All procedures involving animal samples have been reviewed and approved by the appropriate animal ethical review committee.
1. Materials and equipment
NOTE: See the Table of Materials for all reagents and materials required for this protocol. In addition, manuals and websites associated with the respective vendors can also be consulted as needed.
- Flow chamber
- Fasten a superfusion chamber with a glass coverslip onto a platform composed of anodized aluminum. Secure the platform with the chamber onto an aluminum microscope stage.
- Set a micromanipulator holding a pinning pipette at each end of the platform on the aluminum microscope stage.
NOTE: If necessary, use a transferable stage apparatus with a flow chamber unit to move a secured, isolated endothelial tube from one microscope apparatus to another for experimentation.
- Microscopes
- Set up the experimental apparatus by arranging an inverted microscope (objectives: 4x, 10x, 20x, 40x, and 60x) and a manual aluminum stage on a vibration isolation table.
- Intracellular Vm recording equipment
- Connect the electrometer to a compatible headstage. Use accessories, such as a function generator and stimulator, for protocols requiring current injection.
- Connect amplifier outputs to a data digitizer system, oscilloscope, and audible baseline monitors. Secure the reference bath electrode (Ag/AgCl pellet) near the flow chamber exit.
- Assemble a photometric system with integrated components of a fluorescence system interface, high-intensity arc lamp and power supply, hyperswitch, photomultiplier tube (or PMT), and camera to measure [Ca2+]i in endothelial cells.
- Assemble a temperature controller equipped with an inline heater to raise and maintain a physiological temperature (37 °C) throughout the experiment.
- Assemble a multichannel platform connected to a valve controller with an inline flow control valve to control the delivery of solutions to endothelial tubes secured in the chamber.
- Micropipettes and sharp electrodes
NOTE: The experimenter will need an electronic glass puller and a microforge to prepare pinning pipettes.- To secure the endothelial tube in the superfusion chamber, prepare heat-polished pinning pipettes with a blunted, spherical end (outer diameter: 50-70 µm) prepared from thin-wall borosilicate glass capillaries.
- To record the Vm of an endothelial cell, prepare sharp electrodes with a tip resistance of ~150 ± 30 MΩ from glass capillaries using the glass puller only.
2. Solutions and drugs
- Physiological salt solution (PSS)
- Prepare a minimum of 1 L of PSS using 140 mM NaCl, 5 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 10mM N-2-hydroxyethylpiperazine-N'-2-ethanesulfonic acid (HEPES), and 10 mM glucose.
- Prepare necessary solutions lacking CaCl2 (zero Ca2+ PSS) for dissection of cerebral arterioles and isolation of endothelial tubes.
NOTE: Prepare all solutions in ultrapure deionized H2O, followed by filtration (0.22 µm). Ensure that the final product contains a pH 7.4 with osmolality between 290 and 300 mOsm.
- Fura-2 and pharmacological agents
- Prepare Fura-2 AM stock in dimethyl sulfoxide (DMSO; 1 mM). Prepare 500 µL of working concentration (10 µM) by adding 5 µL of the stock to 495 µL of PSS for loading.
- Prepare at least 50 mL of working concentrations of pharmacological agents in PSS or DMSO as appropriate.
- Conducting solution
- Prepare 2 M KCl by dissolving KCl in deionized H2O (7.455 g of KCl in 50 mL of H2O). Pass the solution through a syringe with a 0.22 µm filter prior to backfilling the sharp electrodes.
3. Utilization of arteriolar endothelial tubes for the examination of cellular physiology
NOTE: Isolated and secured arteriolar endothelial tubes can be used for intracellular recordings of [Ca2+]i dynamics and Vm using photometry and sharp electrode electrophysiology, respectively, as previously illustrated (Figure 1). [Ca2+]i and Vm can be measured as separate or combined experimental variables (Figure 1). However, arteriolar endothelial tubes are more delicate than arterial endothelium, and experimentation time should not exceed 1 h.
- Measurement of [Ca2+]i
- Turn on the equipment and software for [Ca2+]i recordings while maintaining continuous superfusion at a flow rate of 5-7 mL/min.
- Load the endothelial tube with the Ca2+ dye Fura-2 AM for 30 min at room temperature. Wash the cells with superfusion solution for another 20-30 min while gradually raising the bath temperature to 37 °C. Maintain the temperature at 37 °C throughout the experiment.
- Manually adjust the imaging window using photometry software to focus on ~20 endothelial cells (Figure 1A). In the absence of light, turn the PMT on the fluorescence interface and begin acquisition of [Ca2+]i by exciting Fura-2 alternately (≥10 Hz) at 340 nm and 380 nm while collecting fluorescence emission at 510 nm. Once a stable baseline recording of [Ca2+]i is established, apply pharmacological agents (e.g., purinergic receptor agonists) per the experimental objective (Figure 1B).
- Measurement of Vm
- Turn on the equipment and software for Vm recordings and set the data acquisition rate (≥10 Hz) while maintaining continuous superfusion at a flow rate of 5-7 mL/min. Gradually raise the bath temperature to 37 °C and maintain it until the end of the experiment.
- Pull a sharp electrode using a borosilicate glass capillary, backfill with 2 M KCl, and secure it over a silver wire coated with chloride in the pipette holder attached to an electrometer that, in turn, is held by a micromanipulator.
- While viewing through the 4x objective, use a micromanipulator to carefully position the sharp electrode tip just over a cell of the arteriolar endothelial tube into the flowing PSS in the chamber.
- Gradually increase magnification to 400x and reposition the electrode tip as needed.
- Using the micromanipulator, gently insert the tip of a sharp electrode into one of the cells of the endothelial tube and start recording Vm using an electrometer (Figure 1A).
Once the endothelial resting Vm is stable (−30 to −40 mV), apply the desired pharmacological agents per experimental objective (Figure 1C).