Peptide Purification: An RP-HPLC-based Technique to Extract Peptides from Digested Protein Lysates

0 weergaven3:21 min • April 30th, 2023

To isolate and purify peptides, begin with a suspension of protein lysate containing a mixture of digested peptides.

Treat the mixture with trifluoroacetic acid or TFA-based reagent. This anionic or negatively charged reagent forms an ion-pair complex with basic or positively charged residues on peptides, thus masking their charge and improving peptide hydrophobicity.

Now, assemble a reversed-phase high-performance liquid chromatography, or RP-HPLC, column containing a porous silica-based stationary phase, immobilized with non-polar hydrophobic ligands. Add a peptide-compatible equilibration buffer to the column to prepare the stationary phase before sample application.

Now, load the acidified sample into the column. Within the column, the hydrophobic peptides selectively adsorb to the hydrophobic ligands of the stationary phase. Run a wash buffer through the column to elute any unbound or unwanted particles.

Next, pass a less polar organic elution buffer, with high hydrophobicity, through the column. Owing to their selective affinity towards the elution buffer, the bound peptides desorb from the column matrix.

Collect the peptide-containing eluate from the column. Lyophilize the eluate to remove the water content and obtain a dry powder of pure and stable peptides.

To acidify the sample, add approximately 20 microliters of 5% trifluoroacetic acid, or TFA, per milliliter of lysate. Mix the tube well and use a pH strip to measure the pH. If necessary, add extra 5% TFA to adjust the pH to 2.5. Next, connect to the shorter end of a C18 column to a vacuum manifold. After setting the vacuum between 17 and 34 kPa, use 3 milliliters of 100% acetonitrile and a glass pipette to wet the column.

Equilibrate the column by using a glass pipette and applying two 3-milliliter aliquots of 0.1% TFA. Then, load the acidified sample onto the column. Use three 3-milliliter volumes of 0.1% TFA to wash the column. Then, apply 2 milliliters of 40% ACN, 0.1% TFA to elute the sample and collect two 2-milliliter fractions into glass culture tubes. With parafilm, cover the eluate tubes, and use a 20 gauge needle to punch three to five holes into the cover before lyophilizing the samples overnight.

06:34

Een op maat HPLC Zuivering Protocol dat zijn vruchten High-zuiverheid Amyloïd Beta 42 en Amyloid Beta 40 Peptiden, in staat oligomeervorming

Gerelateerde video's

0 Bekeken

13:35

Een handige methode voor de extractie en analyse met hogedruk-vloeistofchromatografie Catecholamine Neurotransmitters en hun metabolieten

Gerelateerde video's

0 Bekeken

09:01

Extractie en kwantificering van oplosbare, radiogelabelde Inositol Polyfosfaten van verschillende plantensoorten met SAX-HPLC

Gerelateerde video's

0 Bekeken

02:54

An Affinity Chromatography Technique for the Purification of a Recombinant Bacterial Protein

Gerelateerde video's

0 Bekeken

04:00

Immunoaffinity Based Extraction of Ubiquitinylated Peptides: A Technique to Selectively Extract Ubiquitin Tagged Peptides from Purified Peptide Fractions

Gerelateerde video's

0 Bekeken

03:54

Nickel Affinity Chromatography-Based Protein Purification: A Technique to Purify Polyhistidine-Tagged Recombinant Proteins from Bacterial Cell Lysate

Gerelateerde video's

0 Bekeken

05:43

A STAGE Tip Procedure for Desalting and Purification of Peptides

Gerelateerde video's

0 Bekeken

09:35

Het oplossen van Affiniteit gezuiverde eiwit complexen door Blue inheemse PAGE en Protein Correlatie Profiling

Gerelateerde video's

0 Bekeken

11:54

Detectie van eiwitubiquitinatiesites door peptideverrijking en massaspectrometrie

Gerelateerde video's

0 Bekeken

07:10

In Vesiculo synthese van peptide membraan precursoren voor autonome Vesicle groei

Gerelateerde video's

0 Bekeken

Laatst bijgewerkt: 22 augustus 2026