The study was conducted in accordance with the Declaration of Helsinki, and the protocol was approved by the Ethics Committee of Xinhua Hospital (XHEC-C-2023-128-2). All items required in this protocol are listed in the Table of Materials.
CAUTION: Avoid forceful guidewire advancement, deep catheter engagement, and high-pressure balloon inflation during the procedure. The balloon should be inflated gently and only to the minimum pressure necessary for catheter stabilization and partial venous occlusion. Continuous fluoroscopic monitoring of the guidewire, catheter tip, balloon position, and contrast flow is essential throughout the procedure. VOM manipulation should be stopped immediately if resistance, contrast extravasation, vessel staining, suspected dissection, or unstable catheter positioning is encountered. VOM blood sampling should be abandoned in any patient in whom safe catheter positioning cannot be consistently maintained.
1. Preoperative preparation
NOTE: The pre-ablation evaluation and anticoagulation strategy follow the central standard for routine ablation procedures; activated clotting time (ACT) should be monitored during the procedure and maintained within the institutional therapeutic range (typically 300–350 s) during left atrial instrumentation. Anticoagulation with heparin during surgery to make the ACT greater than 300.
- After establishing venous access, place a ten-pole electrode to the CS and get access to the LA through the atrial septum, ensuring that fluoroscopy and contrast conditions are available, and ensure that fluoroscopy and contrast conditions are available.
2. VOM recognition and intubation
- Entering CS: Place the Judkins catheter and Runthrough guidewire into the CS along the ten-pole electrode.
- CS imaging localization VOM: Identify and localize the vein of Marshall (VOM) under fluoroscopic guidance after successful engagement of the coronary sinus (CS).
- Advance the CS sheath or guiding catheter into the proximal CS and perform CS venography using a small amount of contrast medium.
- Acquire venographic images in multiple fluoroscopic projections, including right anterior oblique and left anterior oblique views, to clearly define the CS anatomy and identify venous branches.
- Locate the VOM ostium near the region of the valve of Vieussens, typically close to the junction between the CS and the great cardiac vein.
- Recognize the VOM as a small oblique venous branch that usually courses superiorly and posteriorly toward the left atrium and left pulmonary vein region. Carefully distinguish the VOM from adjacent CS branches by evaluating its origin, direction, caliber, and venographic course.
- If the VOM ostium is not clearly visualized on initial CS venography, rotate or reposition the guiding catheter and repeat contrast injection under a different fluoroscopic projection.
- Use a subselective catheter to explore the proximal CS region when branch anatomy remains unclear.
- Confirm the VOM by selective contrast injection before advancing the guidewire or OTW balloon catheter. Avoid forceful catheter or guidewire manipulation during this step to reduce the risk of venous dissection, perforation, or contrast extravasation.
3. VOM and control site blood collection
- VOM blood collection: Inflate the OTW balloon slowly and stepwise under fluoroscopy in the target segment to stabilize the position (while minimizing mixing with the CS backbone), and slowly collect the VOM blood sample through the OTW system. After blood collection, gently rinse the lumen with heparinized saline to prevent clotting.
- AIV blood collection (ventricular vein control): Withdraw the catheter system to the CS/large cardiac vein direction and enter the AIV under contrast imaging. Use the OTW balloon positioning method to withdraw blood as a non-atrial source control.
- CS main stem and LA cavity blood collection: Obtain CS main stem blood by withdrawing at the CS sheath/catheter position. Collect LA blood after successful transseptal puncture and placement of the transseptal sheath in the left atrium, preferably before ablation energy delivery.
4. Sample processing
- ANP: Collect the blood in pre-cooled EDTA tubes, and protease inhibitors such as aprotinin are added as soon as possible and placed on ice; Centrifuge the blood at 4 °C within 1 h (e.g. 1600 g x 15 min), separate the plasma and store it at -20 °C or lower (long-term recommendation -80 °C).
5. Criteria for minimizing CS-VOM blood mixing
- Advance the OTW balloon catheter into the VOM, inflate the balloon gently with diluted contrast or saline until the catheter position is stable, and minimize reflux from the VOM into the CS trunk.
- Use only the effective and safe balloon inflation volume and pressure required for partial venous occlusion, and confirm adequate balloon positioning by fluoroscopy and contrast injection.
- Verify stable balloon placement, reduced contrast backflow into the CS trunk, and absence of vessel staining or extravasation.
- Aspirate VOM blood slowly through the OTW catheter using gentle negative pressure; avoid rapid or forceful aspiration to prevent venous collapse, hemolysis, and CS contamination.
- Stop aspiration if the sample volume is insufficient or aspiration resistance increases, adjust the catheter or balloon position gently rather than increasing suction force, discard samples with obvious hemolysis, visible clotting, or suspected CS contamination.