Source: Vy M. Dong and Diane Le, Department of Chemistry, University of California, Irvine, CA
This experiment will demonstrate the use of a polarimet…
1. Preparing the Polarimeter
2. Preparation of Analyte Sample
3. Measuring Optical Rotation
4. Calculation of Specific Rotation
![Specific rotation formula [α] = α/(l*c) for optical activity analysis, chemistry equation.](/files/ftp_upload/10348/10348eq3.jpg)
Polarimeters are widely used in organic and analytical chemistry to assess the purity of a chemical product and investigate its properties.
Polarimeters detect the presence of enantiomers: mirror-image variants of a compound that may have wildly divergent biological activities. Distinguishing between enantiomers is critical in many applications, including pharmaceuticals, since one enantiomer is typically responsible for biological effects while the other is usually inert, less active, or, as in the case of the drug thalidomide, harmful.
This video will illustrate the principles of polarimetry, demonstrate setup and operation of a polarimeter, and discuss some applications.
Polarimetry is useful for studying organic compounds containing stereocenters.
Stereocenters are carbon atoms that are bonded to four different atoms or groups. In this example, the carbon atom is bonded to hydrogen, fluorine, chlorine, and bromine, forming bromo-chloro-fluoro-methane.
Compounds containing stereocenters are called "chiral," meaning they exist as mirror-image isomers: non-equivalent physical structures that cannot be rotated or oriented to superimpose on each other. The mirror-image isomers are called "enantiomers," and they have identical physical properties, with one exception related to optics.
In optics, non-laser light sources emit light waves that oscillate in a variety of planes. Such light waves are called "unpolarized." However, certain materials are capable of filtering light waves based on their plane of oscillation, transmitting only those light waves that oscillate in one specific plane while absorbing those oscillating in other planes. The transmitted light has been "plane polarized."
Enantiomers have different effects on plane polarized light. If they are struck by plane polarized light, one enantiomer will rotate the plane of oscillation clockwise, while the other will rotate the plane of oscillation by an equal angle counterclockwise. The former is called the "dextrorotatory" enantiomer, and its name prefixed with a plus sign. The latter is called the "levorotatory" enantiomer, and its name is prefixed with a minus sign. The ratio of rotation angle to concentration is unique for each compound, and is called "specific optical rotation."
A polarimeter detects whether one or both enantiomers are present in a sample. It consists of a light source, a polarizer, a sample cell, a detector, and an analyzer. The light source emits light waves that are unpolarized but monochromatic, meaning they have the same wavelength. The light waves then encounter the polarizer, which transmits only those oscillating in one specific plane, yielding a plane-polarized beam. The plane-polarized light then interacts with the sample in the sample cell.
If the sample contains only one enantiomer of the chiral compound, the polarized light will rotate. The angle is called the "optical rotation," and it depends on the specific optical rotation of the compound, its concentration, and the length of the sample cell. If, on the other hand, both enantiomers are present in equal concentrations, they form a "racemic mixture" that cannot rotate polarized light. Finally, if one enantiomer is present in greater concentration than the other, an "enantiomeric excess" results, and the plane of oscillation will be rotated in proportion to the excess.
After the polarized light passes through the sample, it is detected. The analyzer measures the optical rotation.
Now that you've seen the principles, let's examine a typical operating procedure.
The first step to using the polarimeter is zeroing the instrument.
First, turn on the polarimeter and let it warm up for 10 min.
Set the instrument to optical rotation mode.
The sample cell is typically a tube 1 dm long with a volume of 1.5 mL. Prepare the cell by cleaning with acetone and lab wipes.
Gently place the empty sample cell into the holder and press "zero." This establishes the baseline.
Next, calibrate the polarimeter using a pure sample of the chiral compound under investigation.
In this example, the dextrorotatory enantiomer of carvone is used. Pipette 1.5 mL into the sample cell. Insert the cell into the holder, and press "measure." The optical rotation is displayed. Dividing the measured optical rotation by concentration, or density for pure substances, and cell length yields the specific optical rotation of the compound.
The specific optical rotation of a purified unknown can be found similarly, by dissolving the unknown in an optically inactive solvent and measuring the optical rotation. The specific optical rotation of the compound is then determined by dividing by the concentration. The compound is then identified by comparing its specific optical rotation to literature values.
Now that you know how to perform measurements, we will explore some practical applications.
In the pharmaceutical industry, polarimetry is used for quality control. For instance, it has been used to measure the concentration and enantiomeric purity of ephedrine in commercial cough suppressants. Even in the presence of other ingredients, this technique can be used to determine the ephedrine concentration to within 1%.
In the food and beverage industries, sucrose concentrations and purities are monitored continuously with specially-designed flow polarimeters. Sucrose, one of the most common ingredients in foods, has a specific optical rotation of 66.5 degrees. By dividing the optical rotation of the sucrose stream by the specific optical rotation of sucrose, the concentration can be determined. Fluctuations in the optical rotation would indicate fluctuations in sucrose concentration.
Polarimetry has also been used to study reaction kinetics, including kinetics for enzyme systems such as the penicillin-penicillinase system. In this case, the sample cell contains both enzyme and substrate, and the optical rotation is measured with respect to time. The change in optical rotation is directly proportional to the change in substrate concentration. This not only reveals the reaction kinetics, but also permits simultaneous determination of enzyme and substrate concentrations in future assays.
You've just watched JoVE's introduction to the polarimeter. You should now understand its principles of operation, the steps for setup and measurement, and some of its applications. Thanks for watching!
View the full transcript and gain access to JoVE Science Education videos
Q1: What are stereocenters and why do they matter in polarimetry?
Stereocenters are carbon atoms bonded to four different atoms or groups. Compounds containing stereocenters are chiral, meaning they exist as mirror-image isomers called enantiomers. Polarimetry is useful for studying organic compounds containing stereocenters because enantiomers rotate plane-polarized light in opposite directions, allowing their detection and identification.
Q2: How do enantiomers differ in their effects on polarized light?
When struck by plane-polarized light, one enantiomer rotates the plane of oscillation clockwise, called dextrorotatory and prefixed with a plus sign. The other rotates the plane counterclockwise, called levorotatory and prefixed with a minus sign. The ratio of rotation angle to concentration is unique for each compound and is called specific optical rotation.
Q3: What components make up a polarimeter and how do they work together?
A polarimeter consists of a light source, polarizer, sample cell, detector, and analyzer. The light source emits unpolarized but monochromatic light. The polarizer transmits only light oscillating in one plane, creating plane-polarized light. The sample interacts with this light, and the analyzer measures the resulting optical rotation.
Q4: What is a racemic mixture and how does it affect optical rotation?
A racemic mixture contains both enantiomers in equal concentrations. Because one enantiomer rotates plane-polarized light clockwise while the other rotates it counterclockwise by an equal angle, their effects cancel out. Therefore, a racemic mixture cannot rotate polarized light, making it optically inactive.
Q5: How is a polarimeter zeroed and calibrated before measurement?
First, turn on the polarimeter and let it warm up for 10 minutes. Set it to optical rotation mode. Place an empty sample cell into the holder and press zero to establish the baseline. Then calibrate using a pure sample of the chiral compound by measuring its optical rotation and dividing by concentration and cell length to determine specific optical rotation.
Q6: What is enantiomeric excess and how does it relate to optical rotation?
Enantiomeric excess occurs when one enantiomer is present in greater concentration than the other. In this case, the plane of oscillation will be rotated in proportion to the excess. This allows polarimetry to quantify the purity and composition of chiral compounds in pharmaceutical and food applications.
Q7: How is polarimetry used for quality control in pharmaceuticals?
Polarimetry measures the concentration and enantiomeric purity of active compounds in commercial products. For example, it has been used to determine ephedrine concentration in cough suppressants to within 1%, even in the presence of other ingredients. This technique provides rapid, accurate assessment of drug purity without separating individual components.