Electrochemiluminescence Assay for Quantification of Target Protein Levels in Brain Lysate

0 weergaven • 6:25 min. • July 8th, 2025

For quantitative detection of a specific protein in mouse brain lysate using the electrochemiluminescence, ECL, immunoassay, begin with a multi-well assay plate.

The wells comprise conducting working and counter electrodes, completing the electrical circuit. A dielectric layer separates the electrodes, improving the assay sensitivity.

Add primary mouse monoclonal antibodies against the target protein to the wells. The working electrodes, with greater binding capacity, facilitate the antibodies to immobilize on them.

Incubate with a protein-containing solution, binding to the remaining binding surfaces of the well, reducing background interference.

Add mouse brain nuclear fraction lysate. The target protein in the lysate binds specifically to the monoclonal antibodies.

Add unlabeled polyclonal antibodies, recognizing and binding to epitopes on the protein, which is bound to primary monoclonal antibodies on the electrode. Add specific secondary antibodies labeled with ruthenium complex, binding to the unlabeled antibodies. Add a suitable buffer containing surfactant — providing a suitable ECL generation environment — and tripropylamine, TPA.

Place the multi-well plate in the ECL dete