The cryo-EM resolution revolution has transformed structural biology, enabling atomic-resolution visualization of proteins, nucleic acids, and macromolecular complexes without the need for crystallization1,2. Three technical advances drove this transformation: the development of DEDs with high frame rates, improvements in electron optics and microscope mechanics, and the optimization of GPU-accelerated image processing software3,4,5,6. Together, these advances have made cryo-EM the method of choice for an expanding community of structural biologists.
The demand for cryo-EM access has prompted significant investment in both large national centers and smaller institutional facilities. National centers—such as NCCAT, S2C2, PNCC, and MCCET in the USA, and eBIC, NeCEN, and SciLifeLab in Europe3,5,6, provide access to the highest-end instruments for approved projects. In-house facilities complement these resources by supporting iterative, project-specific research with faster turnaround times and closer user-staff interaction. However, establishing a functional in-house facility requires far more than purchasing a microscope: the supporting IT infrastructure is equally critical, and its inadequacy is a leading cause of operational failure in new facilities6.
A single high-end cryo-EM session can generate 2–5 TB of raw movie data in 24 h. With multiple microscopes and users, the data burden rapidly reaches petabyte scale. Processing this data requires GPU-accelerated pipelines that must be operational concurrently with data collection7,8. Delays in preprocessing mean delayed feedback on data quality and wasted instrument time. Despite these realities, most published guides on cryo-EM facility management devote minimal attention to the IT components, and none provide a step-by-step operational protocol for setting up and validating a complete data infrastructure.
This protocol addresses that gap. We describe the physical and technical requirements for facility construction, the selection criteria for microscopes and detectors, and—most importantly—the design and implementation of the IT backbone that makes a cryo-EM facility operationally viable6. Particular attention is given to on-the-fly data processing: the automated, concurrent preprocessing of data during collection that enables real-time quality control, guides collection decisions, and dramatically reduces post-session turnaround time9,10,11. By following this protocol, facility developers will be equipped to build a complete, integrated cryo-EM operation that serves a broad scientific community sustainably and efficiently.