Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor

10.6K views

Cited by 2

07:51 min

September 29th, 2020

10.3791/61670-v

September 29th, 2020

10.6K views

The goal of this study was to evaluate the accuracy of capnography sampling lines used in conjunction with a portable bedside capnography monitor. Sampling lines from 7 manufacturers were evaluated for tensile strength, rise time, and ETCO2 accuracy as a function of respiratory rate or supplemental oxygen flow rate.

Explore More Videos

Capnography Sampling Line

Chapters in this video

0:04

Introduction

0:46

Sampling Line Tensile Strength Measurement

1:32

Rise Time and Sampling Line Accuracy Measurement

3:11

End-Tidal Carbon Dioxide (ETCO2) Accuracy Measurement

4:40

ETCO2 Accuracy Measurement in the Presence of Supplemental Oxygen

5:36

Results: Representative ETCO2

7:13

Conclusion