All animals received humane care in compliance with the National Health and Medical Research Council (Australia) guidelines. All animal procedures were approved by the Animal Ethics Committee of the Garvan Institute of Medical Research (Sydney, Australia).
1. Preparation of Krebs-Henseleit (KH) buffer
- Prepare 2 L of a 20x stock solution of KH buffer containing 2.36 M NaCl, 0.094 M KCl, 0.024 M MgSO4, 0.024 M KH2PO4. Aliquot into 50 mL tubes and store at -20 °C.
- Prior to the experiment, prepare two 2 L Erlenmyer flasks of 1x KH buffer containing 1900 mL of deionized water, 100 mL of 20x stock KH buffer, 4.20 g of NaHCO3 (25 mM), and 3.96 g of glucose (11 mM).
- Place buffer flasks in a 37 °C water bath and gas the solution with carbogen (95% oxygen, 5% carbon dioxide, flow rate set to 2 L/min) via air stones (slip-on inlet filter) for 5 min, then add 2.8 mL of 1 M CaCl2 (1.4 mM). Continue bubbling the solution for at least 1 h prior to the start of the experiment. Use of an air stone assists with uniform aeration of the buffer.
- Place a small volume (~50 mL) of buffer into a round metal dish at -20 °C to cool and form a layer of iced buffer.
2. Preparation of perfusion circuit
- Ensure all tubing and glassware are clean with no visible growth/contamination.
- Set up heights of glass chambers to correspond to required perfusion pressures based on 1 cm H2O being equivalent to 1.36 mmHg: Langendorff-to-heart 100 cm (approximately 75 mmHg), pre-load working side-to-heart 15 cm (approximately 11 mmHg), heart-to-afterload chamber 100 cm (approximately 75 mmHg).
- Start circulation of hot water around the circuit at least 10 min before heart perfusion.
- Place microfiber filters (1.2 µm) into filter holders and secure them tightly.
- Prime the circuit with KH buffer, ensuring all air is removed, and use a flow control tubing clamp to adjust the flow on the Langendorff side (Figure 1A, B) to ~50 mL/min. The slower flow allows for ease of aortic cannulation and prevents the heart from being dislodged before tying. Place a thermometer within the circuit via a T-shaped tubing connector to monitor the temperature of the buffer at ~37 °C.
3. Setting up of Lab Chart (data analysis) software
- Ensure Powerlab (data acquisition system) and flowmeter are switched on and connected to the computer prior to opening the data analysis software.
- Open a new file.
- Set up data collection by selecting the Settings drop-down menu and selecting Channel Settings. Select for Aortic Pressure, Aortic Flow, and Heart Rate (Figure 2A).
- Set up each channel with the settings shown in Figure 2B.
4. Preparation of animal for rat heart isolation
NOTE: The methodology for rat heart isolation and perfusion for non-DBD and non-DCD hearts for cold static storage is outlined below. The methods for heart instrumentation onto the perfusion circuit are relatively similar regardless of which donation method is used. The differences in the protocol for DBD and DCD rat heart donation primarily occur from the time the animal anesthesia has taken effect to when the heart is excised from the animal.
- Weigh the animal, ensuring that its weight is >330 g. The ideal weight range for male rats is 330-420 g.
- Administer an intraperitoneal injection of ketamine (80 mg/kg) and xylazine (10 mg/kg). Ensure adequate surgical plane of anesthesia based on the absence of a pedal reflex (toe-pinch reflex) and eye blink reflex.
- Place the animal in a supine position, securing the limbs with tape. Administer oxygen to the animal via a nose cone.
- Have the KH buffer ice-slush metal container and a small metal container with ice at the dissecting station. Soak a 7.5 cm x 7.5 cm gauze in KH buffer and place it on the container with ice.
- Once adequate surgical plane of anesthesia is confirmed, shave the skin at the surgical incision site and clean the skin with aseptic such as betadine or 70% alcohol. Make a lateral incision across the abdomen (laparotomy) and move the stomach and intestines to the side to allow access to the left renal vein. Inject 1500 IU heparin and apply pressure to the injection site until the bleeding stops. Allow 1-2 min for heparin to circulate before continuing with heart retrieval.
- Open the chest cavity by cutting through the rib cage on either side and cutting across the diaphragm. Carefully hold and lift the heart-lung bloc while cutting along the descending aorta, ensuring that the aortic arch remains intact. This will allow for enough aorta length to remain for cannulation.
- The rat is euthanized by the removal of the heart and lungs while under anesthesia. Submerge the heart-lung bloc in the KH buffer ice slush to allow the heart to arrest.
5. Cannulating the heart onto the isolated working heart perfusion circuit
- Transfer the heart-lung bloc to a gauze on ice. Dissect excess fat around the aorta, trim the aorta to leave 3-5 mm length for cannulation, and make a small incision ~1-2 mm in the pulmonary artery. This will allow a path for the coronary effluent.
- Place the trimmed heart-lung bloc (while still on ice) directly underneath the cannulas. Adjust the flow from the Langendorff side using a flow control tubing clamp to facilitate aortic cannulation. Carefully cannulate the aorta onto the left cannula, hold it with a black clamp, and tie it with a 2-0 silk suture.
- Increase flow from Langendorff perfusion by completely opening the flow control tubing clamp. Open the afterload clamp. Ensure that there is continuous flow from the Langendorff chamber. Start monitoring heart activity with software.
NOTE: if the aortic flow value on the Flowmeter display is <10 mL/min, ensure that the pulmonary arteriotomy is of sufficient size. A low aortic flow trace/value could also indicate that the aortic cannula has been inserted too far.
- Adjust the heart-lung bloc so that the anterior surface of the lungs is facing the operator. Tie the left lung lobe with a 2-0 silk suture and cut off excess left lung tissue. Tie the remaining lung lobes in one sweep with a 2-0 silk suture and cut the tissue. Rotate the heart so that the tied lungs are towards the back.
- Using small Vannas scissors, cut a 2-3 mm edge of the left atrial appendage, ensuring that a small opening into the atria is achieved. Carefully cannulate the left atrial appendage to the left cannula, hold it with a black clamp, and tie it with a 2-0 silk suture. Figure 3A shows the placement of both cannulas.
- Adjust the heart chamber water jacket so that the heart sits within the center of the chamber.
- Open the pressure transducer to air and zero the pressure transducer in the software by selecting the Aortic Pressure drop-down menu > Bridge Amp > Zero. Click OK.
- Perfuse the heart in Langendorff mode for 10 min. Ensure that the Langendoff aortic flow measures between 10-20 mL/min (ideal 20 mL/min).
- Prior to switching the working mode, ensure that the air stone (slip-on inlet filter) from the Langendorff KH buffer reservoir is placed in the working KH buffer reservoir.
- Switch the heart into working mode by closing the large clamp leading to the aortic cannula and opening the flow leading to the left atrial appendage. Open three-way tap leading from working perfusate. Transfer the common return tubing (taking KH buffer back to the reservoir) from the Langendorff reservoir to the working reservoir.
- Perfuse the heart in working mode for 15 min. Aortic flow values must ideally be >30 mL/min at the end of 15 min perfusion. Once the heart has stabilized, start recording data in data analysis software.
- Measure coronary effluent (coronary flow) at 1 min, 5 min, 10 min, and 15 min.
NOTE: Coronary flow values should be between 10-20 mL/min. Coronary flow greater than 22 mL/min generally indicates a leak originating from an untied pulmonary vessel. Heart rate should be between 200-300 bpm. Coronary effluent can be collected for downstream analysis of tissue injury markers (e.g., troponin-I, lactate dehydrogenase release).
6. Administering cardioplegia for cold static storage of the heart
- Fill the cardioplegia chamber (positioned at a height of 60 cm to yield perfusion pressure of approximately 40 mmHg) with 50 mL of designated cardiac preservation solution (with or without supplementation).
- Lower the water jacket and close the three-way tap on the working mode side.
- Stop the flow towards the left atrial cannula by closing the clamp.
- Open the flow to the aortic cannula.
- Close the clamp from the Langendoff side and on the afterload.
- Open the clamp from the cardioplegia line and flush 15 mL of cardioplegia out from the pressure line. Close the tap at the pressure line.
- Allow the heart to be flushed with cardioplegia via the aortic cannula and collect the coronary effluent. Run cardioplegia for 3 min and record the volume of cardioplegia administered.
- Stop recording data in the software and save the file.
- After 3 min, close the cardioplegia clamp and the small blue clamp leading to the aortic cannula.
- Place a bulldog clamp on the aortic cannula and a second bulldog clamp on the left cannula. Disconnect the heart-cannula block from the circuit and place it in a beaker containing cardioplegia in an ice box (Figure 3B, C).
- Store the heart in the fridge for 6 h (or desired cold storage time).
7. Heart reperfusion after cold static storage
- Prepare KH buffer as per section 1.
- Prime the perfusion circuit as per section 2.
- Open the saved file from baseline perfusion.
- Carefully remove the heart-cannula block from the ice box and secure it onto the perfusion circuit. Before reconnecting the aortic cannula, use a blunt-end 18 G syringe filled with KH buffer to remove any air from the cannula. Connect the aortic cannula first, open the tubing clamp on the Langendorff side, and ensure the afterload tubing clamp is open.
- Reconnect the left atrial cannula, ensuring that air has been removed from the cannula before connecting to the perfusion circuit. Keep the working side clamp closed until it is time to switch from Langendorff to working mode.
- Perfuse the heart in Langendorff mode for 15 min. Ensure to start recording data at the start of Langendorff reperfusion.
- Prior to switching to the working mode, place both bubbling stones in the KH buffer, open the 3-way tap on the working side, and transfer common return tubing to the working reservoir.
- Close the white Langendorff clamp, open the white working side clamp, and perfuse the heart for 30 min. Measure and collect coronary effluent at designated time points for downstream analysis (e.g., cardiac injury markers). Ensure to stop-restart software recording to start a new segment (or alternatively, add a comment).
- At approximately 15 min working mode reperfusion, close the three-way tap.
- At the end of reperfusion, save the data file, stop the flow through the circuit, and disconnect the heart block. Collect left ventricle samples for histological processing and/or snap frozen for future studies (e.g., protein extraction for western blot studies).
- Remove the filters from the filter holders, drain the buffer from the circuit, and flush twice with 5 L of distilled water.
NOTE: Although not necessary, it does help to pre-warm the distilled water in a large container within the water bath.
8. Analysis of functional recovery
- Open the LabChart file.
- Highlight a region of at least 30 s of recording, then click Command > Add data selection OR Multiple add to data. This will add the desired parameters to a DataPad file within the software.
- Select Window drop-down menu > DataPad.
- Highlight and copy the recovery values from DataPad to a spreadsheet for functional recovery analysis. Calculate functional recovery as a percentage of its respective baseline value1,6,7.