, D. R., Stewart, P. L. A Script-Assisted Microscopy (SAM) package to improve data acquisition rates on FEI Tecnai electron microscopes equipped with Gatan CCD cameras. Journal of Structural Biology. 164 (1), 166-169 (2008).
  23. Marsh, M. P., et al. Modular software platform for low-dose electron microscopy and tomography. Journal of Microscopy. 228, Pt 3 384-389 (2007).
  24. Zhang, J., et al. JADAS: A customizable automated data acquisition system and its application to ice-embedded single particles. Journal of Structural Biology. 165 (1), 1-9 (2009).
  25. Pramanick, I., et al. Conformational flexibility and structural variability of SARS-CoV2 S protein. Structure. , (2021).
  26. Zhou, D., et al. Structural basis for the neutralization of SARS-CoV-2 by an antibody from a convalescent patient. Nature Structural and Molecular Biology. 27 (10), 950-958 (2020).
  27. Hillen, H. S., et al. Structure of replicating SARS-CoV-2 polymerase. Nature. 584 (7819), 154-156 (2020).
  28. Wrapp, D., et al. Cryo-EM structure of the 2019-nCoV spike in the prefusion conformation. Science. 367 (6483), 1260-1263 (2020).
  29. Thoms, M., et al. Structural basis for translational shutdown and immune evasion by the Nsp1 protein of SARS-CoV-2. Science. 369 (6508), 1249-1255 (2020).
  30. Kumar, A., Sengupta, N., Dutta, S. Simplified approach for preparing graphene oxide tem grids for stained and vitrified biomolecules. Nanomaterials. 11 (3), 1-22 (2021).

Tags

Automated Data Acquisition Software Single-particle Cryo-electron Microscopy Structural Characterization Biological Macromolecules Pathogen Biology Mycobacterium HIV SASCO 2 Latitude-S Software Digital Micrograph Technique Manager Session Settings Folder Selection TEM Condition Camera Condition Image Or Stock Option Atlas State Palette Magnification Illumination Condition
User-friendly, High-throughput, and Fully Automated Data Acquisition Software for Single-particle Cryo-electron Microscopy
Play Video
PDF DOI DOWNLOAD MATERIALS LIST

Cite this Article

Kumar, A., P., S., Gulati, S.,More

Kumar, A., P., S., Gulati, S., Dutta, S. User-friendly, High-throughput, and Fully Automated Data Acquisition Software for Single-particle Cryo-electron Microscopy. J. Vis. Exp. (173), e62832, doi:10.3791/62832 (2021).

Less
Copy Citation Download Citation Reprints and Permissions
View Video

Get cutting-edge science videos from JoVE sent straight to your inbox every month.

Waiting X
Simple Hit Counter