To use DCAIS correctly, the system must be calibrated. There are several calibration methods that can affect the accuracy of the implant placement. This study aimed to assess the potential impact of different calibration methods on the accuracy of DCAIS.
Based on the interventions performed so far, the use of DCAIS allows a high-precision implant placement. In our early studies, we compared 41 clip calibrated dynamic navigated implant placements with 17 tracer calibrated dynamic navigated implant placements.
According to our early data (Table 1, Table 2, Table 3, Figure 7, Figure 8, Figure 9, and Figure 10), when using the two calibration methods, the results showed that there is no significant correlation between the platform and angular deviation in buccolingual (BL) and mesiodistal (MD) directions. Comparing the planned and final position of the implants, the calibration with a clip proved more accurate compared to one made with a tracer, but the difference is not significant (Table 1, Table 2, Table 3, Figure 7, Figure 8, Figure 9, and Figure 10). Based on previously published data by Block et al., implant placement with a dynamic navigation system allows high-precision implant placement1. The accuracy of the intervention can be improved by training8.

Figure 1: Calibration with clip. Calibration by holding the tip of the measured drill to the surface of the labeled clip. Please click here to view a larger version of this figure.

Figure 2: Drill calibration. Measuring the drill length by touching the drill to the go plate. Please click here to view a larger version of this figure.

Figure 3: Drilling process in the mouth. Drilling the bone under dynamic navigation control. Please click here to view a larger version of this figure.

Figure 4: Real-time live view of the drilling process on the monitor. Real-time control view of bone drilling. Please click here to view a larger version of this figure.

Figure 5: Real-time live view of the drilling process on the monitor. Real-time control view of bone drilling. Please click here to view a larger version of this figure.

Figure 6: Calibration with tracer. Identifying the selected anatomical points in the mouth by touching them with a probe tool. Please click here to view a larger version of this figure.

Figure 7: Mean deviation of the measured values (difference between the planned and final position of the implants) using the two different calibration methods. Global Platform deviation (mm): spatial distance between the center of the implant platform of planned and placed implants. Platform B/L deviation (mm): spatial distance between the center of the implant platform of planned and placed implants in buccolingual dimensions. Platform M/D deviation (mm): spatial distance between the center of the implant platform of planned and placed implants in mesiodistal dimensions. Platform depth deviation (mm): spatial distance between the center of the implant platform of planned and placed implants in depth dimensions. Platform non-depth deviation (mm): the resultant of the Platform B/L and M/D deviations. Apical non-depth deviation (mm): the resultant of the apical B/L and M/D deviations. Global Apical deviation (mm): spatial distance between the center of the implant apex of planned and placed implants. Apical B/L deviation (mm): spatial distance between the center of the implant apex of planned and placed implants in buccolingual dimensions. Apical M/D deviation (mm): spatial distance between the center of the implant apex of planned and placed implants in mesiodistal dimensions.Apical depth deviation (mm): spatial distance between the center of the implant apex of planned and placed implants in depth dimensions. Please click here to view a larger version of this figure.

Figure 8: Standard deviation of the measured values. The difference between the planned and final position of the implants using the two different calibration methods. Please click here to view a larger version of this figure.

Figure 9: Standard error of mean deviation of the measured values. The difference between the planned and final position of the implants using the two different calibration methods. Please click here to view a larger version of this figure.

Figure 10: Analysis of the measured values. The difference between the planned and final position of the implants using the two different calibration methods. Please click here to view a larger version of this figure.
| Mean deviation |
| Global Platform (mm) | Platform B/L Deviation (mm) | Platform M/D deviation (mm) | Platform depth deviation (mm) | Platform non-depth deviation (mm) | Apical non-depth deviation (mm) | Global Apical (mm) | Apical B/L Deviation (mm) | Apical M/D deviation (mm) | Apical depth deviation (mm) |
| clip | 1.68 | 0.14 | -0.24 | 0.53 | 1.1 | 1.29 | 1.81 | 0.18 | 0,00 | 0.45 |
| tracer | 1.99 | 0.11 | 0.32 | 0.86 | 1.21 | 1.62 | 2.28 | 0.31 | 0.43 | 0.86 |
Table 1: Mean deviation of the measured values. The difference between the planned and final position of the implants using the two different calibration methods.
| Standard deviation |
| Global Platform (mm) | Platform B/L Deviation (mm) | Platform M/D deviation (mm) | Platform depth deviation (mm) | Platform non-depth deviation (mm) | Apical non-depth deviation (mm) | Global Apical (mm) | Apical B/L Deviation (mm) | Apical M/D deviation (mm) | Apical depth deviation (mm) |
| clip | 1.03 | 0.79 | 1.14 | 1.29 | 0.89 | 1.16 | 1.22 | 0.79 | 1.52 | 1.26 |
| tracer | 0.84 | 0.94 | 1.3 | 1.3 | 0.94 | 1.23 | 1.07 | 1.12 | 1.61 | 1.27 |
Table 2: Standard deviation of the measured values. The difference between the planned and final position of the implants using the two different calibration methods.
| Standard error of the mean deviation |
| Global Platform (mm) | Platform B/L Deviation (mm) | Platform M/D deviation (mm) | Platform depth deviation (mm) | Platform non-depth deviation (mm) | Apical non-depth deviation (mm) | Global Apical (mm) | Apical B/L Deviation (mm) | Apical M/D deviation (mm) | Apical depth deviation (mm) |
| clip | 0.16 | 0.12 | 0.18 | 0.2 | 0.14 | 0.18 | 0.19 | 0.12 | 0.24 | 0.2 |
| tracer | 0.2 | 0.23 | 0.32 | 0.32 | 0.23 | 0.3 | 0.26 | 0.27 | 0.39 | 0.31 |
Table 3: Standard error of mean deviation of the measured values. The difference between the planned and final position of the implants using the two different calibration methods.
| Dynamic navigation implantation systems |
| advantage (+) | disadvantage (-) |
| · Very precise implant placement | · A system failure that interferes with the spatial relationship between the reference points and the patient can lead to errors in implant bed design and implant positioning |
| · Less invasive, shorter healing time, fewer complaints | · Longer training period required to use the system correctly |
| · Less risk of complications (e.g. nerve damage) | · Costly |
| · Easy to use in small mouth openings and in the molar region | |
| · Does not require a separate surgical instrument set | |
| · Efficient use of time, planning and surgery can be performed on the same day | |
| · Possibility to change the position and size of the previously planned implants during surgery | |
| · Can also be used in narrow interdental spaces | |
Table 4: Advantages and disadvantages of dynamic navigated implant systems.