Mice were cared for in accordance with the procedures outlined in the National Institutes of Health and Institutional Animal Care and Use Committee (IACUC) protocol at Meharry Medical College. This institution is accredited by the Association for Assessment and Accreditation of Laboratory Animal Care International and complies with the Public Health Service Policy for the treatment and use of laboratory animals under pathogen-free conditions. The details of the reagents and the equipment used are listed in the Table of Materials.
1. Optimization for cell density
NOTE: Each cell type exhibits unique metabolic characteristics and considerable variation in size. Both factors are critical in determining the appropriate number of cells to load per well to obtain optimal results. Before running an initial experimental assay, prepare and run a test plate with the chosen cells of interest plated at varying densities.
- For adherent cells (e.g., cancer cells), test the seeding densities ranging from 20,000–100,000 cells per well. For highly metabolic and/or large cells (e.g., human cancer cells or fibroblasts), density optimization begins at 10,000 cells or fewer per well.
- For non-adherent cells (e.g., lymphocytes), optimize the seeding density within the range of 100,000–500,000 cells per well. For all cell types, aim for a uniform single-cell monolayer to achieve optimal results (refer to Figure 1 for example plate layout of samples).
- Run the Mito Stress assay (as below). Data from this preliminary assay can be used to determine the optimal cell density for the main experiment. Select the density that shows the greatest difference between assay conditions.
NOTE: This protocol does not apply to the analysis of 3D samples (e.g., cell spheroids, organoids). It can be modified, however, to account for the sample thickness and uneven surface.
2. Preparation for the assay
- On the day before running the assay, gently lift the sensor cartridge out of the utility plate and add 200 µL of RNAase/DNAse-free water to each well of the utility plate, and lower the sensor cartridge into the utility plate until sensors are completely submerged in water.
NOTE: The water level must be high enough to make sure sensors are completely submerged overnight, as the sensor cartridge fluorophores will not work if not properly hydrated. If using an XF Hydrobooster to hydrate the cartridge, the cartridge can be put directly in calibrant solution and the water hydration step excluded.
- Place the hydrated sensor cartridge/utility plate in a 37 °C CO2-less incubator overnight. To prevent evaporation from the plate, also place a container with double-distilled water into the incubator.
- On the day of the assay, remove the sensor cartridge/utility plate from the CO2-less incubator and gently remove the sensor cartridge from the utility plate, placing the cartridge sensor-side up on the bench.
- Discard the water from the utility plate and shake dry of any loose droplets, and add 200 µL of XF Calibrant Solution to each well.
- Gently place the sensor cartridge back in the utility plate so that the sensors are completely submerged in calibrant solution, and place it back in the 37 °C CO2-less incubator.
3. Assay media preparation
- Mito Stress media preparation:
- In a 50 mL tube, add the following to 45 mL of XF RPMI medium: 500 µL of 100x sodium pyruvate (final concentration: 1 mM), 500 µL of 100x L-glutamine (final concentration: 2 mM), and 550 µL of 45% glucose solution (final concentration: 25 mM)
- Adjust the pH to 7.4. Top up with XF RPMI medium to a final volume of 50 mL. Keep the prepared medium in a 37 °C water bath or dry bath prior to using.
- Glycolysis assay media preparation:
- In a 50 mL tube, add 250 µL of 100x L-glutamine to 45 mL of XF RPMI medium.
- Adjust the pH to 7.4, then top up with XF RPMI medium to a final volume of 50 mL. Keep the prepared medium in a 37 °C water or dry bath prior to using.
NOTE: The pH of both Mito Stress and Glycolysis assay media must be readjusted to 7.4 after adding supplements that alter pH, as even small deviations can significantly affect assay results.
4. Washing cells and changing media prior to analysis
NOTE: The washing step of XFe96/XF Pro Cell Culture Microplates will vary depending on whether using adherent or non-adherent cell types. Please refer to step 9, Adherent (cancer cells, primary BMDMs cells) and non-adherent cells (T cells and other lymphocytes), to determine the best washing method for the cell type used.
- Remove 160–180 µL of growth media from each well using a multi-pipettor, working one row/column at a time to prevent cells from drying. Avoid touching the bottom of the wells.
- Using a single-tip pipetman, remove the remaining 20–30 µL from the wells, avoiding touching the cell layer at the bottom of the well.
- Add 180 µL of Glycolysis or Mito Stress assay media (see step 3) to each well, following the layout in Figure 1. To avoid disturbing the cell layer, gently dispense the media along the side wall of each well until all wells are filled. Both assays are in the same plate.
NOTE: The wells that do not contain cells and control wells must also be filled with assay media.
- Place the microplate into a 37 °C CO2-less incubator for 40–45 min prior to starting the assay, and start preparing the drugs.
5. Drug preparation and loading the cartridge
CAUTION: Some drugs (FCCP, Antimycin A/ Rotenone, Oligomycin) are hazardous and should be handled with care. Appropriate measures are needed to protect personnel; these include the use of gloves and other personal protective equipment (PPE) as well as disposing of materials in an institution-approved chemical waste bin or container.
NOTE: Optimal FCCP/oligomycin concentration is cell type–dependent and prior to the assay, empirically titrated when working with particularly sensitive cells. Final concentrations may vary for drugs depending on cell type. Prepare drugs and assay media fresh on the day of use. Do not freeze or reuse. Assay drug preparation can be completed within this 45-min time period for optimal results.
- Using the pre-warmed Mito Stress and Glycolysis Assay media from step 3, dilute the compounds according to Table 1 and/or Table 2 to the concentrations appropriate for each cell type. Prepare each compound in a separate 15 mL tube (if using all six compounds, six tubes will be needed).
- Remove the sensor cartridge/utility plate from the 37 °C CO2-less incubator and remove the lid, carefully placing the XF loading guide provided with the kit, the XF loading guide corresponding to the injector port (Guide A for Port A) securely above the top of the sensor cartridge (refer to Figure 2).
- Using a multichannel pipettor and reagent reservoir for each drug, slowly add the volume needed to the corresponding port (refer to Table 1 and Table 2 for volume). Hold the pipettor strictly upright when loading the drugs (refer to Figure 2).
- Change out or reuse after thoroughly rinsing and drying the loading guide as needed for each port (Port A, B, C, and D) and discard pipette tips after each load.
NOTE: Pipette tips are to fit firmly in the guide port, but not too tightly, as the pipette tips may become stuck in the guide! It is recommended to test out the pipette tips/pipettor used on the guides before loading the compounds to avoid misloading samples.
- After loading all compounds into their respective ports, remove the guide, visually confirm that all ports are filled, and secure the cartridge with its lid. Drugs should correspond to the assay in each part of the same plate (Figure 1).
NOTE: Final well concentrations recommended for Mito Stress are 1 µM oligomycin, 1 µM FCCP, and 0.5 µM Rotenone/Antimycin; for Glycolysis, 10 mM glucose, 1 µM oligomycin, and 50 mM 2-DG.
6. Setting up the Seahorse XFe96 analyzer
- Turn on the analyzer and open the WAVE software (refer to Figure 3).
- Select XF Mito Stress Test.
NOTE: As long as the appropriate assay media and compounds are used, the analyzer can simultaneously perform both mitochondrial and glycolysis tests on the same plate using XF Mito Stress Test.
- Select the number of ‘Groups’ based on the plate setup and label each group. Proceed to the next tab to assign wells for the plate map (refer to Figure 3). Both assays are done in the same plate.
- Highlight the wells for each Group based on how the plate is loaded.
- Adjust the program according to the cell type and experimental purpose. For example, if the setup includes a preactivated group before analysis or an activator is added during the run, extend the incubation time in each cycle to allow activation to occur.
- Select the next tab, ‘Run Assay’ and fill in the information as needed to save the data file.
- Load the pre-hydrated cartridge with drugs (from step 5) into the machine when indicated for initialization. The initialization step will indicate if sensors are properly hydrated and functional.
- Start the assay and load and unload the plates as indicated by the prompts during the program.
NOTE: When loading the cartridge or microplate into the analyzer, make sure all lids are removed! While some analyzer models have a safety feature that can sense lids, earlier models do not. This may damage the machine if the lid is intact during a plate run.
7. Normalization methods: (SRB) and nuclear count (DAPI staining and imaging)
CAUTION: Some reagents in the SRB assay kit are corrosive and will burn if in contact with skin. DAPI may cause an allergic skin reaction. The use of gloves and other PPE is recommended, as is the disposal of materials in an institution-approved chemical waste bin or container.
NOTE: Select the normalization method based on the characteristics and pretreatments of the analyzed cells. Either method is suitable when comparing cells with different proliferation rates, as both methods account for such variations. However, if drug treatments induce senescence, halting proliferation while increasing cell size, DAPI staining is the preferred normalization approach. DAPI may not be suitable for cell types where nuclear staining is difficult to visualize (i.e., adipocytes, erythrocytes) or cells may be multinucleated (i.e., parenchymal hepatocytes, myocytes).
- To normalize cells using the SRB assay (Sulfohodamine B Cell Cytoxicity Assay Kit) after real-time metabolic analysis, remove the plate from the machine and add ¼ of the media volume (eg, 50 µL in 200 µL of culture medium) of the Fixation Solution to each well.
- Incubate at room temperature for 1 h or overnight at 4 °C.
- Remove the solution and gently wash the wells three times with 200 µL of RNase/DNase-free water, taking care not to disturb the cell layer. For short-term storage (if needed), replace the final wash with PBS and store at 4 °C.
- For immediate normalization using the SRB assay, remove the wash solution and allow the plate to air-dry completely at room temperature.
- Add 25 µL of SRB Solution to each well and stain for 15 min at room temperature in the dark or wrap the plate in aluminum foil during the incubation period.
NOTE: SRB solution is light sensitive, and this part of the experiment must be performed in a low-light area.
- Remove the staining solution and add 100 µL of 1x Washing Solution to wash each well 2–3 times.
- Wash as quickly as possible to avoid bleaching and remove wash solutions as much as possible by vacuum suction.
- Add 100 µL of 1x Solubilization Solution to each well and place the plate on a shaker for 10 min at room temperature.
- Measure the Optical Density (O.D.) at 565 nm on a plate reader to determine well density based on color saturation (see Figure 3 for visual reference).
- Export the O.D. values data as an Excel file. This spreadsheet will be pasted into the WAVE software for normalization (step 8).
NOTE: If intense color is observed (> O.D. 3.5) due to cell overload, a suboptimal wavelength (e.g., 490–530 nm) may be used to lower the readings back into the instrument's linear range. Correct background by subtracting the O.D. of the control containing only the culture medium (background control well) from all samples’ readings.
- For DAPI stain, dilute the DAPI stock solution to 300 nM in 1x PBS.
- Wash each well of the cell plate 3 times gently with 200 µL 1x PBS.
NOTE: If there are non-adherent cells, centrifuge the plate at 200 x g for 3 min with no brake to ensure cells are suspended at the bottom of each well after each wash step.
- Fix cells in 2%–4% PFA diluted in 1x PBS for 10–15 min at room temperature.
- Wash each well of the cell plate 3 times gently with 200 µL 1x PBS.
- Add approximately 150 µL of this dilute DAPI staining solution to each, making certain that the cells are completely covered in each well.
- Incubate 1–5 min at room temperature in a low-light area (or cover plate in aluminum foil).
- Remove DAPI stain solution from each well.
- Wash each well of the cell plate 3 times gently with 200 µL 1x PBS.
- Remove any remaining PBS.
- Image the cells at any Excitation/Emission between 358–461 nm or under any DAPI filter setting on a microscope.
- Export the emission values data as an Excel file. This spreadsheet will be pasted into the WAVE software for normalization (step 8).
8. Analysis and normalization of the data
NOTE: For the analysis of data, “Wave” or other software can be used. The protocol provided is applicable to the Wave software.
- Open the results of the analysis in Wave (refer to Figure 3). Normalization is performed using values obtained from the Excel file generated by the microplate reader after SRB staining (step 7).
- Click on “Normalize”.
- In the open window, choose “select all” followed by “paste” to paste the data copied from the Excel spreadsheet generated during the normalization step. Click on “Apply”. This procedure automatically normalized all experimental data.
- To generate a report, click on “Export to”, in the panel that opens on the right, select “XF Cell Mito Stress Test Report Generator.” Save the generated file. During analysis, select only the wells assigned to the Mito Stress Test (refer to Figure 3).
- To generate a report, click on “Export to”, in the panel that opens on the right, select “XF Cell Glycolysis Stress Test” Report Generator.” Save the generated file. During analysis, select only the wells assigned to the Glycolysis Stress Test (refer to Figure 3).
9. Adherent vs. non-adherent cells
NOTE: Adherent and non-adherent cells differ in their growth behavior, surface attachment, and handling requirements. Adherent cells (e.g., cancer cells, BMDMs) attach to the culture surface, so plating and washing do not require special precautions. Non-adherent cells (e.g., T cells, other lymphocytes) remain in suspension and require different centrifugation, resuspension, and plating methods. Understanding these differences is crucial for accurate cell seeding, drug treatment, and assay consistency.
- Ensure that all cells used for the experiment are cultured in a medium and cell culture vessel of choice just prior to the assay being run, but then transfer to the microplate to run the assay.
NOTE: While XF RPMI medium is recommended for the cell types described in this protocol, for other cell types, consult the manufacturer to ensure the appropriate medium is used.
- For adherent cells (cancer cells, BMDM cells), seed the cells to the XFe96/XF Pro Cell Culture Microplate 4–5 h before the assay at an optimized cell density (see step 1) in the media volume 200 µL and place in a 37 °C CO2 incubator.
NOTE: BMDM cells can be plated up to 24 h before the assay, as they do not proliferate and will maintain the same cell density. For fast-growing adherent cells, like cancer cell lines, it is advised to plate on the day of assay if no additional treatment is needed and analyze them as soon as cells are attached to avoid additional cell growth. If this design is not possible, normalization to protein quantity/well or nuclei number/well is a necessary step.
- Ensure that the corner wells for every four corners of the microplate are free of cells but filled with media, as this is measured as the blank control wells.
- After 3–5 h of incubation, observe the microplate to make sure cells are attached to the plate at a uniform level at the bottom of each well.
- If cells are not yet attached, the plate can be placed in the 37 °C CO2 incubator for additional time until cells are thoroughly attached.
- Next, prepare the microplate with adherent cells for washing as mentioned in step 4 to proceed with the assay.
- For non-adherent cells (T-cells and other lymphocytes), coat the microplate with Poly-D-Lysine before adding cells to the plate.
NOTE: While Poly-D-Lysine coating is recommended in this protocol, depending on the cell type, another plate coating (e.g., collagen, fibronectin, gelatin) may be more suitable. If using an alternative coating, please follow the manufacturer’s instructions for use.
- To coat the microplate, prepare 2.5 mL of a 20 µg/mL Poly-D-Lysine solution (50x dilution of 1 mg/1 mL water solution) in 0.1 M sodium bicarbonate, pH 8.0 (bicarbonate provides the optimum pH for adhesive adsorption). This step can be performed prior to performing the assay or on the day of.
- Apply 25 µL of the solution to each well of the plate and incubate on the bench or on a plate rocker at room temperature for 20–30 min.
- Aspirate or pipette off any remaining solution and wash each well twice using 200 µL of RNAase/DNAse-free water.
- Wait until wells are dry (1–2 h) before seeding the cells to the microplate.
- Next, prepare the microplate with non-adherent cells for the washing step by centrifuging the plate at 200 x g for 3 min with no brake to ensure cells are suspended at the bottom of each well.
- Proceed to step 4 and continue the remainder of the assay as mentioned in steps 5–7.