1. Size exclusion chromatography (SEC)
NOTE: SEC is used to increase the purity of extracellular vesicles (EVs) and remove non-vesicular protein.
- Use a small SEC column (10 mL bed volume) for the isolation of EVs from <100 mL of starting material and a larger column (47 mL bed volume) for the isolation of EVs from >100 mL of starting material.
NOTE: The example below will list volumes for the larger column, with volumes for the smaller column in parentheses. - Bring the SEC column and phosphate-buffered saline (PBS) to room temperature over several hours. Stabilize the SEC column in a vertical position using a....