1. MicroParticle Preparation
- Label Antibodies using Biotin NHS Chemistry
- Determination of the Degree of Biotinylation
- Conjugate Streptavidin Beads with Biotinylated Antibody
- Resuspend the beads in the original vial by vortexing for 1-2 minutes
- Aliquot the desired volume of beads into a microcentrifuge tube
- Wash the beads 3 times with PBST:
- Place the tube containing the beads on the magnet for 2 minutes
- Remove the supernatant by aspiration with the tube on the magnet. Be careful not to disrupt the beads
- Remove the tube from the magnet and add 1 mL PBST
- Vortex for 20 seconds to resuspend the beads
- Centrifuge for 5 seconds to remove the beads from the cap
- Repeat 2.3.1-2.3.5 twice more
- After the third wash place the tube on the magnet and aspirate the PBST
- Add biotinylated antibody to the beads and resuspend
- If necessary bring the volume of the biotinylated antibody up to 300 μL with PBS. If the antibody volume is already greater than 300 μL do not add any PBS
- Incubate the bead/antibody suspension for 30 minutes at RT on the vortex shaker or tube rotator
- Wash 3 times with PBST as in step 3.
- Block beads with biotinylated BSA
- After the third wash add 10 μg/mg biotinylated BSA, adjust the volume to 300 μL with PBS if necessary, and incubate with agitation for 30 minutes
- Wash 3 times with PBST as in step 3
- Resuspend the beads in 100 μL of PBS/0.05% BSA. Sodium azide can be added for long term storage
- Using standard NHS Fluorescein protocols, fluoresceinate the detector antibody
- Determine the Degree of Fluorescination
2. HCP - Development
- Prepare working solutions:
- Fluorescein labeled Antibody (Fl-Ab): Most CHO-HCP assays use Fl-Ab at a final concentration of 400 ng/mL. The Fl-Ab is added as ¼ of the final assay volume, therefore the concentration of the working solution is 1.6 μg/mL. Prepare this solution by diluting FlAb stock with BioScale Bead/FlAb Diluent.
- Beads: Most CHO-HCP assays use beads at a final concentration of 2x105 beads/mL. The bead reagent constitutes ¼ of the final assay volume; therefore the concentration of the working solution is 8x105 beads/mL. Prepare this solution by diluting bead stock with BioScale Bead/FlAb Diluent. Vortex the bead stock well to make sure all the beads are in suspension prior to removing an aliquot.
- Running Buffer: The composition of the running buffer may have to be optimized for the antibodies used to develop the assay, however several CHO-HCP assays have been developed using 10 mM HEPES buffer, 1% Tween, pH 7.5. BioScale has developed an additive for the running buffer (supplied with the development kit) which is added to the buffer to a concentration of 1% prior to use.
- Dilute calibrators and samples with BioScale Sample Diluent.
3. ViBE System Set Up
- Turn the ViBE Workstation on
- Connect the system supply and waste
- Prime the system
- Insert a ViBE Cartridge
- Route the tubes
- Prepare assay template by filling out the sheet labeled "Sample Setup". BioScale provides the capability to have one assay template for each ViBE Workstation assay. This allows the user to edit and save specific notes in the "Notes" sheet of the template for different assays keeping a complete record of the experiment.
- Initial assay template set up: Begin by filling out the table (top left corner) of the concentration range of the standards in the assay in the "Sample Setup" sheet.
- Next, enter additional information in the table to the right of the standards (Simple Column Read Setup, One Reagent Add and Two Reagent Add setups) to instruct the ViBE Workstation how long samples need to incubate, how much volume of reagent is to be transferred, how long the sample accumulation needs to be, and whether sensor conditioning and decontamination of probes and cartridge are necessary, etc
- Fill out the plate layout.
- Load sample/reagents into plates. The ViBE will provide the required volumes for the reagent plate. Sample volume can vary, but in half of a total assay volume of 120 μL is a good starting place.
- Apply plate covers. Load plates onto workstation deck.
- Run the assay start up wizard, go through wizard and start assay.
4. Representative Results

Figure 1: AMMP Sandwich Immunoassay
On a ViBE Workstation, an AMMP assay measures antigen concentration. Signal is measured when the bead-antigen-hapten antibody binds to the sensor through the hapten/anti-hapten interaction. Regeneration brings the sensor back to the beginning state by removing the immune complexes.

Figure 2: CHO-HCP AMMP Assay calibration curve
Figure 2 above is an example of a CHO HCP AMMP Assay calibration curve. Analysis can be performed in samples from across the purification stream including the reactor with high precision and reproducibility.

Figure 3: Protein A AMMP Assay calibration curve
Figure 3 is a calibration curve for a Residual Protein A AMMP Assay, performed with a VIBE Residual Protein A Assay Kit. The same kit can be used for a faster assay with a 30 minute incubation period, or for more sensitive results and wider dynamic range with a longer incubation period without loss of precision or accuracy, both of which are typically better than equivalent ELISA's in users' hands

Figure 4: Product Titer AMMP Assay
The ViBE Workstation can also be used for a Product Titer assay, results shown in figure 4. Performed differently than other AMMP Assays, this assay does not require sample treatment other than simple dilution, or beads, and quantitation has been verified in "no spin" samples containing over 10 million cells over a wide range of viabilities.

Figure 5: AMMP Assay standard curve for GST-Gadd34 biomarker assay
Shown in Figure 5 is a standard curve of signal obtained by the ViBE BioAnalyzer (Panel A) or AlphaScreen Technology using recombinant GST-Gadd34. Results on the ViBE platform were approximately a log more sensitive using the same antibodies. AMMP leverages the measurement of fewer larger particles yet still benefits from solution phase capture and detection complex formation.

Figure 6: AMMP Assay standard curve for GST-Gadd34 in tumor sample
In Figure 6, the induction of Gadd34 from CWR22 xenografts was tested in the presence or absence of a boronate proteasome inhibitor as analyzed by either the ViBE BioAnalyzer (Panel A) or AlphaScreen (Panel B). The ViBE platform was able to produce sensitive, reproducible results in the in vivo tumor xenografts whereas the AlphaScreen signal was quenched with increased μg of protein.
This example shows the power of AMMP technology in highly complex samples. Due to the sensitive detection chip and high efficiency of the magnet near the chips, low numbers of magnetic microparticles can efficiently and reproducibly scavenge for low abundance target amidst the complexity of lysed cellular (including lysis reagents) and blood debris.