$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
Spectral imaging is a technique that provides both spectral and spatial information and has been applied to various fields, from microscopy1 to remote sensing2. Multispectral and hyperspectral imaging are distinguished based on whether a few discrete spectral bands or a continuous spectrum are recorded for each pixel of a 2D sensor. Raman imaging is a particular implementation of spectral imaging.
Raman scattering microscopy is a powerful technique to measure vibrational frequencies νR of a system. Vibrational modes are associated with the chemical bonds, and hence Raman spectroscopy3 enables the investigation of the chemical composition of samples in biological4 and materials sciences5. In spontaneous Raman spectroscopy, the sample is illuminated with monochromatic radiation at frequency ν0, called the pump. A fraction of the pump light, interacting with the vibrational modes, may undergo inelastic scattering, resulting in photons at frequencies νS (νAS)=ν0-(+) νR, where νR is the molecular vibrational frequency within the sample. The lower-energy scattered light (νS) is called Stokes radiation, while the higher-energy one (νAS) is the anti-Stokes radiation. At thermal equilibrium, most molecules occupy lower energy states, hence the anti-Stokes components are significantly weaker than the Stokes one. For this reason, Raman scattering microscopy typically detects Stokes photons. As molecules and solids have multiple vibrational modes, the Raman spectrum is characterised by distinct bands that collectively form a unique chemical signature, enabling material identification and analysis. The primary challenge of this approach is the extremely low cross-section of spontaneous Raman scattering, with only one in over 109-1012 photons undergoing this process.
In a standard configuration of a Raman scattering microscope, the excitation light is tightly focused on one point of the sample. After rejecting the elastically scattered pump photons (typically with spectral filters), the remaining light is collected and characterised with a frequency-domain spectrometer (e.g., dispersive grating); the spectral resolution of these devices can reach ~0.5-1 cm-1. The measurement is repeated for each point (x,y) in the field of view (FOV) of a sample by a method called point or raster scanning6. There are two main issues of this approach: (1) the low cross-section of spontaneous Raman scattering typically requires long integration times (0.1-1 s) to collect sufficient photons for each pixel of the spectrometer, thus the Raman map of an extended sample may require long measurement times; (2) the excitation by the pump light also gives rise to photoluminescence (PL), which overlaps to the spontaneous Raman signal and often may overwhelm it.
To increase the acquisition speed, it is possible to adopt the line-scanning7 approach, in which the pump is focused along a line of the FOV (e.g. at coordinate x) and the collected signal (PL or Raman scattering) is dispersed by a spectrometer with a 2D sensor, producing for the given coordinate x a 2D dataset in which one axis corresponds to the spectrum and the other to one spatial coordinate y. Nevertheless, this implementation still suffers from the significant losses introduced by the entrance slit of the spectrometer, whose narrow width provides a better spectral resolution, but also limits the amount of collected light.
A wide-field configuration can be more adequate to reduce measurement time: with this method, a large area of the sample is illuminated and the spectrum is acquired for all pixels of a 2D sensor simultaneously. In this case, spectral information can be collected by using either a set of bandpass filters8 or a tunable spectral filter9 in front of a monochrome imaging camera. Also, this method is based on a frequency-domain approach, and it acquires only a discrete number of spectral bands.
An alternative method to measure a spectrum is the Fourier transform (FT) approach. It is a time-domain technique based on Wiener-Kintchin's theorem10, which states that the power spectral density of a signal is the Fourier transform of its time autocorrelation. In Optics, this is obtained in practice by generating two delayed replicas of the waveform with an interferometer. Their interference is measured by a detector as a function of the relative delay, giving rise to the so-called interferogram11. The FT approach is widely used to measure spectra in the infrared spectral region, where it is called Fourier-transform infrared spectroscopy (FTIR)12. FTIR is routinely used to measure the absorption spectrum of vibrational transitions in the mid-infrared spectral range (~2.5−25 µm wavelength).
The time-domain FT approach offers some benefits compared to conventional dispersive spectrometers: it achieves greater throughput due to the absence of slits (Jacquinot's etendue advantage12); it provides flexible spectral resolution, which only depends on the scan range. Moreover, it is well-suited for wide-field configurations: in fact, when using a 2D detector, the method measures one interferogram for each pixel in parallel, enabling simultaneous recording of continuous spectra across all pixels. On the other hand, an FT imaging system must fulfil two challenging requirements: (1) the relative delay between the two replicas of light must be controlled to a small fraction of the optical cycle; (2) the phase shift between rays that interfere in a single pixel must be limited to guarantee a good coherence and thus high contrast. Some solutions have been proposed based on common-path13,14, or nearly15, interferometers, but they are rather cumbersome or have spectral resolution larger than 100 cm-1.
In this work, a novel wide-field microscope is described, which enables the acquisition of photoluminescence and Raman scattering images. The innovative block of this system is a compact and ultra-stable common-path birefringent interferometer, called Translating-Wedge-based Identical pulses eNcoding System(TWINS)16. It provides delay scans of hundreds of optical cycles with a phase accuracy of a small fraction of the optical cycle itself and high stability. The two light replicas have orthogonal polarizations and travel collinearly through the same common path, therefore, they are not affected by path-length fluctuations, typical of a double-beam interferometer. The scheme, shown in the inset of Figure 1, consists of two blocks of birefringent crystal with orthogonal optical axes: one (B2) with fixed length and the other (B1) split into two wedges that can be moved along the common hypotenuse direction. Incident light is polarized by P1 at 45° with respect to the optical axis of the birefringent crystal, so that half of the energy is along the ordinary polarization, and half along the extraordinary one. While travelling along the birefringent blocks, the two orthogonal components accumulate a relative delay that is finely adjusted by varying the overall thickness of block B1 by changing the insertion of the wedge. The second polarizer P2 projects the field components to a common linear polarization to make them interfere. The delay τ between replicas depends on the wedge displacement x along the common hypotenuse as (Equation 1)
, with c speed of light, α apex angle of the wedges, Δn= ne-no difference in ordinary and extraordinary refractive indexes (birefringence). Since the spectral resolution Δv is inversely proportional to the scan range T=|T2-T1|, with T1, T2 initial and final delay of the interferogram, the high spectral resolution required by Raman spectroscopy calls for high birefringence and long wedge translations. Among the most common birefringent crystals, YVO4 (yttrium orthovanadate) is particularly suited for its high Δn and transparency range in the visible-near infrared. The wedges are designed with an apex angle of 10°, thus, with a scan length of 15.3 mm, the interferometer can impart a delay of 2500 fs at 600 nm, obtaining a spectral resolution of 23 cm-1, over the entire spectral range of interest (the silicon camera sensitivity poses the upper limit). This work will present the strategy for a proper setting of the interferogram sampling (i.e., step and scan range) to achieve the desired spectral resolution and to deal with the broadband background, disentangling the Raman signal from photoluminescence.
As described by Candeo et al.17, this birefringent spectrometer can be coupled with a commercial optical microscope to perform hyperspectral measurements. The requirements are the following: a reconfigurable optical microscope body, a TWINS birefringent interferometer, a low-noise monochrome camera (e.g., CCD or sCMOS), and a narrowband laser source. For the latter, typical excitation wavelengths for Raman and fluorescence spectroscopy are 355 nm, 488 nm, 514 nm, 532 nm, 633 nm, and 785 nm. Despite high-spectral purity lasers, such as the one listed in the Table of Materials, being the best choice, a laser with a broader line width, e.g., ~0.1 nm, can be used as well, given the limited spectral resolution of the instrument. The microscope is upgraded by placing the TWINS in the detection path, between the tube lens and the monochrome detector. This configuration provides a uniform path delay across the field of view and an average contrast of 55%. The excitation path is designed to get a uniform illumination of the sample: in practice, the output tip of a large-core multimode fiber is imaged on the sample plane. By using a mechanical scrambler, the modes are strongly coupled in the core of the fiber, producing a flat-top profile on the sample plane. In addition, a vibrating voice coil averages out the speckle pattern typical of monochromatic radiation. A characterisation of this top-hat illumination is shown in Supplementary Figure 1. The beam is focused on the sample by an infinity-corrected objective, which also collects backscattered Raman and luminescence light toward the detection path. The filters required to reject illumination will be detailed.
To show how to measure broadband and narrowband spectral features with this hyperspectral microscope, both luminescence and Raman scattering are measured for a test sample specifically selected for illustrative purposes. The sample consists of a mixture of three powder pigments with well-known Raman spectra: two polymorphs of titanium white (TiO2), namely rutile and anatase, and cadmium sulphide yellow (CdS). However, the same procedure can be used for a broader range of applications, from materials science to biology.
The scope of this article is to illustrate the steps to perform a hyperspectral measurement, while the assembling and alignment of the TWINS interferometer are detailed by Candeo et al.17. Therefore, an interested experimenter should start from a pre-assembled interferometer based on this technology.
The protocol starts with the sample preparation and how to set up the microscope to excite and collect the Raman scattering and PL photons. Then, the scan parameters (scan length, sampling step) and acquisition parameters (exposure time, hardware binning) are set according to the requirements of the signal under investigation. Finally, the post-processing pipeline will be presented, detailing the procedure for retrieving the spectrum at each pixel (spectral hypercube, Figure 2B) from the acquired temporal dataset (temporal hypercube, Figure 2A), and outlining the algorithms for noise reduction and advanced spectral analysis. The entire workflow is shown in Figure 3.