High-resolution imaging of the pulmonary structure is an essential part of diagnostic work-ups for many pulmonary conditions. Typically, this is performed using Computed Tomography (CT) imaging, which is ideally suited to generate high-resolution images of the lungs1. However, CT imaging delivers a non-trivial dose of ionizing radiation, making it ill-suited for regular repeat imaging, imaging at multiple different respiratory phases, or imaging certain populations (e.g., pediatrics). Magnetic resonance imaging (MRI) does not carry the same risk of ionizing radiation, and thus is amenable to such imaging tasks. However, it is challenging to image the lungs using MRI owing to low tissue density, respiratory and cardiac motion, and very fast signal relaxation2,3,4.
One MRI technique that is able to mitigate these challenges is ultra-short echo time (UTE) MRI4,5,6. In UTE MRI, the MRI signal is sampled immediately following signal excitation, which reduces the impact of fast signal relaxation. Moreover, this technique samples k-space from the center outward, which leads to significant oversampling at the center of k-space. This oversampling at the center of the k-space makes this imaging technique robust to motion. In addition to this inherent robustness to motion, repeated sampling of the center of k-space encodes information about respiratory motion, which enables the self-gating of images7,8,9. This self-gating can be used to generate images at a variety of respiratory phases. Because humans spend the majority of the respiratory phase at expiration, it is common to generate an image for end-expiration, as this phase has the most imaging data acquired.
There are a variety of strategies for respiratory self-gating in pulmonary MRI. The first distinction to be made is image-based vs. k-space-based gating10 (Figure 1). In image-based gating, a set of images with high temporal resolution is generated by reconstructing small temporal subsets of the imaging data. Subsequently, the position of the diaphragm in these images is used to identify the respiratory phase for a given set of image projections10,11. In k-space-based gating, data from the center of k-space ("k0") is examined8,9,12. The signal intensity of the image is encoded in k0, and thus, the intensity of the k0 point varies with respiration. Projections can thus be binned into different respiratory phases based on the intensity of k0. In both image-based and k-space-based gating, projections with like-respiratory phases are grouped for image reconstruction. It has been suggested that image-based gating provides improved fidelity in estimating the respiratory phase, thereby providing images with reduced blurring10,13.

Figure 1: Image-based and k-space based self gating techniques. (A) In image-based gating, low spatial resolution, high temporal resolution images showing the diaphragm are generated from temporal subsets of the overall data. Using a line over the diaphragm, respiratory motion can be visualized and binned for image reconstruction. (B) In k-space-based gating, the first point on a center-out k-space projection ("k0") is used to visualize respiratory motion. After smoothing k0, signal intensity differences based on the respiratory cycle are clearly visible and can be used to identify different respiratory phases. Please click here to view a larger version of this figure.
Both image and k-space-based gating can be performed using either hard gating or soft gating11,14. In hard gating, only the projections corresponding to the desired respiratory phase are reconstructed. However, this discarding of unwanted projections can lead to reduced image signal-to-noise ratio (SNR) and increased undersampling artifacts. These undesired effects can be mitigated by using soft gating. In soft gating, all projections are used for image reconstruction, but projections from an unwanted respiratory phase are weighted such that they have a lesser impact on the final image. In doing so, images can be reconstructed with minimal artifacts and high SNR while still suppressing the impact of respiratory motion.
Through the combination of UTE MRI acquisition with post-acquisition self-gating, high-quality images can be generated that, while not equivalent to CT, have a contrast and resolution that is approaching that of CT imaging6,15,16,17,18,19. Herein, a simple protocol is provided for collecting and reconstructing UTE MRI images to generate high quality images of pulmonary structure.
This protocol is written primarily for 3T MRI scanners; 3T is the most common field strength used for research MRI. Lower magnetic field strengths such as 1.5T or the recently available 0.55 T20 can provide improved image quality and signal intensity within the lungs, as signal relaxation within the lungs is slower at these field strengths.
While every attempt has been made to provide clarity and simplicity in this protocol and the provided image reconstruction code, the protocol will likely require a dedicated MRI physicist (or similar MRI expert) to establish an appropriate UTE MRI sequence on the MRI scanner. The MRI sequence should implement a 3D non-Cartesian encoding strategy with Center-out k-space trajectories. Examples include 3D radial or 3D spiral (e.g., "FLORET")21,22 imaging sequences. Importantly, the order of projections should have good temporal stability: Over any given subset of time, the projections should cover the full range of k-space23. Examples of projection ordering strategies with good temporal stability are golden means or Halton-randomized Archimedean spiral. If a projection ordering with poor temporal stability is used, post-acquisition self-gating will omit large regions of k-space, leading to image artifacts. Finally, the sequence should be capable of achieving an echo time (TE) of <100 µs. The T2* relaxation time in the lungs at 3T is <1 ms24, so using a very short TE is essential to generating high-quality images.