$$\rightleftharpoonup{xx}$$
$$\longleftharp{xx}$$,
$$\longrightharp{xx}$$,
1. Sample amplification using RT-PCR
- Extract viral RNA from clinical material or a viral isolate using the Qiagen MinElute Virus Spin Kit in conjunction with the QIAcube automated nucleic acid extraction platform. Extractions are performed on 200 μl specimen with a final elution volume of 60 μl. Store extracts at -70°C or lower for later use.
- In a template free-area, prepare the RT-PCR master mix on ice according to the manufacturer's protocol. To incorporate biotin during RT-PCR, use a biotinylated dNTP mixture in place of the manufacturer supplied dNTP mix. The cost of using a biotinylated dNTP mixture is less than $1 USD/assay. Alternative methods of biotin incorporation such as the use of a biotinylated primer can also be used. Add primer mixes at final concentrations of 1.0 μM for Flu A, 1.0 μM for Flu B, and 0.14 μM for the internal control. The Flu A primer set amplifies the matrix gene segment (1032 nt product) and the Flu B primer set amplifies a non-structural gene segment (811 nt product). Each FluChip primer set contains one primer with a 5' phosphoryl group to facilitate the generation of single stranded DNA via enzymatic digestion with lambda exonuclease following PCR. The resulting single stranded product requires much shorter hybridization times than the double stranded equivalent and significantly shortens the overall assay time. A pre-prepared mixture of these primers as well as the internal control template RNA are available from InDevR to interested researchers on a limited basis.
- Briefly vortex and distribute 18 μl of the master mix into thin-walled PCR tubes.
- Transfer tubes on ice to a suitable work area for template addition, and add 2 μl template to each reaction tube.
- Transfer PCR tubes to a thermal cycler, and run the following thermal profile: reverse transcription at 50°C for 30 min, enzyme inactivation/activation at 95°C for 15 min, 40 PCR cycles of 95°C for 30 s, 55°C for 30s, and 72°C for 1 min, and a final extension at 72°C for 10 min.
2. Hybridization of RT-PCR products to low density microarrays
- In order to generate single stranded DNA for FluChip hybridization, prepare the enzymatic digestion mixture by combining 1.0 μl lambda exonuclease enzyme, 2.2 μl of the accompanying reaction buffer, and 0.8 μl of nuclease-free water. These amounts are for a single sample, but can be simply scaled for the total number of samples to be digested. Remove samples from the thermal cycler and add 4 μl of the prepared mixture to each reaction to digest the phosphorylated strand of the PCR product. Return samples to the thermal cycler and program the thermal cycler to 37 degrees Celsius for 15 minutes followed by 95 degrees Celsius for 10 minutes to complete enzymatic digestion and heat fragmentation steps.
- The custom influenza microarrays used are printed on aldehyde functionalized glass slides by Applied Microarrays Inc. (Tempe, AZ). 5'-amino terminated capture sequences are combined with an optimized spotting buffer and printed at a final concentration of 20 μM (except for the positive control sequence which additionally has a 3'-biotin modification and is spotted at a final concentration of 500 nM). A non-contact spotting method is used, with an optimal spot diameter of 300 μm and center to center pitch of 700 μm.
- Remove microarrays from the storage box and apply single-use hybridization wells around the microarrays by removing the protective sheet and pressing firmly around well perimeter.
- Place slides in a wash bin containing 110 ml of purified water for a 5 min pre-hybridization wash using an orbital shaker at 60-90 rpm. Dry the array by gently touching a tissue wiper to the edge of the well and allowing water to be wicked away.
- Combine 22 μl of 2x Hybridization Buffer with each of the fragmented ssDNA products, and pipet 40 μl of the hybridization solutions into the microarray wells.
- Allow slides to hybridize in closed humidity chamber for 60 minutes.
- Remove the slides from the humidity chamber, briefly rinse arrays with Wash Buffer D in a rinse bottle before placing into the slide rack. Typically the rinsing volume of Wash Buffer D is 2 ml per array. Place slide rack containing slides in wash bin containing 110 ml Wash Buffer A. Place bin on orbital shaker at 60-90 rpm for 1 minute.
- Remove slide rack from Wash Buffer A, briefly rinse with Wash Buffer D, transfer slide rack to a bin containing Wash Buffer B, and shake at 60-90 rpm for 5 minutes.
- Gently dry the array on each slide, and place dried slides in humidity chamber in preparation for ampliPHOX Colorimetric Detection steps.
3. ampliPHOX: labeling hybridized product and calibration chips with ampliTAG
- Combine 10 μl of ampliTAG, 20 μl of 2x ampliTAG buffer, and 10 μl of purified water for each array to be processed. Reagent volumes can be simply scaled for the total number of samples. Make sure to prepare enough labeling mixture to account for all sample arrays and calibration arrays needed. The number of calibration chips needed depends on whether you are performing a full calibration (for a new instrument or new lot of reagents), or just an instrument check. For a full calibration, 3 calibration chips should be labeled, and for a reagent check, just a single calibration chip should be labeled. Please note that before the calibration arrays are labeled, they should go through the pre-hybridization washing step that was previously described for the influenza arrays.
- Transfer 40 μl of label mixture to each array and allow labeling reaction to proceed in a closed humidity chamber for 5 minutes.
- Immediately rinse arrays with Wash Buffer D in a rinse bottle before placing slides in a slide rack. Transfer rack into bin containing 110 ml of Wash Buffer C, and shake at 60-90 rpm for 5 minutes.
- Using a second wash bin filled with purified water, perform three consecutive brief water dips to remove salt residue. Dry the arrays by gently touching a tissue wiper to the edge of the wells.
- The microarrays are now properly labeled with ampliTAG, and the remainder of the ampliPHOX detection procedure can be performed. Photoactivation and imaging of labeled arrays should be completed within 24 hours, and additional arrays can be stored in a dark slide box until use.
4. ampliPHOX: calibration, signal amplification, and imaging
- Turn ampliPHOX reader on and ensure ampliVIEW software is ready for photoactivation.
- Determine the optimal photoactivation time using calibration chips. In addition to the brief introduction that follows, this procedure is also outlined in detail in the ampliPHOX Operation Manual. The calibration chips contain a series of dilutions of a biotinylated control sequence that are utilized to optimize the sensitivity of the assay, with the goal of maximizing the number of rows of the calibration chip that produce a positive signal as determined by the software.
- Remove ampliPHY from 4°C, allow to warm to room temperature, and vortex briefly to mix. Pipet 3 μl of ampliPHY enhancer into the ampliPHY vial, and vortex thoroughly for 10 seconds.
- Evenly transfer 40 μl of ampliPHY solution into the microarray well containing the ampliTAG-labeled array, ensuring that no bubbles are present. Close the ampliPHY vial between each application. Insert the microarray slide into the photoactivation bay of the ampliPHOX Reader.
- For the first calibration array, use the default photoactivation time in the 'Time' box on the lefthand panel of the software, and click the green 'Start' button to start the photoactivation. Once completed, remove the array and rinse with purified water to remove excess ampliPHY. Clear polymer formation on some of the spots should now be visible.
- Allow the polymer spots to dry for 2 minutes, then distribute two drops of ampliRED onto the array and allow staining to proceed for 2 minutes.
- Next, quickly rinse microarray with purified water and dry with a tissue wipe.
- Insert microarray into the imaging bay of the ampliPHOX Reader, and click the 'Capture New Image' button in the Imaging tab. Once imaged, adjust the crop and save the image.
- In the Analysis tab, select the Calibration Chip mask and the 'Auto Placement' button to start the analysis. The software will automatically produce a 'Summary Record' showing the quantified results. Based on these results, the photoactivation time is adjusted by 10 seconds for the second calibration array, and the procedure is repeated.
- Once the optimal photoactivation time has been determined, the sample arrays can be processed using the same signal amplification and imaging protocol used for the calibration arrays.
- Once the image is captured for the sample arrays, a mask for your particular array layout such as the influenza array layout described here can be created.The ampliVIEW software is capable of performing automated image analysis and generating influenza subtyping results for each sample.
5. Representative Results:

Figure 1. Schematic illustration of the ampliPHOX colorimetric detection method. (A) Biotinylated target DNA is hybridized to each spot in the array, and (B) labeled with ampliTAG. (C) ampliPHY solution is then added, and (D) exposed to light to form visible polymer spots. (E) The polymer spots formed are subsequently stained with a non-toxic dye to improve contrast.

Figure 2. (A) Influenza low density microarray layout. Sequences 1-7 and 10-13 target influenza A, and sequences 8, 9 target influenza B. (B) ampliPHOX and (C) fluorescence images of a 2009 novel H1N1 ('swine flu') specimen showing the same detection pattern by both methods.

Figure 3. From left to right, representative ampliPHOX images for influenza A H3N2, human-origin H1N1, 2009 novel H1N1 (swine-origin), and a negative specimen. All 3 subtypes show visually distinct patterns on the array. Notice in the negative that only the MS2 internal control is seen, indicating the RT-PCR amplification was not inhibited.