In Vitro Evaluation of Digitally Designed Surgical Guides for Free-end Mandibular Implants: Accuracy and Stability of Non-cross-arch Designs

142 views

⸱

DOI:

10.3791/70485

⸱

June 30th, 2026

 ,  ,  , 

In This Article

Summary

This in vitro study compared four static surgical guide designs and dynamic navigation for mandibular free-end implant placement. Progressive guide-design modifications reduced most placement deviations, while dynamic navigation demonstrated selected advantages in angular and apical control at the distal #37 site.

Abstract

Digital surgical guides are widely used in implant dentistry, but implant placement in mandibular free-end regions remains challenging because distal tooth support is limited and guide displacement may occur during drilling. This in vitro study evaluated the placement accuracy and stability of four digitally designed static surgical guide configurations and compared static computer-assisted implant surgery with dynamic navigation in a standardized mandibular distal-extension resin model. A partially edentulous mandibular model involving implant sites #36 and #37 was generated from intraoral scan and cone-beam computed tomography data. Four static guide designs were tested: a conventional unilateral cross-arch tooth-supported guide (GT I), two compact non-cross-arch guides with different supporting-tooth configurations (GT II and GT III), and a non-cross-arch guide with distal micro-screw fixation (GT IV). Postoperative scan data were superimposed onto the virtual plan to quantify angular, coronal, apical, depth, and lateral deviations. Progressive modification from GT I to GT IV reduced most angular and linear deviations at both implant sites. GT IV showed the lowest angular deviation and a favorable overall accuracy and stability profile in most comparisons, although GT III showed a slightly lower apical three-dimensional deviation at site #37. In the static-versus-dynamic comparison, dynamic navigation showed site-specific advantages in angular deviation and apical global deviation at #37, whereas static guidance showed comparable or better values for several measurements at #36. The combined-site comparison did not support uniform superiority of either modality. These findings indicate that guide-support configuration and distal stabilization influence placement accuracy and guide stability in a mandibular free-end resin model. Clinical validation is required before routine treatment recommendations can be made.

Introduction

Dental implant restoration has become an established treatment option for partial and complete tooth loss, and the accuracy of implant placement remains a key determinant of prosthetic fit, peri-implant safety, and long-term functional outcomes1. The development of cone-beam computed tomography (CBCT), intraoral scanning, three-dimensional planning, and additive manufacturing has supported the broader use of computer-assisted implant surgery (CAIS) in clinical implantology2,3. By integrating radiographic and surface-scan data, static surgical guides allow the planned implant position to....

Access restricted. Please log in or start a trial to view this content.

Protocol

The mandibular model was generated from anonymized intraoral scan and cone-beam computed tomography data obtained from one adult volunteer. The use of human-derived imaging data was reviewed and approved by the Human Research Ethics Committee, Universiti Sains Malaysia (JEPeM-USM; USM/JEPeM/PP/23110816). Written informed consent was obtained from the volunteer for the use of anonymized intraoral scan and CBCT data for in vitro model fabrication, surgical guide design, and accuracy analysis. All identifying information was removed before data processing, and only anonymized STL and DICOM files were used in the experiment. The chemicals, reagents, and tools use....

Access restricted. Please log in or start a trial to view this content.

Results

This in vitro study evaluated implant placement accuracy and guide stability in a standardized mandibular free-end resin model. The primary accuracy outcomes were angular deviation, coronal three-dimensional deviation, apical three-dimensional deviation, depth deviation, and lateral deviation. Analyses were performed at implant sites #36 and #37. The static guide-design comparison included GT I, GT II, GT III, and GT IV, while the modality comparison included static CAIS and dynamic navigation.

Access restricted. Please log in or start a trial to view this content.

Discussion

This in vitro study evaluated the effects of static guide-design optimization and dynamic navigation on implant-placement accuracy and guide stability in a standardized mandibular free-end resin model. The main finding was that progressive modification from GT I to GT IV generally reduced angular, coronal, depth, lateral, and micromotion values. However, the improvement was not uniform across all outcomes. GT IV showed the most favorable overall profile in most measured parameters, but GT III showed a slightly l.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

The authors have nothing to disclose.

Acknowledgements

This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Calibrated torque wrenchStraumannRatchet / torque control deviceUsed for 8 N·cm GT IV micro-screw fixation and 25 N·cm implant insertion.
Dental model resinFormlabsModel Resin V3Used for printing rigid mandibular resin models.
Dental phantom headNissin Dental ProductsSimple Manikin IIIUsed for standardized fixation of the mandibular resin model and restricted posterior access simulation.
Desktop laboratory scanner3ShapeE4 Lab ScannerUsed for printed-model verification and postoperative STL scanning.
Digital displacement gaugeMitutoyoABS Digimatic Indicator, Series 543, 0.001 mm resolutionUsed for guide micromotion measurement.
Digital force gaugeMark-10Digital force gauge, 0–50 N rangeUsed to verify 5 N, 10 N, and 3 N loads during stability testing.
Dynamic navigation systemClaroNavNavident EVOUsed for dynamic navigation-assisted implant placement.
Fit-checking siliconeGC CorporationFit Checker Advanced BlueUsed to verify surgical guide seating and detect high-contact areas.
Guided implant driverStraumannTissue Level RN guided implant driverUsed for implant insertion to the planned platform level.
Guided surgery drill handlesStraumannGuided drill handles, Ø2.2–4.2 mmUsed with the Ø5 mm guide sleeve during sequential guided drilling.
Guided surgery drill setStraumannStraumann Guided Surgery Instruments / VeloDrill systemUsed for pilot, intermediate, and final drilling.
Implant planning and guide-design softwareStraumann / Dental WingscoDiagnostiXUsed for STL–CBCT registration, virtual implant planning, and guide design.
Isopropyl alcoholSigma-Aldrich / Merck2-Propanol, ≥99.5%Used for washing printed resin models and guides; prepared as 95% isopropyl alcohol for protocol use.
Mechanical model clampGeneric laboratory fixtureStainless-steel adjustable clampUsed for rigid fixation of the resin mandibular model.
Mechanical testing frameMark-10 / InstronManual force test standUsed for standardized loading during micromotion testing.
Metallic guide sleeveStraumannT-sleeve, Ø5 mm, H5 mm, REF 034.053Used for all static guide designs to standardize the drilling path.
Navigation calibration kitClaroNavNavident calibration kitUsed for handpiece and drill-axis calibration.
Navigation tracker setClaroNavNavident model tracker and handpiece trackerUsed for model and handpiece tracking during dynamic navigation.
Open-source statistical environmentR FoundationR version 4.3 or laterUsed for supplementary statistical checks and effect-size calculations.
Resin curing stationFormlabsForm CureUsed for 405 nm post-curing.
Resin washing stationFormlabsForm WashUsed for isopropyl alcohol washing after printing.
Scan bodyStraumannRN scanbody compatible with Tissue Level RN platformUsed for postoperative implant-position reconstruction.
Spreadsheet softwareMicrosoftExcel 365Used for data organization and table preparation.
Statistical softwareIBMIBM SPSS Statistics 29.0Used for normality tests, ANOVA, t-tests, non-parametric tests, and ICC analysis.
Surgical guide resinFormlabsSurgical Guide ResinUsed for printing GT I–GT IV surgical guides.
Surgical motor and handpieceNSK / W&HImplant surgical motor with 20:1 contra-angleUsed for guided drilling at 800 rpm under continuous external saline irrigation.
Three-dimensional inspection softwareZEISSZEISS INSPECT Optical 3DUsed for surface-deviation inspection and planned-versus-placed implant analysis.
Three-dimensional modeling softwareMaterialise3-matic Medical 19.0Used for STL cleanup, mesh repair, and mandibular model preparation.
Three-dimensional printerFormlabsForm 3B+Used for stereolithography printing of mandibular resin models and surgical guides.
Tissue-level implantStraumannStandard Plus Implant, Ø4.1 mm RN, SLActive, 10 mm, REF 033.562GUsed at sites #36 and #37 in all groups.
Titanium micro-screwDentium / Jeil MedicalOrthodontic micro-screw, 1.5 mm × 8.0 mmUsed only for distal fixation of GT IV.
Table note: Equivalent products may be used if they provide the same technical function and measurement precision. Scanner accuracy, printer layer thickness, sleeve dimensions, implant dimensions, guide resin type, navigation registration threshold, torque settings, and displacement-gauge resolution should remain consistent with the protocol.