1. Isolating carotid arteries
- Euthanase10 week old Sprague Dawley rats via CO2/O2 asphyxiation.
- Excise left and right common carotid arteries with aorta and heart ensuring minimal stretching of the vessels.
- In ice cold Krebs buffer separate the carotid arteries from the aorta and heart and perform close dissection.
- Keep isolated vessels in Krebs on ice before mounting.
Approx. time = 45 mins
2. Priming the vessel chamber
- At the proximal and distal connectors of the vessel chamber flush Krebs buffer maintained at physiological pH by infusing carbogen gas (95% O2; 5% CO2) through the buffer at 37°C.
- Ensure tubing and cannulas are flushed completely and are aligned.
- Flush Kreb buffer through the P1 (proximal) and P2 (distal) transducers. Close taps to ensure no air bubbles in the transducers.
- Connect the transducers to the corresponding connectors and again flush more Krebs buffer ensuring no air bubbles. Close off taps to the chamber.
Approx. time = 15 mins
3. Pressurizing the vessel chamber
- Turn on pressure equipment: pressure servo, pressure monitor, peristaltic pump (Figure 1).
- With the peristaltic pump on pressure and the pressure servo on automatic run Krebs buffer through the tubing at 20 mmHg (dial at 1), ensuring no air bubbles.
- Connect tubing to the closed P2 transducer. Pressure will become stable.
- Open P2 tap to the vessel chamber to flush out any bubbles then close off.
- Fill the bath with Krebs buffer (5 - 7 mL).
Approx. time = 10 mins
4. Mounting the vessel
- Under a dissecting microscope place black polyester ties onto each cannula (Figure 2).
- Move cannulas and cannula holders apart and mount ¼ vessel (aortic arch end) onto the P1 cannula. Ensure not to tear the vessel.
- Using a syringe filled with Krebs buffer gently flush excess blood out of the vessel via P1 transducer. Close off P1 to chamber.
- Secure vessel to the P1 cannula with the polyester tie.
- Move cannulas/cannula holders closer for mounting onto the P2 cannula.
- Mount distal end of the vessel (~¼ of the length) onto distal cannula. Secure with polyester tie.
Approx. time = 20 mins
5. Pressurizing the vessel
- Adjust the cannula holders to ensure no bend or stretch in the vessel.
- With P1 and P2 closed to the chamber switch the pressure servo from automatic to manual. Check on the pressure servo there is no continual decrease in pressure within 10 seconds. If there is, a leak is occurring and the connections and transducers need to be secured in place.
- Adjust the pressure servo back to automatic then open P2 to the chamber. Under the microscope observe the vessel dilate when the tap is opened to the chamber.
- Switch the pressure servo to manual. Again check for continual decrease in pressure within 10 seconds. If there is a leak then the system is not pressure tight and there is likely to be a hole in the vessel.
- Switch back to automatic open P1 to the chamber. On the manual setting check that the pressure remains stable.
- If no leak, on manual setting increase dial to 2 (40 mmHg).
- Adjust to automatic and observe the vessel under the microscope. If necessary, adjust the cannula holder to ensure no bend in the vessel.Check for leaks on manual setting.
- Repeat steps 5.6 - 5.7 increasing the dial to 3 (60 mmHg), 4 (80 mmHg) and so on until desired pressure is reached.
Approx. time = 15 - 30 mins
6. Incubating the pressurized vessel
- Connect the temperature controller set to 37°C.
- With a second peristaltic pump perfuse Krebs buffer (1 mL/min) into the vessel chamber bath.
- Connect aspirator to chamber ensuring only the top layer of the bath is removed.
- Incubate pressurized vessel at 37°C for 1 hour with the pressure set to automatic.
- Check the pressure is not leaking (switching to manual) periodically.
- The effects of various pharmacological interventions can be observed by simply adding the compound to the bath during incubation.
Approx. time = 1 hour
7. Perfusing the pressurized vessel with whole blood
- During incubation obtain at least 7.5 mL human whole blood/vessel collected in 40 U heparin/mL blood. Incubate in a 50 ml falcon at 37°C.
- 10 minutes before the end of the incubation label blood with VybrantDil (1:1000) for 10 minutes at 37°C in dark.
- After 10 minutes, collect blood in a syringe clearing any bubbles and attach onto a syringe pump with a heat jacket set at 37°C.
- Close off P1 transducer to the chamber/vessel and connect syringe and waste tubing.
- Purge blood at 1000 μL/min through the waste tubing to remove any bubbles.
- Open P1 to the chamber. The pressure servo will adjust automatically to maintain desired pressure.
- Perfuse blood at 100 μL/min.
- Using a fluorescent microscope coupled to a digital camera record two fields of the perfused vessel for 15 seconds at 1, 3, 5, 7.5 and 10 minutes.
- Post perfusion endothelial integrity and function can be assessed via pharmacological techniques such as myograph and adhesion molecule expression can be determined by immunohistochemistry to further validate the inflammatory response.
Approx. time = 15 mins
8. Representative Results:
A schematic diagram of the pressure chamber setup is shown in Figure 1. With a digital camera coupled to a fluorescent microscope results can be visualised instantly via live video recordings. Representative video images are seen in Figure 3 where leukocytes are considered to be adherent to the endothelium if they remained stationary for 10 seconds. With video recordings on continuous loop adhered cells can be counted as an average per field. While both low (Figure 3A) and high (Figure 3B) pressure will cause some amount of adhesion a significant increase in leukocyte adhesion at the higher intraluminal pressure is seen and this is also demonstrated quantitatively (Figure 4).

Figure 1. Pressurised ex vivo vessel chamber schematic. A cannulated vessel connected to a proximal (P1) transducer and a distal (P2) transducers that enables blood pressure to be manipulated within the vessel. Perfusion is through the P1 transducer and pressure is maintained via P2 transducer.

Figure 2. Cannuala with ties. Black polyester ties are attached to each cannula.

Figure 3. Representative video images. Dynamic cell adhesion (red arrow) under fluorescence at 80 mmHg (A) and 120 mmHg (B) after 10 minutes of perfusion.

Figure 4. Leukocyte adhesion in Sprague Dawley carotid arteries after 1 hour incubation at low (80 mmHg) and high pressure (120 mmHg). ***P<0.001, as analysed by 2-way repeated measures ANOVA using Bonferroni post hoc test.