Summary

Non-invasive 3D-Visualization with Sub-micron Resolution Using Synchrotron-X-ray-tomography

Published: May 27, 2008
doi:

Summary

We used synchrotron X-ray tomography at the European Synchrotron Radiation Facility (ESRF) to non-invasively produce 3D tomographic datasets with a pixel-resolution of 0.7µm. Using volume rendering software, this allows the reconstruction of internal structures in their natural state without the artefacts produced by histological sectioning.

Abstract

Little is known about the internal organization of many micro-arthropods with body sizes below 1 mm. The reasons for that are the small size and the hard cuticle which makes it difficult to use protocols of classical histology. In addition, histological sectioning destroys the sample and can therefore not be used for unique material. Hence, a non-destructive method is desirable which allows to view inside small samples without the need of sectioning.

We used synchrotron X-ray tomography at the European Synchrotron Radiation Facility (ESRF) in Grenoble (France) to non-invasively produce 3D tomographic datasets with a pixel-resolution of 0.7µm. Using volume rendering software, this allows us to reconstruct the internal organization in its natural state without the artefacts produced by histological sectioning. These date can be used for quantitative morphology, landmarks, or for the visualization of animated movies to understand the structure of hidden body parts and to follow complete organ systems or tissues through the samples.

Protocol

Animals used in this study

Specimens of the parthenogenetic oribatid mite Archegozetes longisetosus (Acari, Oribatida) were taken from our laboratory culture. The culture grows on a plaster of Paris/charcoal mix (9:1) in plastic jars, in constant dark at 20-23°C with approximately 90% of air humidity.

Sample preparation

  1. Specimens were taken from the culture, cleaned with a fine brush and placed in a 6:3:1 mixture of 80% ethanol, 35% formaldehyde and 100% acetic acid for 24 hours.
  2. Afterwards, specimens were dehydrated in a graded ethanol series of 70%, 75%, 80%, 85%, 90%, 95%, and 100% with 3 changes at each concentration, and 10 min between the steps.
  3. Finally, samples were placed in fresh 100% ethanol overnight and critical point dried in CO2 (CPD 020, Balzers). Dried specimens were attached to the tip of plastic pins (1.2 cm long; 3.0 mm in diameter).

Synchrotron X-ray tomography

X-ray tomography was performed at beamline ID19 (ESRF, Grenoble, France, experiment SC-2127).

  1. The samples were mounted in the sample-holder and adjusted to a central position in the beam.
  2. Samples were measured with an energy of 20.5 keV. The radiographs were recorded with a cooled CCD (ESRF FReLoN camera) with a 14-bit dynamic range, 2048×2048 pixels and an effective pixel size of 0.7 μm. 1500 projections were recorded over the 180° sample rotation with an exposure time of 0.35 s for each projection. The detector-to-sample distance was 20 mm.

Using a certain distance between sample and detector enables a differential imaging of materials with low X-ray attenuation coefficients (Cloetens, et al. 1996), which would produce insufficient contrast in absorbtion imaging (where the sample is located directly in front of the detector). Most biological matters are phase objects, composed of materials with low absorbtion and/or only small differences in atomic number (Betz, et al. 2007). However, phase enhanced tomography requires a high spatial coherence of a homogeneous X-ray beam. Therefore, synchrotron radiation is better suited than desktop-scanners for these kind of measurements.

Data analysis

  1. The resulting 2D radiographs were transformed to 3D voxel data (8-bit grey-values) with a filtered back-projection algorithm (Cloetens, et al., 1997)
  2. The voxel data were analysed with the software VGStudio Max 1.2.1. (Volume Graphics, Heidelberg, Germany).
  3. Grey-values from the background were removed from the histogram for 3D-visualization.
  4. Pre-defined camera-paths were used to generate rotational animations and animated clipping planes.
  5. A user-defined camera path was generated to follow the digestive system of A. longisetosus.

Discussion

In this presentation, we focused on the 3D-visualization of the internal anatomy of a chelicerate micro-arthropod. The synchrotron X-ray measurements allow a pixel-resolution of down to 0.3µm, depending on the size of the sample. Here, we have shown data with 0.7µm pixel-resolution. Generally, synchrotron X-ray tomography can be useful for analysing small biological materials (or tissues) with low X-ray attenuation. The pixel resolution almost reaches that of conventional light microscopy. The technique can be applied to any kind of material for which the internal organisation is of interest and which must not be destroyed by sectioning. Histological sectioning, however, has the advantage that tissues can be stained differentially, which is not possible with X-ray tomography. But here, different gray-values correspond to tissues with different X-ray attenuation, and the grey-value distribution can be expanded using quantitative phase tomography (holotomography; Cloetens et al.,1999; Heethoff & Cloetens, 2008) . The technique of synchrotron X-ray tomography is extraordinary valuable due to the following reasons:

  1. sample preparation is easy and limited to fixation and drying, no histological sectioning is necessary
  2. the method is non-invasive: the internal organization of the sample can be observed in natural state
  3. the resulting dataset can be analysed in any desired orientation which is different from histological sectioning where the sections are fixed to one orientation.

We suggest considering this technique whenever the internal structures in natural state of small samples or of samples which must not be destroyed (e.g., fossils in amber or type-specimens) are to be analysed. Of course, this technique is not restricted to biological materials, but here it is of the biggest advantage due to the possibility of analysing materials with X-rays which have low X-ray attenuation.

Disclosures

The authors have nothing to disclose.

Acknowledgements

We thank Paavo Bergmann, Michael Laumann, and Sebastian Schmelzle for their help at the ESRF. This work was supported by the European Synchrotron Radiation Facility project SC-2127 through the allocation of beam time.

References

  1. Betz, O., Wegst, U., Weide, D., Heethoff, M., Helfen, L., Lee, W. -. K., Cloetens, P. Imaging applications of synchrotron X-ray phase-contrast microtomography in biological morphology and biomaterial science. I. General aspects of the technique and its advantages in the analysis of millimetre-sized arthropod structure. J. Microscopy. 22, 51-71 (2007).
  2. Cloetens, P., Barrett, R., Baruchel, J., Guigay, J. P., Schlenker, M. Phase objects in synchrotron radiation hard X-ray imaging. J. Phys. D: Appl. Phys. 29, 133-146 (1996).
  3. Cloetens, P., Pateyron-Salome, M., Buffiere, J. Y., Peix, G., Baruchel, J., Peyrin, V., Schlenker, M. Observation in microstructure and damage in materials by phase sensitive radiography and tomography. J. Apll. Phys. 81, 5878-5886 (1997).
  4. Clotens, P., Ludwig, W., Baruchel, J., van Dyck, D., van Landyut, J., Guigay, J. P., Schlenker, M. Holotomography: quantitative phase tomography with micrometer resolution using hard synchrotron radiation X-rays. Appl. Phys. Lett. 75, 2912-2914 (1999).
  5. Heethoff, M., Cloetens, P. A Comparison of aynchrotron X-ray phase contrast tomography and holotomography for non-invasive investigations of the internal anatomy of mites. Soil Organisms. , (2008).

Play Video

Cite This Article
Heethoff, M., Helfen, L., Cloetens, P. Non-invasive 3D-Visualization with Sub-micron Resolution Using Synchrotron-X-ray-tomography. J. Vis. Exp. (15), e737, doi:10.3791/737 (2008).

View Video