$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
1. Apparatus
For all tests, a soft padded surface is placed at the base of the apparatus to cushion any mice that fall off. NB This is not shown in the figures, in order not to obscure any part of the apparatus. During the testing of hundreds of mice, both normal and mutant, we have never observed any injury caused by falling from such apparatus, which is typically 30-50 cm high. Also, be sure to clean and sterilize the equipment between each mouse tested.
1.1 Rotarod
There are several commercial versions of this apparatus on the market, but some have disadvantages, such as failing to accelerate at an adequate speed to detect motor incoordination (rather than endurance). Another common fault is that the rod is not high enough above the base, resulting in mice tending to drop or jump off.
The rotarod shown in Figure 1 was made to the design of the author by the engineering sections of the department of Pharmacology, University of Oxford. The rod is 3 cm in diameter, supported 30 cm above the base of the apparatus. The surface is knurled in a series of parallel ridges along the longitudinal axis, enabling the mice to grip it effectively (Figure 2). NB The depth of the ridges is a vital detail; if the mouse cannot get a good grip the test will be very much harder; on the other hand, if the grip is too good the mouse will "cartwheel" around the rod by passively holding on to it. For "amateur rotarod makers" it may be useful to know that knurled wooden dowelling of an appropriate size for mice is available commercially; it would of course need to be treated with a good waterproof varnish before use. The start speed is adjusted to 4 rpm, the acceleration rate to 20 rpm/min. Maximum speed is 40 rpm.
Two flanges prevent the mouse from leaving the rod. They are 30 cm in diameter (this could probably be reduced to 20 cm). Their separation is set at 6 cm (maximum) but may need to be adjusted smaller for sub-adult mice if they tend to turn around on the rod.
1.2 Triple horizontal bars (Figure 3)
The bars are made of brass, 38 cm long, held 49 cm above the bench surface by a wooden support column at each end. The columns are secured to a heavy wooden base. Three bar diameters are available: 2, 4 and 6 mm. The 2 mm bar is the standard one we use, but many mice reach maximum scores on this. Therefore larger diameter bars have been included in an attempt to refine the test, as mice cannot grip these so well.
1.3 Static rods
Five wooden dowels or rods of varying thickness (35 (rod 1), 28, 22, 15 and 9 (rod 5) mm diameter) each 60 cm long are fixed by a G-clamp to a laboratory shelf such that the rods horizontally protrude into space (Figure 4). The end of the rod near the bench has a mark 10 cm from the end, to denote the finishing line. The height of the rods above the floor is 60 cm.
If a shorter and less detailed test is required, use only the largest, smallest and middle rods. Van Dellen et al.14 used only a single rod to assess the progress of Huntington's disease in mice.
1.4 Parallel bars
Two parallel steel bars 1 m in length and 4 mm in diameter are fixed 30 mm (on their centers) apart by wooden supporting columns at their ends (Figure 5). The bars are 60 cm above the floor.
2. Procedure
For all tests, bring mice to the experimental room 5-20 min before testing, to ensure they are fully awake. As a general rule, to allow recovery of muscular strength and a return to normal levels of arousal, rest the mice by a return to the home cage after each motor test.
2.1 Rotarod
Set the rotarod with a start speed of 4 rpm, acceleration rate 20 rpm/min. Holding the mouse by the tail, and place it on the rotating rod, facing away from the direction of rotation so it has to walk forward to stay upright. This is easiest to do if the mouse is brought up towards the rod at an angle of 45° below horizontal; attempting to lower the mouse from above will result in it spreading its hind legs and grasping the edges of the flanges. Quickly release the mouse when it is almost touching the rod, just forward of and above the rods top dead center, enabling it to easily grip the rod. A (probably easier and more reliable) technique is to use a thin (10-15 mm) dowel. Holding the mouse in one hand, lower it on to the dowel, parallel to the long axis. The dowel is angled with the mouse's head downwards about 30° to give it more time to grip the rotarod once it is released. Lower the mouse and dowel between the flanges of the rotarod, and once over the rotating rod the pull the dowel downwards away from the mouse, which will grasp the rod and walk forwards.
At 10 sec after placing the mouse on the rod, start acceleration (as long as the mouse is facing forward: if not, wait till it does face forward before starting acceleration), and note the speed at which the mouse falls off. If it falls off before 10 sec, note the time of fall and try again, up to three times in total, recording the speed at the first fall after the 10 sec point. However, falls before 5 sec which are due to poor placing by the experimenter should not be recorded. The mean speed at fall is the datum; rather than using the maximum speed, this corrects for the extra practice the mouse receives during the failed runs, which it is assumed would assist performance. Thus a mouse that falls before 10 sec once then stays on till 12 rpm the second time scores (4+12)/2 = 8 rpm. A mouse falling twice then staying on for 10 rpm scores (4+4+10)/3=6 rpm. If a mouse fails to grip in 10 sec three times assign it a score of 4 rpm.
One problem is that mice sometimes stop walking forward and grip the rod firmly instead, so that they passively somersault round, although most eventually fall off. It might therefore be useful to also note the speed when the first such inversion occurs. A possible solution might be to use a rod with a larger diameter or less pronounced ridges.
2.2 Horizontal bars
As mice find it easiest to grasp the narrow 2 mm bar, it is advisable to test them first on this one. Holding the mouse by the tail, place it on the bench in front of the apparatus, slide it quickly backwards about 20 cm (this aligns it perpendicular to the bar), rapidly raise it and let it grasp the horizontal bar at the central point with its forepaws only, and release the tail, simultaneously starting the stopclock. This can be difficult to achieve efficiently; some mice grip better if the tail is released suddenly; if they feel they are still supported they may fail to grip. The criterion point is either a fall from the bar before the mouse reaches one of the end columns of the bar, or the time till one forepaw touches a column. Maximum test time (cut-off time) is 30 sec.
If the mouse fails to grasp the bar properly first time and this appears to be attributable to the experimenter's technique, rather than the mouse, try again (after a brief rest while another one or two mice are tested) and do not record this fall. If the mouse falls before 5 sec and this is apparently not due to poor placing by the experimenter, repeat up to three times in an attempt to get a >5 sec score. If >5 sec on the second attempt, do not do a third trial. Take the best score as the datum. A mouse that repeatedly fails to support itself for >5 sec scores only 1.
If the triple-bar version is being used, if the mouse scores 5 on this first 2 mm bar it can then be tested on thicker bars, after a brief rest period while another one or two mice are tested. The scoring system for the 4 and 6 mm bars is the same as for the 2 mm bar, and the final score is the cumulated total. Thus a mouse that scores 5 on the 2 mm bar but falls from the 4 mm bar after 13 sec scores 5+3 = 8.
2.2.1
Scoring the horizontal bars: the first two intervals are less than the last two as once the mice have initially mastered the task they are less likely to fall:
Falling between 1-5 sec = 1
Falling between 6-10 sec = 2
Falling between 11-20 sec = 3
Falling between 21-30 sec = 4
Falling after 30 sec = 5
Placing one forepaw on a bar support without falling = 5
NB as the optional use of 4 and 6 mm bars is a new addition to our protocol we do not yet have extensive experience with them.
2.3 Static rods
Place the mouse at the far end of the widest rod (nose tip one head's length from the end is ideal). Take two measures: orientation time (time taken to orientate 180° from the starting position towards the shelf) and transit time (the time taken to travel to the shelf end (nose beyond the 10 cm mark from the shelf end of the rod).
Orientation is dependent on the mouse staying upright; if it turns upside down and clings below the rod, arbitrarily assign it (for statistical purposes) the maximum orientation score of 120 sec. Do not test it on smaller rods; also the transit time becomes the maximum value as a mouse that transits upright is obviously better coordinated than one which transits while hanging on upside down.
If, after orienting, the mouse falls or it reaches the maximum test time (arbitrarily set at 120 sec) do not test it on smaller rods. For statistical purposes, assign mice falling are assigned 120 for that particular rod and for subsequent ones, as it is assumed they would again fall from a narrower rod. If the mouse turns upside down, also note this event; for successful transit as well as orientation it must stay upright on the rod. Otherwise assign it the maximum score. Remove the mouse after it reaches the end of the rod or falls.
After testing on one rod return the mouse to the home cage to rest while you test another mouse. Then place it on the next smaller size rod and test it in the same way again. Stop testing if the mouse falls off a rod after being on it for more than 5 sec. If it fell off in less than 5 sec replace it and allow another attempt (as falling within 5 sec could be due to faulty placing by the experimenter), for a maximum of three trials, and use the best result.
2.4 Parallel bars
Place the mouse in the center of the two bars with its longitudinal axis perpendicular to that of the bars; both front paws should be on one bar, both hind paws on the other bar. Scoring is similar to that for the static rods. Take two measures: the time taken until the mouse orients 90° to the start position, and the time until it subsequently reaches one of the end supports. If the mouse turns upside down, also note this event. As for the static rods test, orientation and transit must be achieved without it turning upside down, otherwise the maximum time is allotted. It is rare for reasonably healthy mice to fall in less than 5 sec; if this does happen it is probably due to poor placing. Therefore discard the result and retest the mouse.