Adaptation of a Haptic Robot in a 3T fMRI

9.5K views

⸱

DOI:

10.3791/3364

⸱

October 4th, 2011

 ,  ,  ,  , 

In This Article

Summary

The adaptation and use of a haptic robot in a 3T fMRI is described.

Abstract

Functional magnetic resonance imaging (fMRI) provides excellent functional brain imaging via the BOLD signal 1 with advantages including non-ionizing radiation, millimeter spatial accuracy of anatomical and functional data 2, and nearly real-time analyses 3. Haptic robots provide precise measurement and control of position and force of a cursor in a reasonably confined space. Here we combine these two technologies to allow precision experiments involving motor control with haptic/tactile environment interaction such as reaching or grasping. The basic idea is to attach an 8 foot end effecter supported in the center to the robot 4 allowing the subject to use the robot, but shielding it and keeping it out of the most extreme part of the magnetic field from the fMRI machine (Figure 1).

The Phantom Premium 3.0, 6DoF, high-force robot (SensAble Technologies, Inc.) is an excellent choice for providing force-feedback in virtual reality experiments 5, 6, but it is inherently non-MR safe, introduces significant noise to the sensitive fMRI equipment, and its electric motors may be affected by the fMRI's strongly varying magnetic field. We have constructed a table and shielding system that allows the robot to be safely introduced into the fMRI environment and limits both the degradation of the fMRI signal by the electrically noisy motors and the degradation of the electric motor performance by the strongly varying magnetic field of the fMRI. With the shield, the signal to noise ratio (SNR: mean signal/noise standard deviation) of the fMRI goes from a baseline of ˜380 to ˜330, and ˜250 without the shielding. The remaining noise appears to be uncorrelated and does not add artifacts to the fMRI of a test sphere (Figure 2). The long, stiff handle allows placement of the robot out of range of the most strongly varying parts of the magnetic field so there is no significant effect of the fMRI on the robot. The effect of the handle on the robot's kinematics is minimal since it is lightweight (˜2.6 lbs) but extremely stiff 3/4" graphite and well balanced on the 3DoF joint in the middle. The end result is an fMRI compatible, haptic system with about 1 cubic foot of working space, and, when combined with virtual reality, it allows for a new set of experiments to be performed in the fMRI environment including naturalistic reaching, passive displacement of the limb and haptic perception, adaptation learning in varying force fields, or texture identification 5, 6.

Protocol

1. Outside the scanner room

  1. Place the rolling table with the free end supported and the exterior end of the long handle detached.
  2. Check that the robot is switched off.
  3. Place the robot in the table socket and secure the aluminum safety plate over the robot with 2 screws.
  4. Attach the end effecter to the robot handle with the aluminum adapter and check that it moves freely.
  5. Attach the 10' parallel cable with aluminum shielding to the robot and check that the shielding is intact. Add extra foil if needed.
  6. Place the aluminum shielding box over the robot being careful to place the parallel and power cables into the groo....

Access restricted. Please log in or start a trial to view this content.

Discussion

The fMRI compatible robot opens up new possibilities for experiments in the neuroscience of motor control. The most critical step in the setup is the shielding of the robot to prevent artifacts in the fMRI, which we do in two steps. First, the robot itself is about 9' away from the bore with a long, lightweight, handle supported in its middle with a 3 degree of freedom joint . Second, the robot is encased in a 1/16"-1/4" aluminum box with a plastic conical (13" base diameter, 6" top diameter x 42&quo.......

Access restricted. Please log in or start a trial to view this content.

Disclosures

No conflicts of interest declared.

Acknowledgements

We would like to thank Kun Lu and Ronald Kurz for technical assistance. This work was supported by ONR MURI Award No.:N00014-10-1-0072, NSF grant #SBE-0542013 to the Temporal Dynamics of Learning Center, an NSF Science of Learning Center, and NIH grant #2 R01 NS036449-11.

....

Access restricted. Please log in or start a trial to view this content.

Materials

List of materials used in this article
NameCompanyCatalog NumberComments
Phantom premium 1.5/6dof, high force modelSensAble, Geomagic

References

  1. Ogawa, S., Lee, T. M., Kay, A. R., Tank, D. W. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc. Natl. Acad. Sci. U.S.A. 87, 9868-9872 (1990).
  2. Heeger, D. J., Ress, D. What does fMRI tell us about neuronal activity. Nat. Rev. Neur....

Access restricted. Please log in or start a trial to view this content.

Reprints and Permissions

Explore More Articles

fMRI CompatibilityAluminum ShieldingMotor Control ExperimentsNaturalistic ReachingForce Field AdaptationKinematic AnalysisPhantom CalibrationMRI Safety ProtocolsVirtual Reality Integration