1. Optical Setup
The entire optical system was described earlier15. Briefly, the optical tweezers system is based on an ytterbium continuous wave (CW) fiber laser operating at 1064 nm (IPG Laser GmbH). A spatial light modulator (SLM) (LCOS-SLM, model X10468-07 - Hamamatsu) varies the phase of the incoming IR laser beam to control the position of the trapping focus spot on the culture dish by computer generated holograms. The freely available Blue-tweezers software (web link on equipment table) generated holograms projected on the spatial light modulator. The interferometer for force spectroscopy measurements was based on a four-quadrant photodiode (QPD, S5980 with C5460SPL 6041 board - Hamamatsu) and a photodiode (PD, PDA100A-EC - Thorlabs).
The laser dissection source was a pulsed sub-nanosecond UV Nd:YAG laser at 355 nm (PNV-001525-040, PowerChip nano-Pulse UV laser - Teem Photonics). An acousto-optic modulator (MQ110-A3-UV, 355 nm fused silica -AA-Opto-electronic) controlled the power of the UV laser delivered to the sample.
The holographic optical tweezers and laser micro-dissector were integrated on a modified upright microscope (BX51 - Olympus) equipped with a 60X, 0.9 NA water dipping objective.The stage of the microscope is composed of a 3-axis linear DC motor micro-positioning system (M-126.CG1, Physics-Instruments) carrying a separate 3-axis piezoelectric nano-positioning stage (P-733.3DD, Physics-Instruments) to combine coarse movement of the sample with the sub-nanometer resolution of the piezo-stage. The microscope stage system was equipped with two control loops synergistically acting to maintain the trapping focus spot at the right position, depending on the selected working mode (position or force clamp, static or dynamic)16. In particular, an internal feedback loop acts on a piezoelectric stage, to keep the bead at a selected distance from the trap center. The other external loop controls the position of the motorized stage to exploit the region spanned by the piezo-actuator on a larger area than its available stroke17. When the piezo-stage reaches the limit of the available stroke in one direction, the external loop moves the micro stage in the opposite direction, thus the piezo recovers toward its central position because it is tracking the trapped bead adhered to the sample. When the piezo-stage reaches the central position of its course range, the micro stage halted. Further details of the system are reported in Guiggiani et al16,17.
A Peltier device (QE1 resistive heating with TC-344B dual channel heater controller - Warner Instruments) controls the temperature of the cell culture under the microscope (37 °C). In the culture, pH and humidity were maintained at physiological conditions by aerating a custom-designed polydimethylsiloxane (PDMS) sleeve (integrating the microscope objective) with humidified carbogen (95% O2, 5% CO2).
2. Cell Culture Preparation
All the experimental protocols were approved by the Italian Ministry of Health. Primary cultures were obtained from hippocampi of mice (C57BL6J, Charles River) at embryonic day 18 (E18).
- Neurons were plated at a concentration of 25,000 cells/ml on glass-bottom Petri dishes (P35G-0-14-C - MaTek Corporation). The low concentration of cultured cells is needed to avoid the formation of a dense network already in the first days in vitro. At the mentioned cell concentration, the likelihood of finding isolated cells with a longer neurite not connected to other cells is much higher.
3. Bead Coating
- Polymer microspheres (Ø 4 μm, COOH-terminated- Bangs Laboratories, product code PC05N/6700) were coated with poly-D-lysine following the procedure described in the polyLink Protein Coupling Kit (Polysciences, product code 19539). The beads are coated with the same molecule used to cover the culture support and favor cells adhesion.
4. Choose Isolated Neuron. Detach A Bead from the Culture Substrate, Trap and Move it Next to the Neuron
- Put the culture Petri dish on the microscope stage. After setting the focus on the sample by stage motion, correct the position of the illuminating objective condenser to set Kohler-illumination. Aligning the illumination optics to set the Kolher-illumination condition is necessary to improve the bright field imaging quality. Using such alignment conditions, it is also necessary to maximize the IR laser collection efficiency (through objective condenser), from the scattering trapped probe, to perform its interferometric tracking by the QPD and PD.
- Pipette a few μl of the coated microsphere stock solution into the culture dish. Microspheres will deposit and adhere to the culture substrate. The number of μl injected into the culture dish depends on the microsphere concentration of the stock solution. The ideal condition is to have between one and two microspheres per imaging field of view. When too many microspheres are added to the culture dish, they can already attach randomly to the cultured neurons.
- Move around in the culture dish, searching for an isolated neuron with a longer neurite (the axon), and save the position of the microscope stage.
- Move around and search for a bead adhered to the culture support.
- Turn on the IR trapping laser. At this stage, no holograms are projected on SLM, and the position of IR laser spot coincides with the UV laser spot position. Set the axial position of the IR spot 2-3 μm above the culture support surface, by microscope stage motion.
- Set the UV laser power delivered to the sample to less than 1 μW. The UV laser dissector has been previously calibrated14:
5-6 μW the glass support is ablated
4 μW a neuronal connection is completely dissected
2.5 μW a neurite is partially lesioned
‹ 1 μW a shock wave is produced in the culture dish. The optical shock wave is strong enough to detach the bead from the glass support.
- Turn on the UV laser, while leaving the IR laser on, and move the UV spot above the adhered bead by microscope stage motion. The bead detaches from the surface, and it is trapped by the IR laser. Then, turn off the UV laser.
- Move the trapped bead several micrometers above the glass support (20-30 μm). Lifting the bead to this position will prevent it from contacting other cells while it is being moved toward the previously chosen neuron. Bring the bead to the previously saved stage position. Set the stage speed to a low value (10 μm/sec) so the drag fluid force does not exceed the optical trapping force.
5. Move the Trap Position with Respect to the Laser Dissector Spot and Axon Position by Computer Generated Hologram, and Calibrate the Optical Tweezers Stiffness
- Move the bead toward the glass support to visualize the axon. The bead is held above the cell to avoid contact with it (about 5 μm above the glass support).
- Move the microscope stage to move the UV focus spot on the center of the width of the axon. Save the current stage position.
- Move the IR focus spot position by computer-generated hologram. In this step, the stage is halted in the position chosen in the previous step. When the computer generated hologram is projected on the SLM, the trapped bead is moved with respect to the axon. We move the IR laser spot to have the trapped bead aligned on the center of the neurite, with the UV laser spot centered on the same neurite too. The IR spot and thus the trapped bead are positioned along the neurite 5-10 μm away from the UV spot. We move the position of the IR with respect to the UV spot depending on the neurite geometry. In case the optical tweezers are not equipped with an SLM system, the position can be varied by tilting mirrors, or acoustic optic deflectors, on the IR optical path. The IR laser spot may be positioned 5-10 μm away from the UV spot to avoid optical interaction of the dissecting beam with the trapped probe, which could influence its Brownian motion and alter the force spectroscopy traces.
- After defining the new IR spot position with respect to UV spot and axon position, move the stage to position the trapped bead away from the axon and avoid collision with it. Move the axial position of the trapped bead to about 2 μm above the cover glass.
- Align the QPD to center the interference fringes on it: x and y QPD differential signals are zeros when the QPD is centered. Acquire 5 sec of the Brownian motion of the trapped bead by the interferometer, at 50 KHz sampling rate. Obtain the stiffness (pN/nm) and sensitivity (V/nm) of the optical trap by the power spectrum method18.
6. Attach the Bead to the Axon. Perform Axotomy and Simultaneous Force Measurement
- Raise the trapped bead position about 4 μm above the cover glass, and move it to the previously saved stage position (see step 3.2).
- Move the trapped bead down toward the axon until it contacts the neurite. The collision between the trapped bead and the axon is monitored by the z QPD signal detecting displacement of the trapped bead in the axial direction.
- Wait 10 sec with the trapped bead pushed against the axon to favor its adhesion to the neurite membrane.
- Move the micro stage to displace the trapped bead from the axon, and thus verify its adhesion to the cell membrane. If the bead adhered, it escapes from the optical trap.
- Move back the laser trap on the adhered bead and switch on the force-clamp loop with force condition equal to zero. In such a way, the piezo-stage repositions the bead, attached to the cell, on the optical trap center. If the system does not have a feedback loop control, the laser trap position can be moved by the microscope stage to center it on the adhered probe. The center of the trap is reached when the QPD signals are the same as when the bead is trapped and not adhered to the cell (x and y QPD signals equal to zero, and the z QPD signal gives the same voltage sum of the four quadrants as when the bead is trapped far from any surface) .
- Set a force-clamp condition on the z axis, positioning the trapping laser slightly over the center of the bead (about 100 nm), to generate a pretension on the adhered trapped probe. In case the system is not equipped with a force-clamp control, the optical trap position can be raised up at steps of 25 nm by the microscope stage. The z QPD signal allows monitoring. Therefore, use this signal to set the position of the laser trap with respect to the adhered probe. Such pre-tension is needed to sense the strain on the membrane after the axonal lesion, and the consequent decrease of Brownian motion of the adhered probe. A similar approach has been proposed to measuring the release of tension in a membrane tether by video imaging19.
- Switch-off the force-clamp feedback to measure the force on the adhered probe in the position-clamp condition (the piezo-stage is blocked).
- Start simultaneous recording of the trapped probe positions through the interferometer (20 KHz sampling frequency), and the cell during laser axotomy time-lapse bright-field imaging (20 Hz frame rate).
- Turn on the UV laser to deliver laser pulses until a lesion becomes visible on the image of the axon (Usually 200-400 optical pulses are needed. Energy per pulse: 25 nJ), then turn off the UV laser.
- Continue recording by the interferometer for about 3 min, until the x, y and z QPD traces reach a plateau.
7. Quantify the Total Tension Release
- Convert the recorded QPD traces (in Volts) by the calibrated sensitivity of the optical trap, to obtain the respective traces in nanometers.
- Filter the x, y, and z displacement traces by a high pass filter with cut off at 10 Hz.
- Calculate the total variance of the filtered traces representing the Brownian motion of the trapped bead. Calculate the variance at 25 msec steps for overlapping time windows of 500 msec (10,000 data points)20. The high pass filtering of traces excludes the changes of Brownian motion due to the membrane fluctuation or cortical actin motion. The Brownian motion of the bead is reduced by the optical forces and the adhesion forces on the cell membrane. When the membrane is strained because of the release of tension its viscosity is increased, and thus the Brownian motion of the bead, detected immediately after axotomy, starts to decrease.
- Define t0: the beginning of decrease of the Brownian motion variance, after the delivery of UV laser energy. The time instant t0 indicates the beginning of membrane strain.
- Define t1: the end of decrease of the Brownian motion variance (when it reaches a plateau). The time instant t1 indicates the end of membrane strain.
- Perform video tracking of debris or a scratch on the cover glass support, to measure any drift of the sample during the force measurement. To track the particle on the glass support, we first apply thresholds to the bright-field images, to obtain a stack of binary images with the particle in white on a black background. Then, we track the particle center of mass with sub-pixel accuracy (using the freeware ImageJ software).
- Subtract the measured drift from the displacement QPD traces. Multiply the drift-corrected QPD traces by the respective calibrated optical stiffness (kx, ky, kz) to obtain the Fx, Fy, Fz traces in piconewtons.
- Calculate the total force trace as Ftot = √(Fx2+Fy2+Fz2).
- Calculate the total release of force between t0 and t1 as Freleased= Ftot(t1) - Ftot(t0). The force measured at t0 has to be subtracted because it is due to the displacement of the probe from the trap center, when the bead adheres to the neurite.
- Calculate the neurite contact area between the trapped bead and the UV laser spot position: on the bright field image measure the long (L) and short (D) axes of the neurite between the trapped bead and the UV laser spot position. The axon contact area is Aaxon = L x D.
- Normalize the total released force Freleased by the axon contact area Aaxon.